Guangqi Li1, Sisi Li1, and Hua Guo1
1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China
Synopsis
Keywords: System Imperfections, System Imperfections: Measurement & Correction, Trajectory correction
Motivation: Gradient imperfections typically cause ringing and blurring artifacts in spiral imaging. Gradient impulse response function (GIRF) characterizes the gradient system and can be applied for gradient prediction in image reconstruction. However, the low-pass characteristic of the gradient system could cause a penalty of the achieved resolution.
Goal(s): This work aims to improve the GIRF-based spiral trajectory correction.
Approach: A joint pre-emphasis and post-processing method was introduced to correct spiral trajectory. The effectiveness of our proposed method was tested on the phantom and in-vivo experiments.
Results: The ringing and blurring artifacts were further mitigated using our proposed method.
Impact: A
joint pre-emphasis and post-processing correction strategy was proposed in this
study. The results of phantom and in-vivo experiments indicate that our
proposed method yields good image quality and effectively reduces the loss of the
actual resolution.
Introduction
In spiral imaging,
gradient imperfections typically result in ringing artifacts and image
blurring. Gradient impulse response
function (GIRF) 1-7 enables the correction of the spiral gradient
waveform by predicting the actual gradient output through convolution of the
nominal gradient waveform (input) with GIRF. The gradient system-related image artifacts
can be significantly reduced in the images reconstructed using the GIRF-based predictions.
However, the actual acquired resolution is compromised due to the low-pass
characteristic of the gradient system 2. Thus, in this study, a joint
pre-emphasis and post-processing method is proposed to further enhance
GIRF-based correction. The overall image quality and the achieved resolution
are improved. Theory
Under the
assumption that the gradient system is linear and time-invariant (LTI), GIRF-based
correction can be performed to improve image quality. The schematic diagram of
our proposed method is shown in Figure 1.
1.
GIRF-based post-processing correction
For conventional GIRF-based
correction, GIRF is used to predict the actual outputs of a gradient system:
$$g_{out}(t)=g_{in}(t)*h(t)$$
The low-pass
characteristic of the gradient system reduces the actual resolution. However,
the conventional GIRF-based correction cannot compensate for this problem.
2.
Pre-emphasis
Pre-emphasis is basically
a high-pass filter designed to counteract the system-induced waveform
distortion by modifying the gradient input. The pre-emphasized
gradient $$$g_{pre}(t)$$$ was played out during spiral imaging, then the actual gradient outputs would be
closer to the nominal waveform, i.e.,
$$$g_{out}(t)\approx g_{in}(t)$$$.
However, the eddy-current compensation may be insufficient because the pre-emphasis
filter was determined by a limited number of exponential terms and the
bandwidth of gradient amplifiers is limited.
3.
Joint pre-emphasis and post-processing correction
In this work, to
overcome the drawbacks of the above two methods, a joint pre-emphasis and
post-processing method was proposed. Both the behavior of eddy-current
compensation and the characteristics of the gradient system can be described by
combined GIRF
$$$h_{com}(t)$$$.
This is subsequently employed to correct the spiral trajectory, i.e., $$$g_{out}(t)=g_{in}(t)*h_{com}(t)$$$.Materials and Methods
All experiments were
performed on an Ingenia CX 3.0T scanner (Philips) using a 32-channel head coil.
1.
GIRF Measurements
The GIRF
measurements were performed with a spherical phantom using our previously
proposed interleaved off-isocenter method 8. Traditional GIRF
$$$h(t)$$$
is determined by measured waveforms and
inputs. Subsequently, to measure the combined
GIRF
$$$h_{com}(t)$$$, input gradients were pre-emphasized and then played out.
2.
Data acquisition
Both phantom and
in-vivo experiments were performed to validate the effectiveness of the
proposed correction method. The spiral waveforms with and without pre-emphasis
were used respectively to acquire data in the following experiments.
Experiment 1:
FOV = 210×210 mm2, resolution = 1.0×1.0×4.0 mm3, TE/TR = 50/3000
ms. A 20-shot uniform-density spiral (UDS) acquisition with a readout window of
8.0 ms was used to obtain axial images of a spherical phantom and ACR phantom.
