Rui Wang1, Qifan Ma1, Mengxiao Liu2, Yang Song3, Robert Grimm4, Voskrebenzev Andreas5, Vogel-Claussen Jens5, Ying Yuan1, and Xiaofeng Tao1
1Department of Radiology, Shanghai Ninth People’s Hospital, Shanghai JiaoTong University School of Medicine., Shang hai, China, 2MR scientific Marketing,Diagnostic Imaging, Siemens Healthineers Ltd, Shanghai, China, Shang hai, China, 3MR Research Collaboration Team, Siemens Healthineers Ltd. Shanghai, China., Shang hai, China, 4MR Application PredevelopmentResearch Collaboration CLS BODY, Siemens Healthineers AG, Erlangen, Germany, Erlangen, Germany, 5Institute for Diagnostic and Interventional Radiology, Hannover Medical School, OE 8220, Carl-Neuberg-Str.1, 30625 Hannover, Germany., Hannover, Germany
Synopsis
Keywords: Lung, Lung, PREFUL MRI
Motivation: The motivation for this study is to develop a non-invasive imaging technique for assessing pulmonary function without the need for ionizing radiation or potentially harmful contrast agents.
Goal(s): To evaluate the impact of different magnetic field strengths (0.55T and 1.5T) on functional parameters derived from the PREFUL MRI technique.
Approach: The study analyzed the data with MRLung software and conducting statistical comparisons to assess the differences in functional parameters, SNR, and CNR between the two field strengths.
Results: VDP, mean perfusion, and mean ventilation exhibited significant changes at the lower field strength, with higher SNR at 0.55T compared to 1.5T.
Impact: We
demonstrate the potential of using a lower field strength (0.55T) MRI system
for assessing pulmonary function, offering insights into the impact of field
strength on functional parameters and quality, which may have
implications for improving non-invasive pulmonary imaging methods.
Introduction
Early
diagnosis and monitoring of treatment responses in pulmonary diseases,
including Chronic Obstructive Pulmonary Disease (COPD), thromboembolic
pulmonary hypertension (CTEPH), and cystic fibrosis (CF), are vital for patient
management. Current pulmonary functional imaging methods often involve ionizing
radiation (e.g., X-ray, CT) and the use of contrast agents, which may have
associated risks1. A potential alternative is the
recently introduced the phase-resolved
functional lung (PREFUL) MRI technique2,
which is
a non-contrast, free-breathing, and dynamic imaging approach to assess lung
function. This study aims to evaluate the functional parameters, SNR and CNR
derived from PREFUL at 0.55T compared to 1.5T MRI.Methods and Materials
A
group of 18 healthy volunteers participated in the study and underwent both
0.55T(MAGNETOM Free.Star, Siemens Healthineers) and 1.5T(MAGNETOM Aera, Siemens
Healthineers) MRI scans within a two-week period.
MRI
data acquisition was performed using the same manufacturer and included routine
structural imaging and quantitative assessment of ventilation and perfusion.
The parameters were as below: PD fBLADE with trigger: TR:6753.11ms, TE:35ms,
FOV:380*380, Voxel size:1.2*1.2*6mm; PREFUL (0.55T based on trueFISP) :TR:276.9ms,
TE:1.6ms, FOV:450*450mm, Voxel size:1.75*1.75*15mm³, Time
resolution:300ms, Scan time: 1min09s; PREFUL (1.5T based on 2D FLASH):TR:2.9ms, TE:1.35ms,
FOV:500* 500mm, Voxel size:2*2*15mm³, Time Resolution:300ms,
Scan time:1min01s.
Image
processing utilized the MRLung research software(Version
2.2.0, Siemens Healthineers, Erlangen,Germany) to
analyze and generate functional parametric maps, including ventilation map((VR),
perfusion map(Q), perfusion defect percentage maps (QDP maps), ventilation defect
percentage maps (VDP maps) and ventilation/perfusion match maps (VQM maps).
