Keywords: Other AI/ML, Machine Learning/Artificial Intelligence
Motivation: MRE is a reliable, quantitative method for the assessment and staging of liver fibrosis. The standard manual MRE image prescription requires proper placement over the liver to ensure consistent MRE quantification. Scan positioning is relatively time-consuming and prone to error and inconsistency.
Goal(s): To develop and implement an automated methodology for MRE prescription from localizers, trained entirely from technologist-prescribed clinical exams.
Approach: Extracted MRE scan coordinates from 354 clinical exams and trained a YOLOv8-nano object detection network to predict prescription planes from a multi-plane localizer series.
Results: We successfully developed a method for automated MRE prescription with implementation on a clinical MRI system.
Impact: Automatic image plan prescription for MRE can minimize technologist-dependent planning errors and scan inconsistency. This may lead to subsequent improvements in both the value and reproducibility of MRE as a quantitative biomarker of liver fibrosis.
We acknowledge support from NIH grant R01EB031886. We wish to acknowledge support from GE Healthcare who provides research support to the University of Wisconsin. We wish to acknowledge support from the University of Wisconsin-Madison Office of the Vice Chancellor for Research and Graduate Education with funding from the Wisconsin Alumni Research Foundation, as well as from the UW Departments of Radiology and Medical Physics.
Dr. Reeder is the John H. Juhl Endowed Chair of Radiology.
This material is based upon work supported by the National Science Foundation Graduate Research Fellowship Program under Grant No. DGE-2137424. Any opinions, findings, and conclusions or recommendations expressed in this material are those of the author(s) and do not necessarily reflect the views of the National Science Foundation.
1. Venkatesh SK, Ehman RL. Magnetic resonance elastography of abdomen. Abdom Imaging. 2015; 40(4):745–759.
2. Guglielmo FF, Venkatesh SK, Mitchell DG. Liver MR Elastography Technique and Image Interpretation: Pearls and Pitfalls. RadioGraphics. 2019; 39(7):1983–2002.
3. Wagner M, Corcuera-Solano I, Lo G, et al. Technical Failure of MR Elastography Examinations of the Liver: Experience from a Large Single-Center Study. Radiology. 2017; 284(2):401–412.
4. Blansit K, Retson T, Masutani E, Bahrami N, Hsiao A. Deep Learning–based Prescription of Cardiac MRI Planes. Radiology: Artificial Intelligence. 2019; 1(6).
5. Geng R, Buelo CJ, Sundaresan M, et al. Automated MR Image Prescription of the Liver Using Deep Learning: Development, Evaluation, and Prospective Implementation. Magnetic Resonance Imaging. 2023; 58(2):429–441.
Figure 1. Process diagram illustrating the generation of an automatic MRE image prescription from a multi-plane localizer acquisition. A localizer series is acquired during end-expiration. The neural network is applied to each 2D localizer image to predict the MRE prescription center and FOV in the corresponding orientation. The coordinates are averaged to produce the fully-parameterized multi-slice MRE prescription. The MRE image prescription appears on the console, after which the technologist may begin data acquisition.
Figure 2. Automatically-prescribed MRE slices (blue) demonstrate good agreement with manually-prescribed volumes (red). Three representative cases from the evaluation set with good agreement between prescriptions and sufficient coverage of the widest transverse liver segment are shown. Each volume consists of 4 slices (thickness=10mm, slice gap=0mm). All automatically-placed MRE slices are well-centered in the AP and LR directions, with accurate positioning in the SI direction to cover sufficient liver tissue while avoiding the lungs and heart.
Figure 3. The automatically-generated image prescriptions in the evaluation dataset show good accuracy across all directions. A boxplot of the differences in position between the centers of each prescription in each direction is shown. The error between the centers of the automatic and manual prescriptions is tightly distributed around 0mm for all directions. The deviation shown is to be expected, as there is variability in the absolute accuracy of the prescriptions in the training and evaluation datasets.
Figure 4. The proposed method for automatic MRE prescription (shown in blue) shows high precision between multiple acquisitions in the same healthy volunteer after removal and repositioning. Each multi-plane localizer series was acquired in a 17-second end-expiration breath hold. For each exam, the automatically-prescribed volume demonstrates sufficient coverage of the widest transverse liver segment.