Ruiyu Cao1, Shaohang Li1, Hongwei Li1, Ying-Hua Chu2, Aiqi Sun3, Ning Jin4, Xu Yan2, Yunzhu Wu2, Daniel Nico Splitthoff5, Hao Li1, and He Wang1
1Institute of Science and Technology for Brain-Inspired Intelligence, Fudan University, Shanghai, China, 2MR Research Collaboration Team, Siemens Healthineers Ltd., Shanghai, China, 3Oden Institute for Computational Engineering and Sciences, University of Texas at Austin, Austin, TX, United States, 4Siemens Medical Solutions USA, Inc., Chicago, IL, United States, 5Siemens Healthcare GmbH, Erlangen, Germany
Synopsis
Keywords: Velocity/Flow, Velocity & Flow
Motivation: 4D flow imaging in the liver suffers from long acquisition time and inefficient motion control.
Goal(s): Address the challenges of prolonged acquisition time and significant motion artifacts in hepatic 4D flow imaging.
Approach: We collected data from all respiratory states without a diaphragm navigator, and retrospectively reconstructed respiratory-resolved 5D flow using simultaneously recorded breathing signal.
Results: This scheme reduced the acquisition time from over 10 minutes to 6-8 minutes, while maintaining consistent image quality with few motion artifacts. High quantitative correlation in hepatic hemodynamic results was found between the results from prospectively navigated data and retrospectively binned data.
Impact: The proposed hepatic 5D flow scheme achieved
high image quality and quantitative metrics, compared with those of diaphragm
navigated results. This scheme can effectively shorten the acquisition time to
6-8 minutes and mitigate motion artifacts.
Introduction
Hepatic hemodynamics serves as a
crucial diagnostic indicator for hepatic diseases [1]. Previous studies have
demonstrated a significant correlation between portal venous hemodynamics and
the liver cirrhosis stage [2,3]. While 4D flow MRI has proven its feasibility
for evaluating portal venous flow [4,5], its long data acquisition duration and
the need for respiratory motion control have limited the widespread clinical
applications. To address these issues, we collected data from all respiration states
without navigator control to reduce scan times to 6-8 minutes. For motion
control, respiratory-resolved 5D flow data were reconstructed by data binning using
simultaneously recorded breathing signals. Hepatic hemodynamics was evaluated for
different phases and compared with navigated 4D flow data.Materials and Methods
The workflow of the proposed scheme is
illustrated in Figure1.
Data
Acquisition
MR imaging was performed on a 3T
system (VIDA; Siemens Healthcare, Erlangen, Germany) equipped with 18-channel
body and spine coil array. Seven recruited healthy volunteers (4 males, 3 females,
age: 22-26 years old) underwent both a continuous acquisition for
retrospectively data binning and a prospectively navigated 4D flow acquisition.
The continuous acquisition used Cartesian
compressed sensing (CS)‐accelerated 4‐point velocity encoding 4D flow sequence,
with prospective peripheral pulse gating, and the data were collected without
navigator control. The breathing signal was recorded simultaneously using a
breathing belt. The other parameters include: orientation = transverse, TR/TE =
39.5/2.95ms, FOV = 430×272×96mm3 , resolution = 2.1×2.1×3mm3,
velocity encoding = 50 cm/s, CS-accelerated factor = 5.7, temporal resolution =
39.5ms, reconstructed 17 cardiac frames, scan time: 5:57min.
The prospectively navigated acquisition
shared identical parameters with the first one, but incorporated a diaphragm
navigator, accepting data only within 5 mm from the end of expiration (scan
time: 10:31min). Navigation efficiency ranged from 55%-80%.
The total acquisition time varies
depending on the volunteer's heart rate.
Image
Reconstruction
We extended the application of the
locally low rank and finite difference (LLR_TV) [6] algorithm with ESPIRIT [7]
to facilitate 4D reconstruction. Data acquired with navigator were
reconstructed using 4D LLR_TV as the hemodynamics quantification reference.
Data acquired during all breathing states were retrospectively separated into
three phases (end-expiration, mid-respiration, end-inspiration) according to
breathing signal. The data at the extreme end inspiration were discarded, and
the division ensured a relatively uniform and random distribution of the data
within each phase. Then the data from three phases (retrospectively gated) and ungated
data (without phase separation) were reconstructed by LLR_TV.
