Ehsan Kazemivalipour1,2, Bastien Guerin1,2, and Lawrence L Wald1,2,3
1A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Harvard-MIT Division of Health Sciences Technology, Cambridge, MA, United States
Synopsis
Keywords: RF Arrays & Systems, RF Arrays & Systems, Transmit radiation, Birdcage coil, Faraday cage
Motivation: Eliminating the Faraday cage would lower installation costs by ~2X and facilitate deployment of MRI in diverse settings, but requires reducing the Tx system’s electromagnetic (EM) radiation.
Goal(s): We employ a parallel transmit (pTx) array and EM absorbers to reduce RF-radiation from a 3T MRI operating without a shielded room, ensuring operation within regulatory limits.
Approach: We model a 3T MRI with a CP body coil and design a 16-channel pTx array and EM-absorber to reduce E-field radiation. Performance is assessed with pTx pulse optimization and L-curve analysis.
Results: We demonstrate a 2270-fold reduction in radiation compared to the birdcage coil.
Impact: By successfully mitigating RF-radiation from MRI
systems operated without Faraday cages, this research advances cost-effective
MRI installations in diverse clinical/research environments, encouraging
further exploration of pTx technology for controlling electromagnetic fields
both inside/outside of the body.
Introduction
Eliminating the
traditional RF-shielded cabin (Faraday cage) in clinical MRI could reduce
installation costs and simplify placement in diverse settings. The Faraday cage
mitigates two key issues: the receive problem, where incoming electromagnetic
(EM) interference increases image noise and artifact levels1-7,
and the transmit problem, whereby EM radiation exiting the scanner’s body Tx
coil might interfere with hospital equipment8. To address the Tx problem, regulatory limits9,10
(IEC-60601-1-2 & CISPR-11) limit peak
E-field on a 10-m radius sphere (peak-E10m) to
below 1-mV/m during 3T MRI. Our previous modeling of EM radiation from conventional
field-strength, body-loaded superconducting scanner geometries without a
Faraday room demonstrated radiation levels significantly above this regulatory
threshold11. Without some replacement mitigation method, MRI
radiation limits would severely limit Tx operation. In
this study, we use EM simulations of far-field E-field levels (peak-E10m)
and patterns for a 3T scanner with a conventional CP birdcage body coil to
propose and examine an active and passive mitigation strategy employing
parallel transmit (pTx) array coil pulse optimization.Methods
Figure 1 illustrates the conventional MRI system simulated,
including superconducting magnet, cylindrical RF shield, and a high-pass CP body birdcage
coil. Without mitigation (by a Faraday shield or other method), this geometry
exceeds the regulatory limits by 4680-fold for a 300Vrms input. Figure 2
shows the two mitigation strategies tested: (1) a 25-channel loop pTx array
positioned around the patient's feet, and an EM absorber placed at the bore’s service-end
(Figure 2B) and (2) a 16-channel head/neck pTx array12
replacing the body birdcage coil together with the same EM absorber at the
service-end (Figure 2C). In the first mitigation strategy, the pTx loops were
strategically positioned away from the head and primarily utilized to minimize peak-E10m. The passive EM
absorber used high-dielectric materials optimized
for graded interface impedance matching, placed at the bore's service-end. All EM
simulations were conducted using ANSYS Electronics with the co-simulation
approach, which allowed efficient tuning, matching, and decoupling. All transmit
channels (birdcage coil and pTx arrays) were tuned and matched to achieve
|S|<-30dB. The pTx channels were ideally decoupled.
For all configurations, the radiated E-fields of
each Tx channel were exported on 1° isotropic grids on
the surface of a
10-m radius sphere. Additionally, B1+-map
of each channel was exported on 2-mm isotropic grids in the central head axial
slice. Q-matrices13 corresponding to the magnitude of the radiated E-field were computed
and compressed using the VOP algorithm14, with a 5% overestimation factor. We
computed and reported peak-E10m
for
a 1Vrms continuous-wave input waveform and for RF-shimming
pulses designed with a target flip-angle of 10° across the imaging slice using
a 5-lobe slice-selective sinc pulse, with 2-ms duration and 100% duty-cycle. To quantify the radiation
mitigation performance of the 25-channel pTx array configuration, we generated an L-curve of the tradeoff
between loops pTx power consumption and peak-E10m. In the case of the 16-channel head/neck pTx
array, we designed the RF pulses using a least-squares optimization15-17 with a target phase profile of the birdcage mode
of the array while constraining the peak-E10m. An L-curve shows the performance tradeoffs
between excitation uniformity and peak-E10m.Results
Figure 3 shows the 10-m E-patterns for the three configurations
of Figure 2. The conventional system exhibited a peak-E10m
of 24dBµV/m higher than the regulation (60dBµV/m at 3T). The utilization
of the 25-channel pTx system in conjunction with the EM absorber reduced this
by 16-fold and thus below the regulation limit (for this RF drive voltage).
