Andrew Dwyer1,2, Angela Walls1, Kirsten Gormly2, Mitchell Raeside2, and Stephanie Withey3
1Clinical and Research Imaging Centre, South Australian Health and Medical Research Institute, Adelaide, Australia, 2Jones Radiology, Adelaide, Australia, 3Siemens Healthineers Pty Ltd, Adelaide, Australia
Synopsis
Keywords: Prostate, Prostate, Diffusion
Motivation: Diffusion-weighted imaging is critical to the diagnosis of prostate cancer but can be limited by noise. A commercially available 200mT/m gradient system may allow revisiting of higher b-value DWI.
Goal(s): To optimise a feasible high b-value prostate MR protocol for clinical use.
Approach: Phantom, simulation and in vivo iterative optimisation were evaluated against technical metrics and novel discrete choice experimental design for radiologist preference.
Results: Radiologist DWI preference is weighted by signal-to-noise. Emerging high gradient performance and deep learning reconstruction can reduced decline in SNR potentiating a signal-optimised high b-value protocol for clinical prostate cancer imaging within feasible scan times.
Impact: Signal-optimised high b-value DWI for prostate cancer using a 200mT/m
gradient 3T MRI system may be clinically feasible and supports a scalable
trial of its diagnostic impact.
Introduction
Diffusion-weighted MRI is critical to the diagnosis
of prostate cancer with hyperintensity at high b values and hypointensity on apparent diffusion coefficient (ADC) images being
a validated imaging marker for clinically significant prostate
cancer. A b-value of 1600s/mm2 has previously been defined as optimal at
3T with endorectal coil, a minimum of 1400s/mm2 recommended in PI-RADSv2.1 and values up to 2000 or higher considered advantageous1. However optimising pelvic DWI contrast needs to balance signal and artefacts with typical b-values in clinical practice ranging from 800-1200s/mm2.
Emerging MRI systems with high performance gradient systems for body imaging
may allow revisiting of higher b-value diffusion particularly in conjunction with developments in deep learning, complex
averaging and Eddy current correction. This paper describes the
initial experiences and optimisation of high b-value prostate MRI using a clinically available system.Methods
Healthy volunteer cohort (no history of prostate
disease, normal PSA) DWI images were obtained on a 3T MAGNETOM Cima.X (Siemens Healthineers, Erlangen,
Germany) capable of 200mT/m gradients at slew rate of 200T/m/s. Slice thicknesses of 2mm, 3mm and 5mm were used with nine acquired b-values in
250s/mm2 increments from 1000s/mm2 to 3000s/mm2
and calculated b-values to 4000s/mm2, vendor selective
field-of-view excitation, complex averaging and deep learning
reconstruction. Prostate signal-to-noise (SNR) was measured as the absolute
signal difference between peripheral zone and obturator internus muscle divided
by the standard deviation of obturator internus, with muscle chosen as the
reference given the impact on background tissue suppression. Geometric
distortion was calculated from the spatial overlap of manually segmented DWI
images. Image analysis was performed in Osirix (Pixmeo, Switzerland) and calculations in Matlab (Mathworks, CA). 156 randomly assigned image pairs (DWI, ADC or both) at mid
prostate slice from differing protocols were generated and sampled for a
discrete choice experiment (axes resolution,
b-value, noise) where 3 experienced prostate radiologists chose between paired images in a method adapted from research on patient healthcare preferences2. Preference weight for a
choice among a sets of alternatives can be used to quantify the impact of underlying attribute levels. Using 5-point Likert scale, the same radiologists also graded
images individually for overall quality, noise,
artefacts, sharpness.Results
From phantom experiments, deep learning and complex averaging methods contributed 26% and 2% incremental gain in signal respectively at b2000. 200mT/m gradient performance allowed
reduction in minimum TE particularly at higher b-values compared to simulation from 45mT/m system (Figure 1). As expected, prostate SNR reduced at higher b-values despite background suppression. Images
up to b3000 with comparable SNR to standard b1250 were possible in clinically
achievable scan time but highest b-value benefited from increase in slice thickness (Figure 2). Representative images from decision tree and direction of radiologist choice is shown in Figure 3. Preference weight was balanced between attributes provided noise was graded not more than 4/5 (having little or no
impact) with trend towards slightly thicker slices (likely to achieve SNR), towards
higher b-values and towards higher SNR. Once noise increased, it was the dominant factor other than a case
with non-diagnostic artefact. It suggests an SNR floor
for image acceptance. Agreement between radiologists was fair
(Krippendorf's alpha 0.67). Calculated peripheral zone ADC value was strongly influenced by the upper b-value, almost 50% lower from b1000 to b3000. Homogeneity of
ADC images reflected in reducing standard deviation also increased at higher b-values. Geometric distortion was not significantly increased with DICE similarity
coefficients using b1000 as reference of 0.974 and 0.963 at b2000 and b3000
respectively (Figure 4).Discussion
The importance of SNR on radiologist preference
in pelvic DWI portends challenges in high b-value acquisitions which are
inherently noisier and may explain the common use of below recommended values
in clinical practice. Deep learning reconstruction methods and gradient performance to minimise TE can reduce decline in SNR. This
provides an opportunity to revisit the role of higher b-values in prostate
cancer diagnosis but compromises are still required in the acquisition. A framework such as ours with optimisation guided by SNR can support deployment of an advanced higher b-value diffusion protocol for clinical research. Dependence of calculated ADC on b-value will require adaptive
thresholds if used for quantitative imaging but the increased homogeneity in
ADC of normal prostate tissue may also improve hypointense lesion detection. Initial results from implementation in a clinical cohort will be
presented.Conclusion
Emerging high-performance diffusion gradients and
reconstruction methods for pelvic imaging may challenge existing norms and allow
new approaches to higher b-value imaging in clinical practice. Ongoing research
such as this is needed to assess the impact on diagnostic accuracy and clinical
outcomes.Acknowledgements
The
authors acknowledge the support of the National Imaging Facility, a National Collaborative
Research Infrastructure Strategy (NCRIS) capability of Australia.References
1. PI-RADS Steering Committee. American College of Radiology. https://www.acr.org/-/media/ACR/Files/RADS/Pi-RADS/PIRADS-v2-1.pdf
2. Hauber B et al. Statistical Methods for the Analysis of Discrete Choice Experiments: A Report of the ISPOR Conjoint Analysis Good Research Practices Task Force,Value in Health,19:4, 2016.