Data Analysis
All reconstructed data were analyzed using
GT flow software (GyroToolsLLC, version 3.2.17, Switzerland) for quantification
of hemodynamic parameters. Three cross-sections were manually placed
perpendicular to the portal vein (PV), right portal vein (RPV), and left portal
vein (LPV), and the regions of interest (ROI) within these vessels were marked.
Average velocity, peak velocity and net flow for these vessels were measured.
The hemodynamic correlation between reference
and reconstruction results of our scheme was evaluated using the Spearman
correlation coefficient. Bland‐Altman analysis was also conducted for the corresponding
data.Results
Figure 2 shows the five reconstructed magnitude
images from one volunteer. The image reconstructed with navigated data exhibits
minimal motion artifacts (Figure 2a), while the image of end-expiration demonstrates
comparable image quality (Figure 2c). Noticeable motion artifacts can be
observed in images reconstructed from data of the other two respiratory phases
and ungated data (Figure 2bde).
For hemodynamic quantification,
although the images from ungated reconstruction were evidently affected by
motion artifacts, all the metrics showed high correlation when compared to the reference
(first column of Figure 3). For retrospective gated reconstruction, the
end-expiration phase showed the highest correlation with the reference (second
column of Figure 3), while the other two phases showed lower correlation. A
greater consistency between end-expiration phase and reference can also be
observed in Bland-Altman plots as shown in Figure 4.
Figure 5 demonstrates an example of post-processing
results.Discussion
The extracted end-expiration phase demonstrates
promising image quality and quantification results and could be optimal for
retrospective gated PC flow acquisitions. Motion correction will be necessary
to fully utilize the data from mid-respiration and end-inspiration phases,
which suffer from more motion artifacts and lower quantitative consistency. Despite
the presence of motion artifacts, ungated flow results without respiratory phase
separation also exhibit high correlation with the reference, which may due to
the contribution from more sufficient data and effective removal of artifacts
by the LLR_TV algorithm.Conclusion
The proposed hepatic 5D flow scheme achieved consistently
high image quality and quantitative metrics, compared with those of diaphragm
navigated results, and effectively shorten acquisition time. Acknowledgements
No acknowledgement found.References
[1] Taourel P, Blanc P, Dauzat M, et
al. Doppler study of mesenteric, hepatic, and portal circulation in alcoholic
cirrhosis: relationship between quantitative Doppler measurements and the
severity of portal hypertension and hepatic failure. Hepatology 1998;28(4):932–936.
3. Zekanovic D, Ljubicic N, Boban M, et al. Doppler ultrasound of hepatic and
system hemodynamics in patients with alcoholic liver cirrhosis. Dig Dis Sci
2010;55(2):458–466.
[2] Hyodo R, Takehara Y, Mizuno T, et
al. Assessing the complicated venous hemodynamics and therapeutic outcomes of
Budd-Chiari syndrome with respiratory-gated 4D flow MR imaging during the
expiratory and inspiratory phases[J]. Magnetic Resonance in Medical Sciences,
2023, 22(1): 1-6.
[3] Hyodo R, Takehara Y, Mizuno T, et
al. Portal vein stenosis following liver transplantation hemodynamically
assessed with 4D-flow MRI before and after portal vein stenting[J]. Magnetic
Resonance in Medical Sciences, 2021, 20(3): 231-235.
[4] Oechtering T H, Roberts G S,
Panagiotopoulos N, et al. Clinical applications of 4D flow MRI in the portal
venous system[J]. Magnetic Resonance in Medical Sciences, 2022, 21(2): 340-353.
[5]Bane O, Stocker D, Kennedy P, et
al. 4D flow MRI in abdominal vessels: prospective comparison of kt accelerated
free breathing acquisition to standard respiratory navigator gated
acquisition[J]. Scientific reports, 2022, 12(1): 19886.
[6]Otazo R, Candes E, Sodickson D K.
Low‐rank plus sparse matrix decomposition for accelerated dynamic MRI with
separation of background and dynamic components[J]. Magnetic resonance in
medicine, 2015, 73(3): 1125-1136.
[7]Uecker M, Lai P, Murphy M J, et al.
ESPIRiT—an eigenvalue approach to autocalibrating parallel MRI: where SENSE
meets GRAPPA[J]. Magnetic resonance in medicine, 2014, 71(3): 990-1001.