Wenyan Zhang1,2,3, Quan Tao2,3,4, Zelong Chen5, Kan Deng6, Zhigang Wu7, Yizhe Zhang1,2,3, Qianqian Zhang1,2,3, Yikai Xu5, and Yanqiu Feng1,2,3,4
1School of Biomedical Engineering, Southern Medical University, Guangzhou, China, 2Guangdong Provincial Key Laboratory of Medical Image Processing, Southern Medical University, Guangzhou, China, 3Guangdong Provincial Engineering Laboratory for Medical Imaging and Diagnostic Technology, Southern Medical University, Guangzhou, China, 4Department of Rehabilitation Medicine, Zhujiang Hospital, Southern Medical University, Guangzhou, China, 5Medical Imaging Center, Nanfang Hospital, Southern Medical University, Guangzhou, China, 6Philips Healthcare, Guangzhou, China, 7Philips Healthcare, Shenzhen, China
Synopsis
Keywords: Quantitative Imaging, Quantitative Imaging
Motivation: Renal T1 mapping has been explored as a tool to visualize localized and diffuse renal diseases. 2D-MOLLI imaging have limited coverage, and 3D imaging are currently focused on cardiac imaging and with long acquisition time.
Goal(s): In this study, we aimed to propose a fast 3D T1 mapping sequence for kidney.
Approach: A saturate-recovery(SR) pulse, followed by 3D-GRASE readout, is used for T1 mapping. We also verified its effectiveness in T1 mapping for phantom and normal volunteers.
Results: The proposed sequence allows for 3D renal T1 imaging within two minutes and measured T1 value in phantom and volunteers were consensus with 2D-MOLLI.
Impact: This
SR-GRASE sequence may provide a 3D fast clinical imaging technique for kidney.
Introduction
Renal
T1 value is an important indicator for detection of renal fibrosis, acute
kidney injury, and chronic kidney disease1. The recent consensus2
recommends 2D-MOLLI as renal T1 standard measurement, but it is single slice imaging
and with low acquisition efficiency. Moreover, it is often affected by magnetic
susceptibility artifacts in renal imaging. Giovanna et al3 proposed
3D SASHA sequence, which is difficult to satisfy the clinical needs due to long
scanning time. This work aimed to develop a fast renal 3D T1 mapping technique,
and investigated its accuracy of T1 quantification.Method
Sequence
design: The proposed sequence (Fig. 1) consists of a non-selective SR pulse
followed by GRASE readout sampling the SR-curve after saturation-time (TS). An B1-insensitive
SR pulse utilizes a WET (water suppression enhanced T1 effects prepulse)
composite saturation pulse4, which applied a four-pulse saturation
train. Furthermore, a SPIR pulse is used for fat suppression.
Ex-vivo
experiment: The phantom contain mixture of agarose and (0.180,0.342,0.531,0.725
mM) NiCl25. All MRI data were scanned with a clinical
3.0T scanner (Ingenia, Philips Healthcare, Best, The Netherlands) and equipped
with 32 channels torso coil.
In-vivo
experiment: This study was approved by local Ethics Committee and informed
consent was obtained from all volunteers. Three recruited volunteers (n=3) underwent
MRI imaging using 2D-MOLLI and 3D-SR-GRASE sequence. Based on phantom
experiments, the imaging parameters were set. The parameters of 3D-SR-GRASE were
followed: nine sets of TS, ranging from 54ms to 3000ms6. TSE profile
order: low-high, TSE Turbo direction: Z phase encoding, EPI direction: Y phase
encoding, scan duration:94s. Scan parameters refer to Fig.2 for details.
