Jan Weis1, Magor Babos2, Sergio Estrada3, and Ram Kumar Selvaraju3
1Department of Medical Physics, Uppsala University Hospital, Uppsala, Sweden, 2Mediso Medical Imaging Systems, Budapest, Hungary, 3Department of Medicinal Chemistry, Preclinical PET-MRI Platform, Uppsala, Sweden
Synopsis
Keywords: Biology, Models, Methods, Data Acquisition, EPSI, flyback readout, ghost artifacts, micro-imaging, water-fat imaging
Motivation: To assess echo-planar spectroscopic imaging (EPSI) with flyback readout gradients in preclinical MR system.
Goal(s): To demonstrate spectral ghost artifacts produced by two, three, and four interleaved gradient echo trains and to measure water, fat, and water-fat shift artifacts free images.
Approach: Flyback EPSI with two, three, and four interleaved gradient echo trains.
Results: The proposed approach with four interleaved gradient echo trains and with four echoes in each train enables high spectral bandwidth in combination with narrow receiver bandwidth and a very good water/fat signals separation. It improves SNR without the undesired consequence of water-fat shift artifacts.
Impact: Echo-planar spectroscopic imaging with
flyback readout enables measurement of water, fat, and water-fat shift artifact-free images. Four interleaved echo trains with
four echoes in each train provide high spectral, and narrow receiver
bandwidth, and a very good water-fat separation.
Introduction
Water
and fat unsuppressed echo-planar spectroscopic imaging (EPSI) with high spatial
resolution was previously used on clinical scanners for water and fat imaging
and spectroscopy1-3. The weakness of such an approach is the presence of Nyquist
ghost artifacts caused by inconsistency between odd and even echoes. An alternative
to the gradient echo train created by trapezoidal positive and negative gradient
pulses is the “flyback” gradient echo train in which only one polarity of the
readout gradient is applied4,5. Gradient waveform is asymmetric and
combines strong rewind gradients with lower ones for the readout. It simplifies
data processing. However, a high-power gradient system is required to achieve
sufficient spectral bandwidth (sBW). In this study, the flyback EPSI sequence
was implemented in a preclinical MR scanner. The main aim of this work is to
demonstrate spectral ghost artifacts and the potential of the interleaved flyback EPSI
in preclinical MR systems. Methods
Experiments were performed on a nanoScan® PET/MRI 3T preclinical scanner
(Mediso Medical Imaging Systems,
Budapest, Hungary). The scanner was equipped with gradients with a maximum amplitude of 550
mT/m and a maximum slew rate of 4500 T/m/s. Flyback EPSI sequence begins with
slice selection followed by phase encoding. Gradient echoes were acquired with
two, three, or four interleaved gradient echo trains with linear ramps
of 0.15 ms duration. The resultant spacing
between echoes was 0.8 ms (sBW 9.74 ppm). Ghost peaks of
water and fat spectral lines were visualized with a phantom (Fig. 1) containing vegetable
oil and water solution of MnCl2 (~0.23 mM)6. Phantom’s T2
relaxation times mimics subcutaneous fat (T2 ~70 ms) and muscle
water (T2 ~30 ms). All experiments were performed with maximum
available FOV 80x80 mm, 192 phase-encoding steps, 256 points/echo (resolution
in-plane 0.31x0.42 mm2) and 2 averages. Phantom experiments were
performed with slice thickness 2 mm, TR/TE1=200/3 ms, and flip angle
30o. Receiver bandwidth (rBW) was either 434028 Hz (readout
gradient gread=127 mT/m) or 114280 Hz (gread=33.7 mT/m).