The software employed motion correction, lung segmentation, filtering, and
phase-resolved analysis.Signal-to-noise ratio (SNR) and contrast-to-noise ratio
(CNR) were calculated for lung parenchyma and pulmonary artery.
Graphpad
Prism 8.0 software (GraphPad Software
Inc., San Diego, CA, USA) was used for statistical analysis. The differences of
quantitative functional parameters including Q, VR, QDP, VDP and VQM were
analyzed with non-parametric Wilcoxon signed-rank test Shapiro-Wilk test
revealed no normal distribution, while SNR and CNR were analyzed with paired t
test due to in accordance with normal distribution.
Additionally,
skewness and kurtosis of the Q and VR values were calculated with normalization
by dividing each value by the 95th percentile of the values. Bland-Altman plots
were created to evaluate the differences of all parameters between two series. Differences
were considered statistically significant at P<0.05.Results
Functional
parameters, including perfusion defect percentage (QDP), ventilation defect
percentage (VDP), and ventilation/perfusion match (VQM), showed no significant
differences between 0.55T and 1.5T MRI(Table 1). However, VDP, mean perfusion,
and mean ventilation exhibited significant changes at 0.55T compared to 1.5T
MRI.
SNR
was significantly higher at 0.55T for both expiration and inspiration states,
while CNR showed no significant differences between the field strengths. (Table
2) Histogram analysis indicated reduced skewness and kurtosis at 0.55T compared
to 1.5T, suggesting a change in the distribution of perfusion and ventilation
values.
The
alterations in ventilation values were more pronounced than those in perfusion
across different field strengths, explaining the significant changes in VDP and
VQM(Fig.2). The mean differences of
functional parameters between 0.55T and 1.5T were as follows: Q (3.03%, -3.88
to 9.94), VR (-4.1%, -9.81 to 1.61), QDP (-3.24%, -20.77 to 14.29),
VDP (13.18%, -1.24 to 27.), and VQM (-9.65%, -27.76 to 8.46).Discussion
Alterations
in perfusion and ventilation parameters are significant, with a 58% increase
in mean perfusion and a 33% decrease in mean ventilation at 0.55T MRI compared
to 1.5T. These findings are consistent with previous research on lower field
strengths. The unique characteristics of 0.55T MRI, such as reduced
susceptibility effects and slower spin-spin interactions, may contribute to the
observed differences in functional parameters. The relationship between
relaxation rates and field strength, as well as the distinct distributions of
T1 and T2* values in lung tissue, could explain the altered perfusion and
ventilation parameters. Also the utilized, fundamentally different pulse sequences may have a
significant effect on the derived parameters. We also noted higher SNR at 0.55T, which may be
attributed to lower field inhomogeneities and susceptibility effects.Conclusion
Perfusion and ventilation
parameters, as well as SNR, are significantly different when comparing 0.55T and 1.5T MRI systems using the
PREFUL method. These changes are likely influenced by the unique
characteristics associated with lower field strength and the fact that a different pulse sequence type was used at 0.55T. The study highlights the importance of further optimizing the
acquisition protocols and
establishing a normative range of parameter values at 0.55T to warrant comparability
of PREFUL parameters with higher field strengths.Acknowledgements
NoneReferences
1. Pöhler GH, Klimes F, Voskrebenzev A, Behrendt L, Czerner C,
Gutberlet M et al. Chronic
Thromboembolic Pulmonary Hypertension Perioperative Monitoring Using
Phase-Resolved Functional Lung (PREFUL)-MRI. Journal of magnetic resonance imaging : JMRI 2020; 52(2): 610-619.
2. Voskrebenzev A, Gutberlet M, Klimeš F, Kaireit TF, Schönfeld C,
Rotärmel A et al. Feasibility of
quantitative regional ventilation and perfusion mapping with phase-resolved
functional lung (PREFUL) MRI in healthy volunteers and COPD, CTEPH, and CF
patients. Magnetic resonance in medicine
2018; 79(4): 2306-2314.