Peter Gatehouse1, Andrew Scott1, Gaby Captur2, Muhammad Usman1, Ronald Mooiweer1, Dudley Pennell1, and Sonia Nielles-Vallespin1
1Cardiovascular MR, Royal Brompton Hospital, London, United Kingdom, 2University College London, London, United Kingdom
Synopsis
Keywords: Phantoms, New Devices, Precision & Accuracy, Field Camera, Virtual Phantom
Motivation: Quality assurance (QA) of quantities derived from MRI requires elaborate phantoms. Instead, we send modulated RF signals into the scanner representing any initial test object (previous “Virtual Phantom ViP”) avoiding physical phantom difficulties.
Goal(s): Proof-of-principle of a novel step for QA of derived quantities, by combining two previous methods: a Field Camera (FC) to govern ViP.
Approach: QA compares outputs from the unmodified scanner against the initial object. Tests evaluated the prototype technically, plus a derived QA example: myocardial bloodflow by first-pass contrast-enhanced myocardial perfusion.
Results: Phase stability without phase-locking to scanner was marginal, while first demonstrating FC+ViP QA of a derived quantity.
Impact: Quality
assurance of derived quantities can require elaborate physical phantoms. Instead,
we demonstrate novel field camera governance of the unmodified scanner's virtual phantom acquisition, reconstruction and analysis. We call for vendor cooperation in a new standardised
inexpensive quality control interface.
Introduction
Quality assurance (QA) of quantities
derived from MRI currently requires elaborate phantoms as reference
standards1-7. There is an unsolved need regarding practical
difficulties of some physical phantoms especially for intersite or intervendor work.
Alternatively, previous work has transmitted modulated RF signals into the scanner (“Virtual
Phantom ViP”)8-11 with controlled modulations for any required
physics, representing the Simulated Test Object (STO) “input”, to be compared with the
scanner “output” to achieve the QA purpose. This work aims for
proof-of-principle of a novel step in capability for QA of derived quantities
by combining two prior methods, using a prototype Field Camera (FC)12-15
to govern ViP (FC+ViP, named Ariel16).
Methods
Like ViP, Ariel starts with the STO modeling the
quantity where QA is required (Figure 1). Using the FC to record the entire
pulse sequence, Ariel calculates by Bloch computation17 the continuous
RF signal that would be emitted by the physical version of the STO during the sequence. Ariel transmits this calculated RF signal for reception,
reconstruction and any further parametric derivations as normally performed by the scanner and further analysis programs used routinely. The analysis output is compared to the input STO for QA. Ariel
operates continuously during the pulse sequence, after one initial trigger to
synchronise with the scanner. Images were received by an
unmodified scanner (3T Vida, Siemens Healthineers, “4-channel Flex Coil, small”
around the Ariel Tx loop), reconstructed as sum-of-squares magnitude, without
gradient nonlinearity correction nor other filtering. For
nutation during BLOCH, the GRARF B1 amplitude was calibrated using the flip
angles and Mz. A conventional phantom enabled the usual scanner prescan “adjustments”.
Tests A-C (parameters, Figure 2) using STO T1=200ms T2=20ms conducted basic technical
evaluation of the prototype, i.e. these are not proposed QA applications.
Test A: Fidelity of GRARF and FE Rolloff: STO = square
of uniform PD, T1 and T2, side 285mm. A total of 30 Ariel scans investigated PE
ghosting and GRARF recordings.
Test B: PSFs and long-term phase stability: STO
= 9 points (2.2x3.0mm FExPE) distributed obliquely over FOV, repeated for 60’.
Test
C: Resolution and Slice-profile (see Figure 3).
Test E (parameters, Figure 2) is the main aim, applying Ariel to QA a
derived quantity (Figure 4): for example, derivation of myocardial blood flow (MBF) from first-pass
contrast-enhanced (GBCA) perfusion18; MBF is clinically interesting and requires QA4.Results
Test A:
Phase-encode ghosting was consistent for repeated scans. It changed for each
GRARF recording, and reduced slightly using 5-averaged GRARF. FE magnitude
rolled off consistently across all Test A-D images (symmetrically to 58% at ±141mm).
Test B: PSF FWHM
along FE was reliable, but PE FWHM of offset points reached 6.7mm. Over 60’ the
PE FOV drifted 1.5cm monotonically.
Test C: (Figure 3):
Phase-encode resolution was again compromised, while slice-profile was reasonable
after correcting for FE rolloff.
Test E: The magnitude, calculated T119, hence [GBCA] of each frame (Figure 4d) enabled deconvolution analysis for MBF showing reproducible
prototype agreement with the input STO values of each myocardial segment (Figure
5). Discussion
While Test E perfusion MBF agreement vindicated Ariel’s
aim, limitations are noted: For this prototype, deconvolution analysis for Test
E was locally written in Matlab, but should usually evaluate all the software
used clinically. Frequency-encode rolloff was associated with slow prototype
microcontroller RFSYN timepoints (6.4us), not the AD9914 IC used. Phase-encode
ghosting, degrading PE resolution, arose from imperfect GRARF recordings, not
OCXO instability. However, temporal phase stability was borderline; complicated
k-space sampling or long scans would demand phase-locked scanner and Ariel
clocks. The OCXO of RFSYN was free-running because Ariel required no
connections other than the standard fibre-optic trigger output of the scanner.
The original notion
of Ariel would calculate and transmit in real-time, even calibrating the
scanner “tuning” adjustments; perhaps feasible20 but beyond the prototype's design. Further, the GRARF imperfections required correction work.
