xiaoyu jiang1, Junzhong Xu2, and John Gore2
1Vanderbilt University Medical Center, Nashville, TN, United States, 2Vanderbilt University Medical Center, nashville, TN, United States
Synopsis
Keywords: Microstructure, Diffusion/other diffusion imaging techniques
Motivation: Noninvasive measurement of liver cell sizes and densities is clinically important.
Goal(s): Assess the repeatability of liver cell size and density mapping using MRI cytometry, a diffusion MRI-based imaging technique.
Approach: In vivo MRI cytometry was repeated twice with intervals of 2-7 days for six healthy subjects.
Results: MRI-derived liver cell size and density exhibited high repeatability, with significant correlations (r2 = 0.89 and 0.91 for cell size and cell density, respectively; [P< 0.0001]).
Impact: Our findings indicate that a liver MRI cytometry
protocol can be implemented on standard clinical 3T scanners with satisfactory
repeatability in under 12 minutes, thus affirming the potential for broader clinical
adoption in the future.
Introduction
Liver
biopsy remains the standard method for evaluating the etiology and extent of
diseases of the liver. Although liver biopsy is generally safe, sampling
errors, rare complications, intra/inter-observer variability, and significant
patient anxiety, may all occur in practice [1]. These
factors have led to keen interest in the development of non-invasive tests and
imaging techniques for the diagnosis and management of liver diseases. The cell
size and cell density, particularly their variations over time, are fundamental
characteristics of liver, and measurements of cell sizes and densities provide
diagnostic biomarkers of several normal and pathological processes.
Measurements of cell sizes and densities thus have high clinical significance
but currently can be obtained only by liver biopsy.
Our laboratory has
developed MRI cytometry, a diffusion MRI-based imaging technique that extracts
microstructural parameters, such as mean cell size and cell density, from measuremnts
of hindered/restricted diffusions in solid tissues[2]. We hypothesize that MRI cytometry can
characterize pathological changes in the microstructure of the human liver, potentially
reducing the need for liver biopsy. Previous research has demonstrated the
feasibility of MRI cytometry on clinical 3T scanners using a clinically
attainable protocol called IMPULSED (Imaging Microstructural Parameters Using
Limited Spectrally Edited Diffusion) [3]. This study
extends this concept by evaluating the repeatability of MRI-derived cell sizes
and cell densities in healthy human subjects. Methods
Theory: MRI cytometry integrates measurements of water diffusion rates over
various time scales, corresponding to probing cellular microstructures at
different distances. The size range of utmost relevance in liver tissues is 5
µm to 25 µm (e.g., hepatocytes ~ 15 – 25 µm, inflammatory cells ~ 5 – 10 µm),
which corresponds to diffusion times of approximately 5 - 70 ms. These
diffusion times can be achieved using a combination of OGSE (oscillating
gradient spin echo) and PGSE (pulsed gradient spin echo) measurements.
Microstructural properties are extracted by fitting multi-b
value-multi-diffusion time fat-suppressed diffusion-weighted MRI signals to a
three-compartment model (blood, intra and extracellular water).
In vivo human imaging: MRI cytometry was repeated twice at 2-7 days apart
for six healthy subjects using a Philips Ingenia CX 3T scanner with a dStream
TorsoCardiac coil. A PGSE
sequence with diffusion gradient duration δ / diffusion gradient separation Δ =
12/74 ms was used to collect diffusion data at a long diffusion time
(70 ms). Shorter diffusion times were achieved using a cosine-modulated
trapezoidal OGSE sequence with gradient frequencies of 25 and 50 Hz. Five b
values (0, 250, 500, 750, and 1000 s/mm2) were used for PGSE and 25
Hz OGSE acquisitions. For 50 Hz OGSE acquisitions, four b values (0, 100, 200,
and 300 s/mm2) were used. Other imaging parameters were TR/TE=2000/110ms;
FOV=192×192mm; in-plane resolution = 4×4 mm; 5 slices; slice thickness=10 mm;
respiratory-gated, single shot EPI; SENSE factor=2; fat suppression with
SPAIR. The total scan time ≈ 12 mins. The SNRs of b0 images are about 20. High
resolution (0.75 x 0.75 x 5 mm3) T1W images covering the whole liver were
collected for registration of MRI-derived microstructural maps obtained at two
different time points for the same subject.
Statistics: Repeatability statistics follow the methods previously
described by Bland and Altman [4]. Kendall’s tau test was performed to test whether
the magnitude of the differences between the first and second measurements was
correlated with the parameter mean of the repeated measurements. The Wilcoxon
signed-rank test was used to test the null hypothesis of no bias (ie, median
difference is zero) between the first and second measurements, with the
Spearman correlation testing the effectiveness of the pairing. The
repeatability of each MRI metric (ie, cell size and cell density) was assessed.
Additionally, the 95% confidence interval (CI), the root-mean-squared
deviation, the within-subject standard deviation, and the repeatability coefficient
(RPC) were calculated for each MRI metric. This repeatability coefficient
defines the magnitude of the maximum difference (absolute value) between
repeated observations expected in 95% of paired observations.Results and discussion
Repeatability scans of the same subject’s liver tissue showed an average
percent difference of 0.06% and 3% in cell size and cell density, respectively,
between the two scans. Repeatability of MRI-derived cell size and cell density
in the liver was not significantly different with the pairing significantly
correlated (r2 = 0.89 and 0.91 for cell size and cell density,
respectively; [P< 0.0001]).
Additionally,
the difference between repeated measurements was independent of the mean.Conclusion
Our
findings demonstrate the feasibility of implementing a liver MRI cytometry
protocol on standard clinical 3T scanners with satisfactory repeatability in
under 12 minutes, affirming its potential for wider clinical adoption in the future.Acknowledgements
No acknowledgement found.References
1. Tobkes, A.I. and H.J. Nord, Liver biopsy: review of methodology and
complications. Dig Dis, 1995. 13(5):
p. 267-74.
2. Jiang, X., et al., MR cell size imaging with temporal diffusion
spectroscopy. Magn Reson Imaging, 2021. 77: p. 109-123.
3. Jiang, X., J. Xu, and J.C. Gore, Mapping hepatocyte size in vivo using
temporal diffusion spectroscopy MRI. Magn Reson Med, 2020.
4. Bland, J.M. and D.G. Altman, Measuring agreement in method comparison
studies. Stat Methods Med Res, 1999. 8(2):
p. 135-60.