Data
analysis: All MRI data were processed by MATLAB R2022b (The MathWorks, Natick, MA),
MR images were de-noised by Non-Local Means(NLM) and registered by principal
component analysis(PCA) using Elastix. Three-parametric exponential decay model
was used to fit the T1 value. Renal cortex and medulla ROIs were outlined on
the middle slice of the left kidney and right kidney.Results
Ex-vivo:
Figure 3 shows the T1 maps acquired by-2D MOLLI and 3D-SR-GRASE, and their
quantification of T1 value. Compared with 2D-MOLLI, the 3D-SR-GRASE showed
relative errors of -4.86%(vial 1)、9.33%(vial 2)、2.26%(vial
3)、-13.29%(vial
4) in T1 measurements.
In-vivo:
Figure 4 shows an example of the renal T1 maps generated by 2D MOLLI and 3D-SR-GRASE,
respectively. The T1 values of cortex and medulla measured using 2D MOLLI were:
1419.5±18.37ms
(left cortex), 1779.16±68.51ms (left medulla),1420.67±16.23ms
(right cortex),1802.5±109.79ms (right medulla); the T1 values of human renal
cortex and medulla measured using 3D-SR-GRASE were: 1535.33±42.19ms
(left cortex), 1903±66.95ms (left medulla),1486.33±57.4ms
(right cortex), 1891.67±58.82ms (right medulla). T1 values estimated by
2D-MOLLI and 3D-SR-GRASE showed strong correlation.Discussion
This
work showed that the proposed sequence is an accurate method for T1 measurement,
and with great consistent with 2D-MOLLI sequence. The phantom experiments
showed a relative 10% deviation and renal T1 values showed a great consistent
with 2D-MOLLI. Due to use of SR pulses, no additional waiting time is required,
and the acquisition efficiency is improved. However, 3D-SR-GRASE has a lower
signal to noise (SNR) ratio than 2D-MOLLI.
There
is an underestimation of T1 in the 2D-MOLLI sequence, which may lead to
inaccurate renal T1. Therefore, a sequence with higher precision needs to be
used as a standard for further exploration. In addition, images fitting in this
paper adopts three-parameter model, but the recovery process of magnetization may
need to take into account the effects of T2 and T2*, which is worthy of further
analysis.Conclusion
The
T1 values of the renal cortex and medulla obtained through the 3D-SR-GRASE are consistent
with 2D-MOLLI, and imaging can be achieved within two minutes. This preliminary
study suggested 3D-SR-GRASE has potential to become a clinically acceptable
renal 3D T1 mapping technique.Acknowledgements
This
study was supported by National Natural Science Foundation of China (U21A6005),
Key-Area Research and Development Program of Guangdong Province
(2018B030340001, 2018B030333001).References
1. Wolf M,
de Boer A, Sharma K, et al. Magnetic resonance imaging T1-and T2-mapping to
assess renal structure and function: a systematic review and statement
paper[J]. Nephrology Dialysis Transplantation, 2018, 33(suppl_2): ii41-ii50.
2. Dekkers I A, de Boer A, Sharma K, et al.
Consensus-based technical recommendations for clinical translation of renal T1
and T2 mapping MRI[J]. Magnetic Resonance Materials in Physics, Biology and
Medicine, 2020, 33: 163-176.
3. Nordio G, Bustin A, Henningsson M, et al. 3D
SASHA myocardial T1 mapping with high accuracy and improved precision[J].
Magnetic Resonance Materials in Physics, Biology and Medicine, 2019, 32:
281-289.
4. Morita K, Oda S, Utsunomiya D, et al.
Saturation recovery myocardial T1 mapping with a composite radiofrequency pulse
on a 3T MR imaging system[J]. Magnetic Resonance in Medical Sciences, 2018,
17(1): 35-41.
5.
Captur G,
Gatehouse P, Keenan K E, et al. A medical device-grade T1 and ECV phantom for
global T1 mapping quality assurance—the T 1 Mapping and ECV Standardization in
cardiovascular magnetic resonance (T1MES) program[J]. Journal of cardiovascular
magnetic resonance, 2016, 18: 1-20.
6. Nordio G, Bustin A, Odille F, et al. Faster 3D
saturation-recovery based myocardial T1 mapping using a reduced number of
saturation points and denoising[J]. Plos one, 2020, 15(4): e0221071.