Scanner Larmor frequency modulation was observed (perhaps “B0 shift compensation”)
which fortunately cancelled out. Multiple/multiplexed RFSYN into separate loops
distributed across the PE direction could support parallel imaging. Conclusions
For the QA aim,
Ariel extended ViP by combining it with a prototype field camera and
continuous operation, needing no prior knowledge of sequence or ADC timings,
k-space ordering, accelerations nor reconstruction strategies (ML might require realistic STO images21,22).
The extra
electronics of Ariel, particularly GRARF, might be eliminated23: we propose24 that a simple inexpensive ‘interface’ for QA (cf. automotive OBD2),
might be standardised among MRI vendors, to eliminate GRARF (by far the hardest
work of Ariel). With little extra hardware, “built-in” standardised QA of derived
quantities could be openly available for all MR.Acknowledgements
The Ariel idea occurred during initial
discussion of the myocardial T1 test phantom (T1MES) with the cardiovascular
MRI research group of Professor James Moon, then at the National Heart Hospital
London.References
1 Cashmore MT, McCann AJ, Wastling SJ et al. Clinical
quantitative MRI and the need for metrology (2021). Br J Radiol 94: 20201215
2 Keenan KE, Ainslie M,
Barker AJ et al. Quantitative magnetic resonance imaging phantoms: a review and
the need for a system phantom (2018). Magn Res Med 79:48-61.
3 Captur G, Gatehouse P,
Keenan KE et al. A medical device-grade T1 and ECV phantom for global T1
mapping quality assurance—the T1 Mapping and ECV Standardization in
cardiovascular magnetic resonance (T1MES) program (2016). JCMR 18:58
4 Chiribiri A, Schuster A,
Ishida M et al. Perfusion Phantom: An Efficient and Reproducible Method to
Simulate Myocardial First-Pass Perfusion
Measurements with
Cardiovascular Magnetic Resonance (2013). Magn Res Med 69:698-707.
5 Stupic KF, Ainslie M,
Boss MA et al. A standard system phantom for magnetic resonance imaging (2021).
Magn Res Med 86:1194-1211.
6 Gadda G, Cocozza S,
Gambaccini M et al. NO-HYPE: a novel hydrodynamic phantom for the evaluation of
MRI flow measurements (2021). Med and Biol Eng and Comput 59:1889-1899.
7 Swailes NE, MacDonald ME,
Frayne R. Dynamic phantom with heart, lung and blood motion for initial
validation of MRI techniques (2011). JMRI 34:941-6.
8 Saint-Jalmes H, Eliat
P, Bezy-Wendling J et al. ViP MRI: virtual phantom magnetic resonance imaging
(2014). Magn Reson Mater Phy 27:419-24.
9 Salvati R, Hitti E, Bellanger J-J et al.
Fat ViP MRI: Virtual Phantom Magnetic Resonance Imaging of water-fat systems
(2015). MRI 2015. doi:10.1016/j.mri.2015.12.002
10 Saint-Jalmes
H, Bordelois A, Gambarota G. Virtual
phantom magnetic resonance imaging (ViP MRI) on a clinical MRI platform (2018).
Med Phys 45:250-7.
11 Barantin L, Le Pape A, Akoka S. A new
method for absolute quantitation of MRS
Metabolites (1997). Magn Reson Med 1997;38:179-82.
12 Brunner DO, Dietrich BE, Çavuşoğlu M et
al. Concurrent recording of RF pulses and gradient fields – comprehensive field
monitoring for MRI (2015). NMR Biomed 29:1162-72.
13 Senaj V, Guillot G, Darrasse L. Inductive measurement of
magnetic field gradients for magnetic resonance imaging (1998). Review of
Scientific Instruments 69: 2400.
14 Sipilä PT. Real-Time
Magnetic Field Monitoring in Magnetic Resonance Imaging (2011).
Doktors-Ingenieur (Dr.-Ing.) Thesis, Technischen Universität München.
15 Dietrich BE, Brunner DO,
Wilm BJ et al. A Field Camera for MR Sequence Monitoring and System Analysis
(2016). Magn Res Med 75:1831-40.
16 Ariel: the archaic aerial demon, not the Mermaid.
17 Benoit-Cattin H, Collewet G, Belaroussi B,
Saint-Jalmes H, Odet C. The SIMRI project: a versatile and interactive MRI
simulator. J Magn Reson. 2005 Mar;173(1):97-115.
18 Kellman P, Hansen MS,
Nielles-Vallespin S et al. Myocardial perfusion cardiovascular magnetic
resonance: optimized dual sequence and reconstruction for quantification. JCMR
2017;19:43
19 Cernicanu A, Axel L.
Theory-Based Signal Calibration with Single-Point T1 Measurements for
First-Pass Quantitative Perfusion MRI Studies (2006). Acad Radiol 13:686–93
20 MARTe2 , for example https://vcis.f4e.europa.eu/marte2-docs/master/html/
21 Buoso S, Joyce T,
Schulthess N, Kozerke S. MRXCAT2.0: Synthesis of realistic numerical
phantoms by combining left-ventricular shape learning, biophysical
simulations and tissue texture generation (2023). J Cardiov Magn Reson
(2023) 25:25.
22 Weine J, McGrath C, Kozerke S. CMRsim - A Python package for MRI simulations incorporating complex
organ motion and flow (2023). ISMRM2023:2393
23 Kennedy M, Lee Y,
Nagy Z. An Industrial Design Solution for Integrating NMR Magnetic Field
Sensors into an MRI Scanner (2018). Magn Reson Med 80:833-9.
24 Gatehouse, Captur et al Submitted to MAGMA.
25 Zur Y, Wood ML,
Neuringer LJ. Spoiling of transverse magnetisation in steady-state sequences
(1991). Magn Reson Med 21:251-63.
26 Sung K, Nayak KS.
Design and use of tailored hard-pulse trains for uniform saturation of
myocardium at 3 Tesla (2008). Magn Reson Med 60:997-1002.