Chie-Hee Cho1,2 and Lutz Lüdemann2,3
1Institute for diagnostic and interventional Radiology, University Clinic Jena, Jena, Germany, 2Department of Radiotherapy, Charite Universitätsmedizin Berlin, Berlin, Germany, 3Department of Radiotherapy, Medical Physics Section, University Hospital Essen, Essen, Germany
Synopsis
Keywords: Large Animals, Nonhuman Primates, Safety, Swine, thermoregulation, temperature curves
Motivation:
Radiofrequency used during MRI exams heat tissue. The temperature increase is unknown and the thermoregulatory responses still need to be analyzed.
Goal(s): To analyse the temperatures within the thresholds. Are any changes seen. Is the biological simulation replicable in living swine.
Approach: SAR levels in the 2.mode were planned in the swine as an animal model.
Results: The thermoregulatory model (temperature going into a plateau) is only one pattern next to 3 other patterns (linear, parabolic, and sinusoidal). So these will change the calculations of simulations for temperatures in humans receiving MRI exams.
Impact: Radiofrequency induced temperature
changes in MRI exams need to be reevaluated as new thermoregulatory patterns
are observed. Increased internal temperatures are subconsciously discerned
causing thermoregulation to avoid tissue damage.
Introduction
Introduction
Voxel
models of human torsos simulating the distribution of energy deposition of
radiofrequency (RF), which is applied in a clinical magnetic resonance (MR)
Scanner are used to predict the local specific absorption rate (SAR). These can
exceed the global SAR (SARwb) up to a factor of 10-20 resulting in hotspots.1,2
According to the IEC 60601 2-33 standard3 different levels of SAR
are allowed as long as body core temperatures are not exceeded, Tab 1.
Mode
SAR
(W/kg)
Max.
local tissue temperature
in °C
Max.
core body temperature
in °C
Max.
increase of core body temperature in °C
Normal
<2
39
39
0,5
1.Level
2-4
40
40
1
2.Level
>4
>40
>40
>1
Tab
1 IEC 60601 2-33 standard different levels of SAR. Normal mode is applied in
regular MRI exams, 1. Level Mode is applied when bigger volumes (i.e. Whole
body MRI, obese patients) are examined, 2. Level Mode is applied in research
terms, needing ethical approval
Following
questions arise: a) as long as body core temperatures are within the limits, do
the temperatures stay within the same limits; b) can changes be seen visually
without further instruments; and c) are the predictions based on the assumption
that all biological models react to temperature in the same way applicable.
Methods
26
swine were positioned in a whole-body resonator (16 ring bird cage type) of a
conventionale 3T MR system operating at 123 MHz in a quadrature mode via a 90°
hybrid to a 35 KW RF power amplifier, for more detailed information on the set
up see Brinker et al.4. A Swine data set was acquired via
computertomography (CT), which was used as a model exposed to RF. The
simulation showed two major hotspot locations (HS). The temperatures in the HS
were measured invasively with fiber optic probes and the core temperature was
measured in the rectum. The swine were anesthetized with propofol (2%) and
0.05mg/ml fentanyl, intubated for securing the airway and breathing freely.
They were assigned 5 energy groups: a) sham SARwb=0 W/kg; b) low energy mean
SARwb=2.7 W/kg, duration of exposure (EXPO) 56.3 min; c) moderate energy mean
SARwb=4.8 W/kg, mean EXPO 31.2 min; d) high energy mean SARwb=4.4 W/kg, mean
EXPO 60.9 min; and e) extreme energy SARwb 11.4W/kg EXPO 8 min. Temperatures
were measured continuously, frequency of heart beat (HR), and frequencies of
breathing (BR) were documented every minute.
Results
Results
The
temperatures of the HS probe 2 and the rectum prior, post and 20 min after RF
exposure were published in Cho et al.5. The core body temperatures
are within in the limits while HS temperatures exceed 42°C leading to
denaturation of protein.
A
B
a
Macroscopic
pathological Character
b
SARwb
3.7 W/kg
4.8 W/kg
c
Core body temperature
at 30 min
39.7°C
37.9°C
d
HS temperature
at 30 min
48°C
46.5°C
e
Energy group
high
moderate
Fig.
1 Column A shows the visual changes of the tissue in swine 23. Column B shows
the visual changes of the Tissue in swine 21.
As
an example swine 23 was exposed to 1.Level mode, core body temperature within the
limit, but macroscopically tissue damage can be seen (Fig 1). Although swine 21
exposed to 2.Level mode the core body temperature was within limits, the HS
temperature was lower than swine 23 exposed to 1.Level mode, but still tissue
damage can be seen (Fig 1).
An
answer to the different temperatures could be due to the cardiovascular
response pattern5 that can be grouped into 4 main patterns: linear,
plateau, parbolic, and sinusoidal. So the previous assumptions in living beings
basing on one pattern only (plateau), described in multiple experimental
setups, need new formulas for calculation and simulation of HS.Conclusion
Conclusions
The
limits in MR safety need to be re-evaluated. More data from patients undergoing
MR exams are needed, since humans have greater thermoregulatory response than
for example swine. The simulation of SAR needs to encompass different patternsAcknowledgements
I am deeply indebted to Eckart Stetter (Siemens) and Gerhard Brinker (Siemens) for organizing the financial, technical, and logistic support. I am grateful for my meanwhile both deceased mentors Peter Wust (Charité) who introduced me to the field of hyperthermia and Christian Grosse-Siestrup (Charité) who was responsible for the official authorisation to carry out the animal experiments, for supervising this project in the field of veterinarymedicine, and for the mental support in evaluating and publishing this work. I thank Jacek Nadobny (Charité) for his experience in the calculations for Hyperthermia and the four veterinarians Katja Reiter, Sarah Ribereau, Friedericke Kobelt, and Markus Piaskowski for their assistance with the animals and execution of the invesitgation. I also thank Gisela Stoltenburg for her counseling in muscle histopathology.References
1.
Nadobny et al., IEEE Transactions of biomedical engineering, Vol.54, No 10,
October 2007
2. Collins et al. Proc. Intl. Soc. Magn. Reson. Med 14 (2006)
3 International Elecrotechnical Commission, IEC 60601-2-33:2002E
4. Brinker et al., Proc. Intl. Soc. Magn. Reson. Med. 18(2010)
5. Cho et al, Bioelectromagnetics. 42(2021)