Hongbae Jeong1, Joshua Guag1, and Ananda Kumar1
1Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, MD, United States
Synopsis
Keywords: Safety, Safety
Motivation: The TF validation process is time-consuming work. Here, we reduced the burden of TF approach using a low power tabletop E-field generator.
Goal(s): Using the TF approach to estimate the RF-induced heating of the stent tip at 3T using a E-field generator and compared predicted values against the estimated values.
Approach: TF of the stent was measured via piX method at 128MHz . TF was validated by exposing the device under diverse test fields using table-top E-field generator. The body coil was used to evaluate the TF approach.
Results: The stent tip heating was estimated using TF approach with a low-power E-field generator.
Impact: TF approach could aid
estimating the RF-induced heating, not only active implantable device, but also
elongated passive implantable device. A compact E-field generator can be used
for TF validation.
Introduction
Radio-frequency (RF)
induced heating is one of the concerns inside MRI, especially for patients with
implantable medical device. Transfer function (TF) based RF-heating estimation
is often used to predict active implantable medical device (AIMD) lead wire heating,
which produces local hot-spot at the distal end of the device1. The traditional TF validation process requires test
measurements in various positions2 in the ASTM phantom which is time-consuming
work. The first goal of this
study is to use an alternative versatile, low power tabletop E-field generator
for RF field exposure to validate TF and to investigate the use of TF approach
to estimate the RF-induced heating of the elongated passive stent tip at 3T3 and compare to. The second goal is to compare predicted RF heating effect value
against the measured value using TF approach. Here, we used TF approach to estimate the
RF-induced heating of a stent and compared with dosimetric probe measurement
inside table-top E-field generator and with the thermal elevation inside a quadrature
body coil (QBC). Methods
The piecewise excitation
method, piX (ZMT, Switzerland), was used to measure TF at 128 MHz (Fig.1). A
stent sample (L=160mm, O.D.=7mm, thickness = 0.2mm) made with Nitinol was
positioned on a racetrack for the TF measurement, submerged under conductive tissue
simulating liquid (s: 0.47 S/m, er: 78) while piecewise excitation
coil for TF measurement swept along the device trajectory every 10 mm and
E-field was measured near the device distal tip using an optical Electric-field
probe (E1TDSx, SPEAG, Switzerland). The measured TF was validated by exposing
the device under diverse Test Exposure (TE) fields using the table-top E-field
generator called the MITS-TT (ZMT, Switzerland) (Fig.2(b)). Customized sixteen different
incident E-field exposures to test RF-induced heating on a device were
generated from MITS-TT by changing the magnitude and phase of the two channel transmit
source and the dosimetric E-field probe (EX3DV4, SPEAG, Switzerland) was
positioned at the tip of the stent to measure the local SAR (W/kg). Sim4Life
software (ZMT, Switzerland) was used to calculate incident E-field along the
device trajectory in MITS-TT, using the simulated tangential E-field (Etan) with
the measured TF, deposited power was estimated using MATLAB (Mathworks, USA).
An additional experiment was done for comparison as a popular method using the
3T QBC (MITS-3.0, SPEAG, Switzerland) by positioning the device inside ASTM gel
phantom (s: 0.47 S/m, er: 78)2 at two positions (Fig.4(b)) to measure the
temperature changes for 15 minutes with rectangular pulse, at 55 dBm input
power at 40% duty cycle (Figure 4 (a)).
Results
The measured TF of the sample stent at 3T is shown in Fig.1(b),
and the device positions superimposed on various TEs along the device
trajectory is shown in Fig.3(a). Fig 3(b) compares the measured SAR (W/kg) in
the MITS-TT with the estimated SAR from Etan calculated at the center of the
stent and at the outer surface. The SAR (W/kg) value deviations from the measured result from
MITS-TT were plotted in Fig 5(a) and the maximum deviation was -3.34 dB. The
temperature rise at the stent tip in an ASTM phantom were measured at two
locations in QBC as shown in Fig.4(c) where maximum 2.6 °C
elevation was observed. The measured
temperature rise at the tip of the stent is plotted in Fig. 4(c-d) for stents
placements shown for circularly polarized (CP) and linear mode QBC excitation. Fig
5(b) shows the temperature rise deviation (in dB) from measured results to the estimated
results based on Tier-3 method4 using Etan calculations.Discussion and Conclusion
The traditional TF
validation process requires test measurements at least in seven positions2
in the ASTM phantom which is time-consuming work, whereas using the table-top
E-field generator that produces multiple incident E-field exposures enabled the
completion of 16 test measurements in approximately 1 hour for a given sample position.
Here we have demonstrated the validity of the TF approach to assess the
RF-induced heating of elongated passive implantable device. We have also employed
an alternative compact E-field generator for the TF validation and the
assessment of the RF-induced heating. The stent heating was relatively low in
the QBC for the applied power to the coil, due to the low conductivity of
Nitinol with which it was made and potentially from the effects of the coating
placed on the stent. We demonstrated that the table-top E-field generator can
be used for RF-induced heating assessment using TF approach with more than 1/3 reduced measurement time, and good agreement on RF deposited power estimation. Acknowledgements
Disclaimer: The
mention of commercial products, their sources, or their use in connection with
material reported herein is not to be construed as either an actual or implied
endorsement of such products by the Department of Health and Human Services.
This presentation reflects the views of the authors and should not be construed
to represent FDA’s views or policies.
References
1. Park,
S.-M., Kamondetdacha, R. and Nyenhuis, J.A. (2007), Calculation of MRI-induced
heating of an implanted medical lead wire with an electric field transfer
function. J. Magn. Reson. Imaging, 26: 1278-1285. https://doi.org/10.1002/jmri.211592.
2.
ASTM 2182-19e2, Standard Test Method for Measurement of Radio
Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance
Imaging
3. Fujimoto
K, Angelone LM, Lucano E, Rajan SS and Iacono MI (2018) Radio-Frequency Safety
Assessment of Stents in Blood Vessels During Magnetic Resonance Imaging. Front.
Physiol. 9:1439. doi: 10.3389/fphys.2018.01439
4.
ISO 10974:2018, ‘Assessment of the safety of magnetic resonance imaging for
patients with an active implantable medical device.’