Additionally, using the 16-channel head/neck pTx array with the EM absorber
yielded a remarkable 2271-fold reduction in the peak-E10m,
keeping it 43dBµV/m below the regulation.
Figure
4 shows the L-curve tradeoff between peak-E10m and maximum pTx voltage for the 25-channel pTx
loop approach, showing that pTx effectively attenuates EM radiation by up to
16-fold without requiring high voltage input. Figure 5 shows the L-curve
comparison between peak-E10m and flip-angle uniformity for the 16ch RF-shimming
approach. At a constant flip-angle RMSE level (9.9%), the 16-channel pTx system
reduces EM radiation by 67.5dBµV/m
(2383-fold) compared to the conventional system.Conclusion and Discussion
Regulations dictate that radiated peak-E10m
should not exceed 60dBµV/m for 3T systems, significantly below that found when
conventional systems are operated without a Faraday room. Here, we proposed and
simulated the ability of pTx methods to reduce radiated peak-E10m
levels. The pTx pulses achieved excellent flip-angle excitations within the
body while reducing far-field RF radiation by up to 67dBµV/m compared to the CP-driven
birdcage coil radiation, ensuring that peak-E10m
remained
below the regulatory threshold (for 10° flip-angle pulses).Acknowledgements
The
authors extend their gratitude to Markus W. May and Boris Keil for the
16-channel head/neck pTx array model, originally designed for 7T imaging, used
in this study.References
[1] Harberts DW, Helvoort MV. Sensitivity of
a 1.5-T MRI system for electromagnetic fields. 2014 International Symposium on
Electromagnetic Compatibility 2014:856-859.
[2] Rearick T, Charvat
GL, Rosen MS, et al. Noise suppression methods and apparatus (US Patent
No. 9,797,971 B2). U.S. Patent and
Trademark Office. 2017.
[3] Huang X, Dong H, Qiu
Y, et al. Adaptive suppression of power line interference in ultra-low field
magnetic resonance imaging in an unshielded environment. J Magn Reson. 2018;286:52-59.
[4] O'Reilly T, Teeuwisse WM, de Gans D, et al. In
vivo 3D brain and extremity MRI at 50 mT using a permanent magnet Halbach
array. Magn Reson Med. 2021;85(1):495-505.
[5] Liu Y, Leong ATL,
Zhao Y, et al. A low-cost and shielding-free ultra-low-field brain MRI scanner.
Nat Commun. 2021;12(1):7238.
[6] Srinivas SA, Cauley SF, Stockmann JP, et al. External
Dynamic InTerference Estimation and Removal (EDITER) for low field MRI. Magn Reson Med. 2022;87(2):614-628.
[7] Zhao Y, Xiao L, Liu Y, et al. Predict
and Eliminate EMI Signals for RF Shielding-Free MRI via Simultaneous Sensing
and Deep Learning. 2022 Asia-Pacific International Symposium on Electromagnetic
Compatibility (APEMC) 2022:213-215.
[8] Harberts DW, Helvoort
MV. Shielding requirements of a 3T MRI examination room to limit radiated
emission. 2013 International Symposium on Electromagnetic Compatibility
2013:1053-1057.
[9] IEC 60601-1-2,
Medical electrical equipment - Part 1-2: General requirements for basic safety
and essential performance – Collateral standard: Electromagnetic disturbances -
Requirements and tests. 2014.
[10] CISPR 11,
Industrial, scientific and medical equipment – Radio-frequency disturbance
characteristics - Limits and methods of measurement. 2016.
[11] Kazemivalipour E,
Guerin B, Wald LL. Simulated radiation patterns of MRI without a shielded room
from 0.5 to 7 Tesla. Proc Intl Soc Mag Reson Med 31 2023:0598.
[12] May MW, Hansen SJD,
Mahmutovic M, et al. A patient-friendly 16-channel transmit/64-channel receive
coil array for combined head-neck MRI at 7 Tesla. Magn Reson Med.
2022;88(3):1419-1433.
[13] Guerin B, Gebhardt
M, Cauley S, et al. Local specific absorption rate (SAR), global SAR,
transmitter power, and excitation accuracy trade-offs in low flip-angle
parallel transmit pulse design. Magn Reson Med. 2014;71(4):1446-1457.
[14] Eichfelder G,
Gebhardt M. Local specific absorption rate control for parallel transmission by
virtual observation points. Magn Reson Med. 2011;66(5):1468-1476.
[15] Zhu Y. Parallel
excitation with an array of transmit coils. Magn Reson Med. 2004;51(4):775-784.
[16] Setsompop K, Wald
LL, Alagappan V, et al. Parallel RF transmission with eight channels at 3
Tesla. Magn Reson Med. 2006;56(5):1163-1171.
[17] Cloos MA, Luong M, Ferrand G, et al. Local SAR
reduction in parallel excitation based on channel-dependent Tikhonov
parameters. J Magn Reson Imaging. 2010;32(5):1209-1216.