Çelik Boğa1 and Anke Henning1
1UT Southwestern Medical Center, Dallas, TX, United States
Synopsis
Keywords: Parallel Transmit & Multiband, Brain, bSSFP, parallel transmission, transmit field inhomogeneity
Motivation: Elimination of the transmit field inhomogeneity effects in the brain for T2 contrast at 7T.
Goal(s): Removing the transmit field inhomogeneity effects in balanced steady state free precession acquisitions using Bilateral Orthogonality Generative Acquisitions method.
Approach: Bilateral Orthogonality Generative Acquisitions method is implemented at dual channel parallel transmit 7T system using several balanced steady state free precession acquisitions with varying scan parameters to eliminate the transmit field inhomogeneity effects and reduce the banding artifacts effects via different combining schemes in the final image.
Results: T2 contrast was obtained without the transmit field inhomogeneity effects but residual banding artifacts exists in the final images.
Impact: T2 contrast can be achieved without the transmit field inhomogeneity effects in the brain using balanced steady state free precession sequence.
Introduction
Balanced Steady State Precession (bSSFP)
sequence has wide range of applications1-3 due to its versatility
and fast acquisition. While it’s a gradient echo (GRE) acquisition, in bSSFP, T2 weighting can be achieved instead of T2*1-3. However, acquired images suffers
from the banding artifacts due to the off-resonance effects1-3.
In this work, Bilateral
Orthogonality Generating Acquisitions (BOGA) method4, which was recently
introduced to yield homogeneous T2*
contrast in a dual channel parallel
transmission (pTx) 7T system without need for B1+
calibration, is implemented with bSSFP for T2
contrast unaffected by the transmit field
inhomogeneity. For reducing the effects of banding artifact, sum of squares and
maximum intensity combining methods are implemented.Theory
For a dual channel pTx system, the flip angle can be written as $$${\alpha}=(s_1+s_2)e^{i\mu_n}{\gamma}e^{i\delta}$$$5. $$$s_1$$$ and $$$s_2$$$ are the channel specific transmit sensitivities and $$$\mu_1$$$ and $$$\mu_2$$$ are the transmission phases of each channel. $$$\gamma$$$ denotes the magnitude of the complex flip angle, $$$\delta$$$ denotes the phase of the complex flip angle. Defining $$$q_n=s_ne^{i\mu_n}e^{i\delta}$$$ the flip angle expression for each channel can be simplified to $$${\alpha}_{1,2}=q_{1,2}\beta$$$.Using small flip angles, the bSSFP signal equation1,2 with low flip angle can be written as $$$S=\frac{{\rho}(1-e^{-TR/T_2}e^{-i\phi})e^{-TE/T_2}\alpha}{1+e^{-2TR/T_2}+2e^{-TR/T_2}cos(\phi)}$$$. Where $$$\rho$$$ is the proton density, $$$\phi$$$ is the phase accumulated due to the off-resonance effects, TR is the repetition time and TE is the echo time. Defining $$$A=\frac{{\rho}(1-e^{-TR/T_2}e^{-i\phi})}{1+e^{-2TR/T_2}+2e^{-TR/T_2}cos(\phi)}$$$, $$$S$$$ can be simplified as $$$S=Ae^{-TE/T_2}\alpha$$$.
For
the bSSFP implementation of the BOGA method, four input images are obtained
with the same TR to fix the locations of the banding artifacts
and effect of T2 decay in the magnitude of the steady state
magnetization. Moreover, 2 sets of input images are obtained where the first
set has no additional RF phase and second set has the additional RF phase of 180o to maximize the distance between banding
artifacts between input sets for the reduction of the banding artifact
mitigation methods. For both sets, input images are
defined as $$$S_1=Ae^{-TE/T_2}(q_1+q_2)$$$
and
$$$S_2=Ae^{-TE/T_2}(-q_1+q_2)$$$
and $$$S_{3,4}=Ae^{-TE/T_2}q_{3,4}$$$.
