Nicolas Arango1, Jacob White1, and Elfar Adalsteinsson1,2
1Massachusetts institute of Technology, Cambridge, MA, United States, 2Institute for Medical Engineering and Science, Massachusetts Institute of Technology, Cambridge, MA, United States
Synopsis
Keywords: Shims, Shims
Motivation: Multicoil B0 shimming improves signal quality in many MRI applications but either requires pre-calibrated ridged shim coils or lengthy coil re-calibration scans.
This has precluded the use of flexible arrays and limited body applications.
Goal(s): Accelerate B0 shim coil re-calibration to enable body-conforming flexible shim coil arrays.
Approach: Field calibration is accelerated using physics-based reconstruction, interleaving few measured slices, and de-tangling contributions of simultaneously active shim coils.
Results: Measured data is used to demonstrate the effectiveness of shim field recovery from few-slice measurements, and simulation is used to demonstrate simultaneous multi shim coil calibration, resulting in a 6.7x total acceleration.
Impact: Physics-based
reconstruction of coil fields from few interleaved
slices,
and detangling
fields
from simultaneous
B0
shim
coil activation,
achieves
rapid calibration. Fast
subject-specific calibration will enable the use of body-conforming,
flexible
B0
shim arrays.
Introduction
Multi-coil B0
shim arrays improve MRI image quality in a variety of applications
[1].
Shim current optimization requires both subject-specific fields and
calibrated shim coil field patterns. In most brain applications coil field precalibration is a multi-hour 1-time subject-agnostic
procedure
that may only be used if the shim array is installed in a fixed
position relative to the scanner.
Recent work in body imaging [2 ,3], implant imaging [4] and flexible
an non-fixed shim coils systems [6] shows the benefits of
body-conforming multicoil shim arrays, but the needed recalibration
is impractically lengthy.
Below we show that field calibration can be accelerated 7 fold using
a combination of physics-based reconstruction, interleaving few
measured slices, and de-tangling contributions of simultaneously
active shim coils.Methods:
We solved a magnetostatic inverse problem using TV and L1
regularization to estimate shim coil fields everywhere in space with
few measurements as shown in Figure 1 [7]. A forward mangetostatic
calculation computes B0 directed magnetic fields in a body domain due
to currents on a surface outside of that domain. Naive inversion is
highly ill-posed while L1+TV regularization of the surface-current
equivalent dipole distribution produces spatially compact, current
distributions with extrapolation power.
Recovering B0 shim fields from few-slices of
calibration data:
We use the L1+TV regularized field
recovery with data from only six planes
peripheral to the volume; that is, two peripheral slices from each
an axial, sagittal, and
coronal coil B0
map.
With a magnetostatic forward model
from a surface dipole distribution A sampled to fields in slices,
Ai
and measured fields due to a unit current through slices bi
we solve for a surface dipole distribution x by solving the
optimization problem:
$$x_{opt} = argmin_x \sum_i\left|A_i x
– b_i\right|_2^2 + TV(x) + |x|_1$$
We solve this optimization problem
using ADMM [5]
and compute the coil field
anywhere in the body domain with
the product $$$Ax_{opt}$$$
We extend the L1+TV regularized field recovery to the simultaneous
multi coil case by slightly modifying the experiment and optimization
problem. We acquire fieldmap planes bi with random
currents W with current wij applied to the jth
active coil on the ith slice. Fields, xj are recovered by
solving:
$$X_{opt} = argmin_X \sum_i\left|\sum_j(w_i^jA_i x^j)
– b_n\right|_2^2 + TV(x) + |x|_1$$
Measurements:
Field recovery experiments were performed using 2D GRE (1.7 x 1.7 x 2
mm, 110x110x75 matrix,TA = 150 s), field maps of each B0 shim coil of
the 32 channel AC/DC multicoil shim array [1], figure 2a). B0 maps
were resampled into axial, sagital, and coronal slices. Simultaneous
multi shim-coil data were synthetically constructed by scaling and
summing resampled fieldmap slices.