Experiment 2:
FOV = 210×210 mm2, resolution = 1.0×1.0×4.0 mm3, TE/TR = 50/3000
ms. The axial, coronal and sagittal slices of a phantom were acquired using a
24-shot variable-density spiral (VDS) acquisition with a readout window of 15.7
ms.
Experiment 3:
FOV = 210×210 mm2, resolution = 1.0×1.0×4.0 mm3, TE/TR = 50/3000
ms. The axial, coronal and sagittal slices covering the brain of a subject were acquired using a 20-shot UDS acquisition with a readout window of 8.0
ms.
3.
Image reconstruction
For comparison
purposes, the spiral images were reconstructed using the following four
different trajectories. (i) nominal trajectories $$$k_{nominal}(t)$$$, (ii) predicted trajectories $$$k_{pred}(t)$$$ obtained by the conventional GIRF, (iii) pre-emphasized trajectories $$$k_{pre}(t)$$$ and (iv) predicted
trajectories $$$k_{pred}^{com}(t)$$$ obtained
by our proposed combined GIRF.Results and Discussion
The magnitude of
measured conventional GIRF $$$H(\omega)$$$, measured
combined GIRF $$$H_{com}(\omega)$$$, and the
spectrum of nominal spiral gradients $$$G_x$$$ and $$$G_y$$$
are shown in
Figure 2. The bandwidth
of $$$H_{com}(\omega)$$$ is wider. Pre-emphasis is favorable for fast acquisitions with rapid gradient switching.
The spherical phantom and
ACR phantom images acquired using UDS are shown in Figure 3. Figure 4 shows the
axial, coronal and sagittal phantom images acquired using VDS. The in-vivo
results reconstructed using four different trajectories are shown in Figure 5. GIRF-based correction and
pre-emphasis can yield good image quality. Residual ringing and blurring artifacts
are removed and more details are retained in the reconstructions using our
proposed method. Furthermore, the maximum
value of $$$k_{pred}(t)$$$
is lower than that of $$$k_{pred}^{com}(t)$$$. The
results reconstructed by $$$k_{pred}^{com}(t)$$$ show better image quality with effectively reduced
loss of actual resolution.Conclusion
In
this study, a joint pre-emphasis and post-processing method was proposed to
further enhance the performance of GIRF-based trajectory correction. The spiral
images reconstructed using our proposed method show superior image quality
with effectively minimized loss of actual resolution.Acknowledgements
No acknowledgment
found.References
1.
Vannesjo SJ, Haeberlin M, Kasper L, et al. Gradient system characterization by
impulse response measurements with a dynamic field camera. Magn Reson Med 2013;
69: 583–593
2.
Vannesjo S J, Graedel N N, Kasper L, et al. Image reconstruction using a
gradient impulse
response
model for trajectory prediction [J]. Magn Reson Med, 2016, 76(1): 45-58
3.
Campbell-Washburn A E, Xue H, Lederman R J, et al. Real-time distortion
correction of spiral and echo planar images using the gradient system impulse
response function [J]. Magn Reson Med, 2016, 75(6): 2278-2285
4.
Rahmer J, Schmale I, Mazurkewitz P, et al. Non-Cartesian k-space trajectory
calculation based on concurrent reading of the gradient amplifiers' output
currents [J]. Magn Reson Med, 2021, 85(6): 3060-3070.
5.
Stich M, Wech T, Slawig A, et al. Gradient waveform pre-emphasis based on the
gradient
system
transfer function [J]. Magn Reson Med, 2018, 80(4): 1521-1532.
6.
Robison R K, Devaraj A, Pipe J G. Fast, simple gradient delay estimation for
spiral MRI [J].
Magn
Reson Med, 2010, 63(6): 1683-1690
7.
Addy N O, Wu H H, Nishimura D G. Simple method for MR gradient system
characterization
and
k‐space trajectory estimation [J]. Magnetic resonance in medicine, 2012, 68(1):
120-129.
8. Li G, Chen S, and
Guo H. Accelerate GIRF measurement using an interleaved off-isocenter method.
In Proceedings of the 27th Annual Meeting of ISMRM. 2019; 1774.