Seung-Kyun Lee1 and Yihe Hua1
1GE HealthCare Technology and Innovation Center, Niskayuna, NY, United States
Synopsis
Keywords: Gradients, High-Field MRI
Motivation: High-performance gradient coils can induce substantial eddy current in the magnet causing heating and artifacts. Motional eddy current at high B0 exacerbates the problem.
Goal(s): To investigate the feasibility of patterned, additional passive shielding to reduce both electromagnetically-indued and motional eddy currents.
Approach: A proof-of-concept experiment was conducted in flat geometry, where shielding efficiency at 3T was compared between a solid copper and a patterned copper plate that mimicked the eddy current image of a driving coil.
Results: The leakage field of patterned copper agreed with theory and was more localized in frequency and space. The shielding efficiency was improved at high frequencies.
Impact: We demonstrated that vibration-induced motional eddy current of a passive conductive shield can be dramatically altered by eddy current cut-outs. The results can be used to build a compact, high-field scanners with reduced magnet-gradient interaction.
Introduction
The
shielding efficiency of an actively shielded gradient coil1,2 is
limited by current discretization as well as design tradeoffs. An additional
passive shield by a continuous conductive layer can improve shielding3,4,
as long as Lorentz-force vibration and motional eddy current are well contained5,6.
Previous attempts to control the motional eddy current focused on structural measures4,7.
Here we present a novel method to reduce the motional eddy current by cutting a
pattern on the conductive shield according to the eddy current image of the
leakage field8. The idea is demonstrated in a flat-geometry testbed
experiment at 3T. Methods
Frequency
response function Figure
1 shows the experimental setup. Gradient leakage field was modeled by the
magnetic field generated by a 20 cm-diameter, 10-turn circular loop coil (“driving
coil”). The field was sensed by a 5 cm-diameter solenoidal pickup coil which
was mounted on a wooden stick that could be translated horizontally (= z
direction) above the driving coil. The experiment consisted of measuring the
frequency response function (FRF), defined as the complex ratio between the pickup coil
voltage and the driving coil current in the frequency domain, in the following
four configurations. (1) First, the FRF was measured at 19 z locations (z =
-9∆z to +9∆z, with ∆z = 2.54 cm) with no barrier between the two coils. (2) Second,
a 1.6 mm-thick solid copper plate backed by 9.6 mm-thick fiberglass, both measuring
55.9 x 55.9 cm2 in plane,
was inserted between the coils (Fig. 1B). (3) Third, the setup (with copper) was
moved inside the bore of a 3T whole-body MRI magnet. (4) Finally, the solid
copper plate was replaced by a copper plate with a spiral pattern (Fig. 1C)
machined according to the simulated eddy current image of the driving coil
(Fig. 1D). The spiral pattern was bridged by a copper strip to make a single closed
loop that allowed current to flow in a predefined way.
Impedance
The
complex impedance spectrum (5 < f < 50,000 Hz) of the spiral pattern was
measured in the magnet, with the copper strip electrically open. This data was
used to calculate theoretical leakage field of the patterned copper including
any Lorentz-force effects. Results
Figure
2 shows FRF at frequencies up to 3 kHz in the two configurations outside the
magnet. Near complete shielding of the driving field by solid copper is
evident. Figure 3(A) shows dramatic degradation of shielding performance of the
copper plate at 3T, caused by Lorentz-force vibration and motional eddy current6.
This effect has been associated with gradient-induced magnet heating in
high-field MRI5. Figure 3(B) contains the main result of this study.
It shows that eddy current pre-patterning of the passive shield substantially
reduces leakage field in peripheral regions (|z/∆z| > 4) and at high frequencies (f >1 kHz). Although the leakage field was higher near z = 0 and f < 1
kHz, inspection of Fig. 3 reveals that the patterned-copper FRF is more localized
in space and frequency, showing consistent spectral shapes across z. This reflects the leakage field originating from a single current mode
defined by the pattern.
The
localization can also be appreciated in Fig. 4, where the root-mean-square of
FRF as a function of z is plotted for different frequency bands. The patterned
copper shield reduces FRF “energy” in the 1 ~ 2 kHz band at most z positions
compared to solid copper.
The
spectral shape of the patterned-shield leakage field could be well predicted from
impedance measurement as shown in Fig. 5. The main magneto-mechanical resonance
peaks observed in Fig. 3(B) are reproduced in the calculated spectrum (Fig.
5(B)).Discussion and Conclusion
The
main idea of patterned passive shield is to preserve the desired, shielding eddy
current while minimizing vibration-induced eddy currents inside the magnet.
This strategy is most effective when the two eddy currents are orthogonal. Our
results show that patterned passive shield fundamentally changes the
spatial/spectral footprint of the motional eddy current, and can significantly
reduce the leakage field in certain frequency bands and spatial regions. Leakage
field peaks may further be reduced through structural reinforcements that
target specific vibration modes. It should be noted that such measures as well
as the exact cut-out patterns depend on the particular leakage field of a given
gradient coil.
In
conclusion, we have demonstrated the concept of motional eddy current-reduced
passive shield of an MRI gradient coil through mode-limiting cut-outs on a
conductive plate. Improved leakage field suppression can reduce magnet-gradient interaction in high-field, high-gradient MRI scanners9,10.Acknowledgements
This project was partly supported by NIH U01EB027696. The abstract does not necessarily represent the views of the funding agency.References
[1] P. Roemer et al., US 4,737,716 (1988), “Self-shielded gradient coils for nuclear magnetic resonance imaging”
[2] S-K. Lee and J. Schenck, Journal of Applied Physics 133, 174504 (2023), “Generalized magnetostatic target field method for shielded magnetic field coils in a separable coordinate system”
[3] G. Mulder et al., WO 00/25146 (2000), “MRI apparatus with a mechanically integrated eddy current shield in the gradient system”
[4] J-B. Mathieu et al., US 10,281,538 B2 (2019), “Warm bore cylinder assembly”
[5] Y. Hua et al., ISMRM 2012, Abstract 2580, “Gradient coil induced Joule heating in a MRI magnet”
[6] L. Jiang and T. Havens, IEEE Transactions on Applied Superconductivity 22(3), 4400704 (2012), “Environmental vibration induced magnetic field disturbance in MRI magnet”
[7] M. Westphal, US 7,514,928 B2 (2009), “Superconducting magnet configuration with reduced heat input in the low temperature regions”
[8] S-K. Lee and Y. Hua, US 11,774,531 B1 (2023), “Systems, assemblies, and methods of suppressing magnet-gradient interaction in magnetic resonance systems”
[9] T. K.-F. Foo et al., ISMRM 2023,
Abstract 4586, “Compact 7T: Progress in construction and assembly of a
low-cryogen, high-performance, head-only high-field MRI system”
[10] N. Boulant et al., Magnetic
Resonance Materials in Physics, Biology, and Medicine 36, 175 (2023), “Commissioning
of the Iseult CEA 11.7T whole-body MRI: current status, gradient-magnet
interaction tests and first imaging experience”