Andrea Dell'Orso1
1Radiologia, Azienda USL Toscana Centro, Empoli, Florence, Italy
Synopsis
Keywords: Image acquisition: Motion Correction, Image acquisition: Whole Body, Image acquisition: Artefacts
In Italy the Healthcare System is Public. The Population is 59,11 millions (2021), 39% are affected by a chronic disease and the main causes of death are due circulatory system 37.49%, and tumors 30.36%.
In Italy there are a total of 1572 MRI scanner. Accessibility to the MRI is guaranteed within 72 hours for urgent services, and within 90 days for programmable exams.
The main field of research for radiographer is on artifact reduction, like the post-processing software O.P.E.R.A. Orthogonal-Phase-Encoding-Reduction-Artifact. It works on DICOM images and doesn't need hardware modifications, hence it could be used on different MR scanner.
In
Italy the Healthcare System is Public.
The Population is 59,11 millions (2021), 39%
are affected by a chronic disease and the main causes of death are due
circulatory system 37.49%, and tumors 30.36%.
In Italy
there are a total of 1572 MRI scanner. Accessibility to the MRI is guaranteed
within 72 hours for urgent services, and within 90 days for programmable exams.
The main
field of research for Italian radiographer is on improve the image quality and artifact
reduction. Artifacts are an important problem leading to reduced image quality,
loss of diagnostic information, and time consuming (1). These problems led to a
degradation of the radiographer's work. Artifacts can be reduced by using
several algorithms, most of them specific for a single anatomical region,
working at K-Space reconstruction level, and often leading to prolongation of
the scan time (2,3).
The software
OPERA: Orthogonal Phase Encoding Reduction of Artifact (4) is a post-process
software for the automatic correction of artifacts generated by motion and
wrap-around. It works on DICOM images and doesn't need hardware modifications,
hence it could be used on different MR scanner.
Purpose
This
prospective study were tested the MR value of OPERA on the reduction of
artifacts, the reduction of the number of re-scan sequences due to severe
artifacts and the reduction of additional sequences required by radiologists.
Method
In MRI the
position of ghosts, motion and aliasing artifacts is predictable along columns
or rows of pixels, by knowing the phase encoding direction. OPERA algorithm
compares the intensity values of the pixels of two DICOM images acquired on the
same anatomy by using the same parameters but with orthogonal phase-encoding
directions. In order to avoid prolongation of the exam time, the Number of
Signal Averages (NSA), (or similar) can be exploited by making an even NSA,
with half of them being distributed along one phase encoding direction and the
remaining half being distributed along the opposite one.
This study
was conducted in a medium-sized general public hospital by using an Espree
Siemens 1.5T MR scanner by comparing 3 full calendar weeks (MR exams n=302:
Group A) with the following 3 weeks
where OPERA procedure was applied to all MR sequences imaging of non-moving
organs (MR exams n=302; Group B). A total of 604 MR exams were acquired on 585
patients (54.8% females; mean age 52.7±17.2 years)]. Imaged anatomical regions
were: brain, n=182; spine, n=138; shoulder, n=64; elbow, n= 26; pelvis, n=96; knee,
n=68; ankle, n=30.
During the
first 3 weeks (Group A), the sequences were acquired with the Standard
parameters required by good clinical practice. For the next 3 weeks (Group
B),OPERA procedure was used. Each sequence was acquired with vertical phase
encoding direction and with horizontal phase encoding direction and
NSAOPERA=NSASTD/2. Both sequences were sent to the PACS and the pair was
processed in a remote-server by OPERA algorithm to obtain a third sequence with
artifact-reduced images that were returned to the PACS.
For both A
and B Groups, the radiographer noted the sequences with several motion
artifacts that were been repeated, and the additional sequences, over the
standard protocol, required by radiologists for the diagnosis. Radiographers
and radiologists have a long-standing
MRI expertise (>15 years.)
The MedCalc
17.2 statistical software was used for statistical analysis.
Results Significant motion artifacts were identified
on sequences in 18.42 % (n= 279) of Group A, and in 13.34% (n= 202) Group
B ( Fig 1).
Additional
sequences were required by radiologist
in 12.48 % (n= 189) of Group A, and in
9.7% (n= 147) of Group B. (Fig 2)
Overall, as
compared to the standard procedure, OPERA algorithm consistently reduced the
percentages of repeated sequences of 5.9% (two tailed p value= 0.0639) while
reducing the percentage of additional sequences of 2.17% (Hodges-Lehmann median
difference 5,9200; 95% Confidence interval -1,2900 to 13,3700; Mann-Whitney U,
10,00; two tailed p value= 0.3379).
Conclusion
The present study confirms the potential of OPERA algorithm in reducing both
the need for MR sequence repetition due to severe artifacts and MR sequences
addiction for better diagnosis. Even the software could be usable on
uncooperative patient or patients with pathology which do not allow motionless
posture during the MR examination. The method was demonstrated to be applicable
to several sequences and anatomical regions.
Elbow
studies were characterized by severe motion artifacts due to the patient
position and the extremely pain in post-traumatic cases.
The
remarkable number of sequences
“not-repeat” could potentially have a positive side effect in saved
time. The peculiarities of the present algorithm allows a wide applicability to
different commercially available MRI machines.
OPERA would
simplify the radiographer work by eliminating the parameter “ phase encoding
direction”, oversampling, or saturation regions.
Acknowledgements
We sincerely thank:
all the radiographers and medical radiologists of San Giuseppe Hospital in Empoli for the opportunity to acquire the images during the clinical practice sessions;
Petronella Samuels and Nina Salaman for valuable technical advice;
the Opera Research Team: Vincenzo Positano, Francesco d'Errico, Francesco De Milato, Giovanni Arisi, Benedetta Banchi, Aldo Taddei, Claudio De Felice.
References
1 . Andre JB, Bresnahan BW, Mossa-Basha M, Hoff MN, et
al.; Toward Quantifying the Prevalence, Severity, and Cost Associated With
Patient Motion During Clinical MR Examinations. J Am Coll Radiol. 2015
Jul;12(7):689-95. doi:
10.1016/j.jacr.2015.03.007.
2. Kruger DG, Glenn SS, et al. ; An orthogonal correlation algorithm for ghost
reduction in MRI. MRM 1997; 38: 678-686;
3. Forbes KP, Pipe JG, Bird CR, Heiserman JE (2001)
PROPELLER MRI: clinical testing of a novel technique for quantification and
compensation of head motion. J Magn Reson Imaging 14:215–222
4.
Dell’Orso, A., Positano, V., De Felice C. et al. OPERA: a novel method to reduce ghost and aliasing artifacts. Magn Reson
Mater Phy 34, 451–467 (2021). https://doi.org/10.1007/s10334-020-00881-1