Denise Lichthardt1, Jens Wetzl2, Michaela Schmidt2, Peter Speier2, Christoph Tillmanns3, Armin M. Nagel1, and Daniel Giese2
1Friedrich-Alexander-University Erlangen-Nürnberg, Erlangen, Germany, 2Siemens Healthineers AG, Erlangen, Germany, 3Diagnostikum Berlin, Berlin, Germany
Synopsis
Keywords: Flow, Motion Correction
Motivation: 3D Phase contrast MRI of the coronary arteries remains challenging due to low SNR and motion.
Goal(s): Develop a 3D PC sequence for flow quantification in the coronary arteries at 3T with improved SNR and reduced motion artefacts compared to 1.5T.
Approach: The sequence was optimized by acquiring datasets separately, enabling inter-scan motion compensation while circumventing SAR limitations. Coronary flow was quantified in 5 subjects at 3T.
Results: Phantom measurements showed reduced image artefacts using the proposed sequence. In-vivo mean peak velocities per slice along the left and right coronary arteries were 15.9±5.0cm/s and 14.1±4.9cm/s respectively.
Impact: High SNR 3D flow measurements
in the proximal coronary arteries are now possible, as SAR restrictions at 3T have
been overcome. This may enable the non-invasive hemodynamic assessment of coronary arteries in
patients with CAD.
Introduction
Phase contrast (PC) MRI can be used for non-invasive acquisitions of
blood flow velocities1. 3D PC-MRI in the coronary arteries is
limited by the need for high spatial resolution due to small vessel sizes,
leading to low SNR, long scan durations, and therefore, increased
susceptibility to motion. Higher field strengths lead to improved SNR, however,
SAR restrictions limit their applicability due to the required preparation
pulses. We propose a sequence to address these challenges and evaluate 3D blood
flow velocities in the left (LCA) and right (RCA) coronary arteries in healthy
volunteers and patients at 3T.Methods
3D PC-MRI (VENC=50 cm/s, undersampling factor = 14, (Δx)3 =
(1.2mm)3, adiabatic T2 prep, SPAIR, ECG-triggering, respiratory
gating) was performed in diastole on a 3T MRI system (MAGNETOM Skyra, Siemens Healthineers
AG, Erlangen, Germany) using a research sequence. Measurements were performed
twice within one scan session on 3 healthy volunteers (2 females, 61.5±17.8 years) and were repeated on a
different day for volunteer 1 to analyze inter-session variability. Further
data were acquired on 2 patients (f, 84y and f, 26y) with suspected CAD and
suspected myocarditis, respectively. Since neither suspicion was confirmed by clinical CMR acquisitions,
data were included in the flow analysis across normal subjects. Additional data were obtained on 1 volunteer
(male, 60 years) using non-adiabatic T2 preparation and fat saturation at 3T (MAGNETOM
Vida, Siemens Healthineers AG, Erlangen, Germany). Due to protocol deviations,
this measurement was not considered in the comprehensive flow analysis. Data in
a static phantom were obtained on a 1.5T MRI system (MAGNETOM Sola, Siemens Healthineers
AG, Erlangen, Germany).
As opposed to conventional PC-MRI, the flow encoded/compensated datasets
were acquired in individual scans (Fig. 1). The resulting decrease in scan time
per scan alleviated SAR limitations at 3T.
Inter-scan motion was compensated through co-registration of
edge-filtered magnitude images (Matlab R2021b, The MathWorks, Natick, Massachusetts,
USA). For edge-filtering, the
approximate Canny algorithm was used. The obtained transformations were applied
to the corresponding magnitude and phase images. Furthermore, spin displacement
in between flow and spatial encoding was corrected2,3. The averaged
magnitude images were used for manual segmentation.Results and Discussion
Figure 2 shows phase difference images of a static phantom acquired with
an interleaved scan compared to phase difference images obtained from four individual
scans. Interleaved acquisitions show less variations in mean velocities across
repetitions. However, using individual scans, artefacts and phase gradients across
the image are reduced, resulting in improved image homogeneity. This could be
due to steady state disruptions in case of interleaved acquisitions, as this
requires alternating between different types of gradient waveforms (flow
compensating vs. flow encoding) for each TR.
Figure 3 compares in-vivo images obtained at 3T using an interleaved acquisition
to individual acquisitions in a volunteer with poor navigator efficiency.
Although overall limited image quality, improved delineation of the coronaries is
observed using the proposed acquisition.
In-vivo acquisitions at 3T show increased SNR compared to 1.5T (Figure 4a)
and improved sharpness after motion correction (Figure 4b), enabling better
delineation of the arteries. Peak velocities per slice (PVPS) along the vessel
centerlines for the LCA and RCA are shown in Figures 4c and d respectively for
an exemplary volunteer, comparing correction strategies. Motion and
displacement correction eliminate outliers and reduce overestimated peak values
caused by surrounding noisy voxels. Correction shifts in pixels along the RL,
FH and TP directions are shown in Figures 4e-g. Mean shifts are -0.3±0.6 voxels, 0.5±1.0 voxels and -0.1±0.5 voxels in
the respective directions.
Figure 5 shows good intra-session agreement with visually shifted
velocity curves for the LCA which could be due to a mismatch of the selected
starting points. Larger inter-session deviations are potentially due to
physiological variations.
Across all subjects and repetitions, the mean values of average PVPS in
the proximal (<25mm along the centerline) arteries are 11.7±3.5cm/s for the LCA and 9.4±3.6cm/s for the RCA. The mean peak PVPS are 15.9±5.0cm/s and 14.1±4.9cm/s respectively. The mean PVPS are in line
with previous results obtained at 1.5T4, therefore, results remain consistent
at higher field strengths. Conclusion
This study demonstrates the feasibility of 3D PC-MRI of the coronary arteries at 3T. Results agree with previous studies at 1.5T while benefitting from higher SNR and increased sharpness after motion correction. The method not only circumvents SAR restrictions at 3T, but also reduces intra-scan motion effects due to shorter scan times. In case of severe artefacts, single datasets may therewith be reacquired at a lower expense in scan time. Future works will evaluate the method in patients with confirmed CAD.Acknowledgements
No acknowledgement found.References
1. Pelc, Norbert J., et al. "Encoding strategies for three‐direction phase‐contrast MR imaging of flow." Journal of
Magnetic Resonance Imaging 1.4 (1991): 405-413.
2. Steinman, David A., et
al. “Combined analysis of spatial and velocity displacement artifacts in phase
contrast measurements of complex flows.” Journal of Magnetic Resonance
Imaging 7.2 (1997): 339-346.
3. Roberts, Grant S., et al. "Virtual injections using 4D flow MRI
with displacement corrections and constrained probabilistic
streamlines." Magnetic resonance in medicine 87.5 (2022):
2495-2511.
4. Lichthardt, Denise,
et al. "3D phase contrast blood flow measurements in the coronary
arteries."
Proc. Intl. Soc. Mag. Reson. Med. 32:
3820 (2023).