We perform single coil recovery experiments with three central planes
and six peripheral planes as shown in figure 2c). Simultaneous Multi
Shim coil recovery experiments were performed with two, three, and
four simultaneous coils using 6, 9, and 12 planes respectively. In
each case the same six peripheral planes were used with additional
evenly spaced planes in each axis.Results and Discussion
Figure 2 shows the recovery of the worst-case coil of the AC/DC shim
array from measurements from six planes peripheral to a brain volume.
The recovered fieldmap is shown as sampled along an axial plane not
used in solving the inverse problem. The worst-case shim coil fields
were recovered over the brain domain with a relative error of 8%.
Over all 32 coils, mean relative error was 4%.
Robust recovery of a single coil requires six planes near the
periphery of the body volume. Only five sets non-identical slices are
sufficiently peripheral for recovery. Single coil recovery from six
planes achieves an acceleration of 1.6x over 32 coils. Due to
geometric symmetries, central slices alone are not sufficient to
accurately recover shim coil fields everywhere in space as shown in
figure 3. We make use of central slices in simultaneous multi shim
coil recovery even if insufficient to recover single coils.
Figure 4 shows an example of recovery of two coil B0
fields from six acquired planes with a relative error of 8% and 5%
for the two channels. Using the same five sets of peripheral coils we
achieve an acceleration of 3.3x. Higher simultaneous factors were
also simulated. 3-channel 9-plane acquisition achieves 5x speedup and
4-channel 12-plane acquisition achieves 6.7x speedup with comparable
error.Conclusion
Physics-based reconstruction
of coil fields from few interleaved
slices, and detangling
fields from
simultaneous B0
shim coil activation,
achieves
rapid calibration. Fast
subject-specific calibration will enable the use of body-conforming,
flexible
B0
shim arrays.Acknowledgements
No acknowledgement found.References
[1] Jason P Stockmann, Thomas Witzel, Boris Keil, Jonathan R
Polimeni, Azma Mareyam,Cristen LaPierre, Kawin Setsompop, and
Lawrence L Wald. A 32-channel combined rf andb0 shim array for 3t
brain imaging. Magnetic resonance in medicine, 75(1):441–451, 2016.
[2] Hsin-Jung Yang, Fardad Serry, Peng Hu, Zhaoyang Fan, Hyunsuk
Shim, AnthonyChristodoulou, Nan Wang, Alan Kwan, Yibin Xie, Yuheng
Huang, et al. Ultra-homogeneous b0 for high field body magnetic
resonance imaging with unified shim-rf coils. 2021.
[3] Lieke van den Wildenberg, Quincy van Houtum, Wybe JM van der
Kemp, Alex Bhogal,Paul Chang, Sahar Nassirpour, and Dennis WJ Klomp.
B0 shimming simulations of the liver using a local array of shim
coils in the presence of respiratory motion at 7 t. Tools for
quantitative MR imaging and
spectroscopy for the improvement of therapy evaluation in oncology,
page 37, 2020.
[4] Fardad Michael Serry, Junzhou Chen, Anthony G Christodoulou,
Yuheng Huang, FeiHan, Won Bae, Christine Chung, Richard Richard
Handlin, John Stager, MatthewMatthew Dausch, Yubin Cai, Yujie Shan,
Yucen Liu, Yibin Xie, Xiaoming Bi, RohanDharmakumar, Zhaoyang Fan,
Debiao Li, Hsin-Jung
Yang, and Hui Han. Improving mri near metal with local b0 shimming
using a uniied shim-rf coil (unic): First case study, hip prosthesis
in phantom. 2021.
[5] Stephen Boyd, Neal Parikh, and Eric Chu. Distributed optimization
and statistical learning via the alternating direction method of
multipliers. Now
Publishers
Inc, 2011
[6] Overson, Devon Karl, et al. "Flexible multi‐purpose
integrated RF/shim coil array for MRI and localized B 0 shimming."
Magnetic Resonance in Medicine, 2023
[7]
Nicolas Arango, Jacob White, and Elfar Adalsteinsson "A
Fast Magnetostaic Inverse Approach for Subject-Specific∆ B 0 Shim
Coil Calibration." ISMRM #1374,
2022