By combining the four signals as in4, $$$C_1$$$ and $$$C_2$$$ can be defined as
$$$C_1=S_3^*S_1+S_4S_2^*$$$ and $$$C_2=S_4^*S_1-S_3S_2^*$$$
. Final image is obtained as $$$I=\frac{\sqrt{0.5(C_1^*C_1+C_2^*C_2)}}{S_3^*S_3+S_4^*S_4}$$$4. Sum of squares and maximum
intensity combination is implemented for the input images or the resulting images to determine
the optimal order of image combination and BOGA method.
Methods
For the acquisition of the data, a 7T
Philips Healthcare whole body human MRI system with two transmit channels is used along with a
32 channel receive Nova Medical head coil. For a healthy volunteer, the four above
described bSSFP acquisitions were acquired each with voxel size of 1x1x1 mm, 256x256x192
acquisition matrix, TFE factor of 3800, compressed SENSE factor of 9 and 5o flip angle. TE1/TR = 12.3/25 ms and
TE2/TR = 2.3/25 ms are used for first two and last two acquisitions.
The final
image has the effective echo time
of 10 ms. For
the second set of input images, 20 Hz offset is applied to generate the aforementioned
RF phase difference of $$$\pi=2{\pi}f_{offset}TR$$$.Results
Figure
1 demonstrates the input images for the input sets with 0 and 20 Hz frequency
offsets for the individual application of BOGA method. Figure 2 demonstrates
the sum of square and maximum intensity combinations of the input images from
Figure 1.
Figure 3 illustrates the T2 images obtained via BOGA
method using the input images demonstrated in the first column of the Figure 1. Second column demonstrates the sum of squares and maximum
intensity combinations of
images in the first column. Whereas
Figure 4 shows the final transmit field inhomogeneity corrected T2
images obtained via BOGA
method with combined input images.
All
images obtained with this
method are free of transmit field inhomogeneity effect, but images at the first column of the Figure 3 are significantly affected from the banding artifacts. While sum
of squares and maximum intensity combinations reduce these affects, they do not
completely eliminate it as can be seen in second column of the Figure 3 and Figure 4. Maximum
intensity combination performs similarly when applied to input images or
resulting T2 images. Whereas, sum of
square combination performs worse when applied to the resulting images.Discussion and Conclusion
In
this work, T2 images without transmit field
inhomogeneity effect are obtained using BOGA method with bSSFP acquisitions. Effect
of the banding artifacts in the final images can be reduced using maximum
intensity and sum of squares combinations but it cannot be fully eliminated.
Future work includes implementation of banding artifact elimination method,
instead of image combination, for obtaining homogeneous
images without the banding
artifact effects.Acknowledgements
This work was performed in the Advance Imaging Research Center at University of Texas Southwestern Medical center Dallas. This work was supported by Cancer Prevention and Research Institute of Texas (CPRIT) grant / RR180056.References
1. Scheffler
K, Lehnhardt S. Principles and applications of balanced SSFP techniques.
European Radiology. 2003;13(11):2409-2418.
doi:https://doi.org/10.1007/s00330-003-1957-x
2. Bieri
O, Scheffler K. Fundamentals of balanced steady state free precession MRI.
Journal of Magnetic Resonance Imaging. 2013;38(1):2-11.
doi:https://doi.org/10.1002/jmri.24163
3. Miller
KL, Hargreaves BA, Lee J, Ress D, Christopher deCharms R, Pauly JM. Functional
brain imaging using a blood oxygenation sensitive steady state. Magnetic
Resonance in Medicine. 2003;50(4):675-683. doi:https://doi.org/10.1002/mrm.10602
4. Boga C
and Henning A. “Bilateral Orthogonality Generating Acquisitions Method for
Homogeneous T2* Images Using Dual Channel Parallel Transmission at 7T” In
Proceedings of the 31st Annual Meeting of ISMRM, Toronto, Canada, 2023
5. Grissom W, Yip C-yu, Zhang Z, Stenger VA, Fessler JA, Noll DC. Spatial domain
method for the design of RF pulses in multicoil parallel excitation. Magnetic
Resonance in Medicine. 2006;56(3):620-629. doi:10.1002/mrm.20978