Animal experiments were performed using four interleaved gradient echo trains with
four echoes in each train. 27 coronal slices (thickness 1.3 mm) were measured
with TR/TE1=500/3 ms, flip angle 60o, and with rBW 114890
Hz. The acquisition time was ~13 minutes. Data processing software was
developed in-house and has been described elsewhere7-9. Results and discussion
The described technique provides magnitude spectra, water, fat, and water-fat
shift (WFS) artifact free images7-9. Water and vegetable oil spectra were computed
from the volume of interests 51 mm3 (Fig. 1). As expected, the interleave of
two, three, and four echo trains produce one, two, and three ghost spectral
lines, respectively (Fig. 2). Distances between ghost peaks and main water or methylene
(-CH2-)n lines are sBW/2, sBW/3, sBW/4 (or 2xsBW/4) for spectra
acquired with two, three, and four interleaved echo trains, respectively. Only
ghost peaks from the highest methylene (-CH2-)n line
can be recognized. Ghost peaks of other fat spectral lines are too small to be detected
in our case. Water and fat images were computed by integration of water and (-CH2-)n spectral lines. Therefore, suitable sBW has to be
chosen to avoid the superposition of water and fat ghost peaks with water and fat (-CH2-)n
lines. Figure 2b demonstrates such superpositions. The advantage of the EPSI
micro-imaging approach over the conventional imaging is the fact that WFS artifacts are
eliminated during data processing7. Since signal-to-noise ratio (SNR)
is directly proportional to WFS1/2 and WFS is indirectly proportional to
receiver bandwidth (WFS ~1/rBW), narrow rBW, i.e. larger WFS artifacts during acquisition can be chosen for SNR improvement without penalty of WFS
artifacts. Narrow rBW can be achieved by increasing the number of the echo trains. Four
interleaved echo trains enabled rBW 114890 Hz in our case. The resulting spectra
and images are shown in Figs. 3 and 4. Intensities of both water and fat ghost
peaks are less than truncation artifacts up to ~32 echoes (Fig. 3 a, b).
Truncation artifacts are small already for 64 echoes (Fig. 3c). Figure 4 shows
examples of the rat’s images. Separation of water and fat
intensities is very good although only 16 echoes were used for imaging. Conclusion
This work demonstrates the applicability of
echo-planar spectroscopic micro-imaging with flyback readout gradients in
preclinical MR systems with small magnets and high-power gradients. The proposed
approach provides higher sBW and very good water and fat signal separation
already for 16 gradient echoes while intensities of ghost peaks are negligible.
Narrow rBW was achieved by increasing the number of interleaved echo trains. It
improves SNR without the undesired consequence of increased WFS artifacts.Acknowledgements
No acknowledgement found.References
1. Du W, Du YP, Bick U, et al. Breast MR imaging
with high spectral and spatial resolutions: preliminary experience. Radiol. 2002;224:577-585.
2. Du W, Du YP, Fan X, et al. Reduction of
spectral ghost artifacts in high-resolution echo-planar spectroscopic imaging
of water and fat resonances. Magn Reson Med.
2003;49:1113-1120.
3. Medved M, Ivancevic
MK, Olopade OI, et al. Echo-planar spectroscopic imaging (EPSI) of
the water resonance structure in human breast using sensitivity encoding
(SENSE). Magn Reson Med. 2010;63:1557-1563.
4. Hennel F, Nedelec J-F.
Interleaved asymmetric echo-planar imaging. Magn Reson Med. 1995;34:520-524.
5. Cunningham CH,
Vigneron DB, Chen AP, et al. Design of flyback echo-planar readout
gradients for magnetic resonance spectroscopic imaging. Magn Reson Med. 2005;54:1286-1289.
6. Thangavel K, Saritas EÜ. Aqueous
paramagnetic solutions for MRI phantoms at 3 T: a detailed study on
relaxivities.Turk J Elec Eng & Comp Sci. 2017;25:2108-2121.
7. Weis J, Ericsson A, Hemmingsson A, Chemical shift artifact-free microscopy:
spectroscopic microimaging of the human skin. Magn Reson Med. 1999;41:904-908.
8. Weis J, Courivaud F, Schacht Hansen M, et al. Lipid content in the
musculature of the lower leg: evaluation with high-resolution spectroscopic
imaging. Magn Reson Med. 2005;54:152-158.
9. Weis J, Bruvold M, Ortiz-Nieto F, et al. High-resolution echo-planar spectroscopic imaging of the human calf.
PLOS ONE. 2014; 9:e87533.