R. Allen Waggoner1, Makoto Uji1, Kenichi Ueno1, Chisato Suzuki1, Fumiaki Sato1, Xuemei Li1, Hiroyuki Kamiguchi1, and Masako Tamaki1,2
1RIKEN Center for Brain Science, Wako-shi, Japan, 2RIKEN Cluster for Pioneering Research, Wako-shi, Japan
Synopsis
Keywords: Multimodal, Multimodal, EEG, fMRI
Motivation: Evaluate the impact of EPI read-out gradient slew-rate on EEG amp heating, for EEG-fMRI experiments at 7T.
Goal(s): Identify slew-rates that will minimize heating of EEG amps, for EEG-fMRI experiments at 7T.
Approach: Test a range of EPI slew-rates while measuring the temperature of the EEG amp, during phantom tests at 7T.
Results: EPI read-out slew-rates of 120, 88, 61 and 46 T/m/s were tested and the slew-rates of 61 and 46 T/m/s were found to induce minimal heating of the EEG amp during EEG-fMRI experiments at 7T.
Impact: These results should help neuroscientists doing EEG-fMRI studies at 7T, to conduct their experiments without damaging the EEG amps.
Introduction
Whenever
non-standard devices are introduced into the MRI environment, testing to ensure
that the MRI environment will remain safe for both subjects and equipment is
important. There are numerous studies
combining EEG and fMRI at 3T and lower fields.
These experiments demonstrate that with appropriate caution, EEG-fMRI
studies can be performed safely at these field strengths. Moving to higher magnetic fields requires
further safety testing. Jorge et al. reported
that heating of the electrodes in the EEG cap were minimal, but significant
heating of the RF amp was observed.[1] In this study we explore the
impact of the slew rate of the EPI read-out gradient pulses, on heating induced
in the EEG amps at 7T.Methods
A phantom
EEG-MRI experiment was conducted on a 7 Tesla MRI system (Signa 7T, GE
HealthCare). The Brian Products EEG cap
was placed on a spherical water phantom, which was place in the 32-channel head
coil (Nova Medical). The EEG amplifiers
(BrainAmp MR Plus and BrainAmp ExG MR; Brain ProductsGmbH) were placed behind
the RF coil with the EEG cap attached. A
fiber optic temperature probe (Rugged Monitoring model L201), was placed
between the EEG and ExG amps, as illustrated in Figure 1. Previous measurements had shown this to be
the location of maximum heating. EPI
images were acquired for at least 5 minutes, with read-out gradient slew rates
of 120, 88, 61, and 46 T/m/s, while the temperature between the EEG amps was
continually monitored with the fiber optic temperature probe.
In addition, to evaluate the
impact of changing the slew rate on image quality, EPI images were acquired
with a human subject (30 yr. old male).
Data sets were acquired with slew rates of 120 and 61 T/m/s. The data sets were acquired with an isotropic
resolution of 1.5mm3, TE=28ms, multi-band factor=6, and TR = 800,
900ms respectively.Results
Figure 2
shows the measured temperature changes, for the EPI scans with each slew
rate. Both the slew rates of 120 and 80
T/m/s showed substantial heating of the EEG amps over the 5 min. test
periods. For the slew rate of 61 T/m/s,
only very modest heating was observed, with a maximum of 0.25 °C. For the slew rate of 46 T/m/s, the observed
temperature change, never rose above baseline.
Reducing the slew rate for 120 to 61 T/m/s, does not have
a pronounced impact on image quality, as demonstrated by the images in Figure
3. Since the EPI sequence employed in
this study uses ramp sampling, reducing the slew rate has only a minimal impact
on the overall read-out window, therefore there is not a dramatic loss of image
quality.Conclusions
Previous
studies have observed heating of the EEG amps during simultaneous EEG-fMRI at
7T. The results presented here
replicated that observation. In
addition, we demonstrated that by reducing the slew rate of the EPI read-out
gradient, the heating of the EEG amps can be dramatically reduced or even
eliminated. Further we demonstrated the
use of the reduced slew rates do not lead to a noticeable loss of image
quality. The results of the experiments
presented here demonstrate that by reducing the slew rate of the read-out
gradient, we can significantly reduce or even eliminate heating of the EEG amps. This demonstrates that EEG-fMRI can be done
safely at 7T, without excessive wear on the EEG amps.Acknowledgements
No acknowledgement found.References
1. Jorge
J, Grouiller F, Ipek Ö, et al. Simultaneous EEG–fMRI at ultra-high field:
Artifact prevention and safety assessment. NeuroImage 2015;105:132-144.