Jaime Mata1, John P. Mugler III1, Bastiaan Driehuys2, David Mummy2, Jamie MacLeod3, and Yun Michael Shim1,3
1Radiology & Medical Imaging, University of Virginia, Charlottesville, VA, United States, 2Radiology, Duke University, Durham, NC, United States, 3Medicine, University of Virginia, Charlottesville, VA, United States
Synopsis
Keywords: Lung, Hyperpolarized MR (Gas)
Motivation: Study on the effects of e-cigarette use is needed to understand its damaging effects on the pulmonary system.
Goal(s): Anticipating a need for multi-center clinical trials in the future this study, performed for the first time, studied the repeatability of Xe-129 gas-exchange MRI of the same subjects at two centers.
Approach: Same subjects were imaged at University of Virginia and Duke University to directly assess technique harmonization and repeatability of Xe-129 metrics.
Results: Note the statistically insignificant differences for each subject, for ventilation and gas-exchange, with signal-to-noise ratio for ventilation scans being the only statistically significant parameter (p=0.027) to differ between the sites.
Impact: Harmonization of Xe-129 MRI gas-exchange and
ventilation imaging methodology among the two sites was successfully achieved,
opening the way for a large clinical trial with these two institutions.
Introduction
Electronic-cigarette
(e-cigarette) use is classified by the Centers for Disease Control and
Prevention as unsafe and is associated with outbreaks of severe lung injury1.
Besides nicotine, which is highly addictive, e-cigarettes contain harmful
unregulated substances. A detailed study on the effects of e-cigarette use is needed
to understand its damaging effects on the pulmonary system. MRI using hyperpolarized xenon-129 (Xe-129) gas
is an FDA-approved, emerging technology capable of providing, in a single
breath-hold, detailed ventilation images of the lungs as well as physiological
maps of Xe-129 distribution into multiple lung compartments (gas-exchange
imaging), specifically into lung parenchyma (membrane) and red-blood-cells
(RBC)2,3,4. Xe-129 MRI studies in healthy subjects and “healthy
smokers” discriminated between these two groups before spirometry detected loss
of lung function5, suggesting that Xe-129 MRI is a sensitive tool
for detecting early-stage lung disease. The current study is the first step to
exploit the potential of Xe-129 MRI ventilation and gas-exchange parameters to detect
lung disease occurring in the early days of e-cigarette use. Anticipating a need for multi-center clinical
trials in the future, this study performed for the first time, repeatability
studies of Xe-129 gas-exchange MR imaging of the same subjects at two
centers.
Methods & Materials
This two-site clinical trial used Xe-129 MRI for
ventilation and gas-exchange imaging to probe multiple lung compartments (air
spaces, membrane and RBC). The same subjects were imaged at the University of
Virginia and Duke University to directly assess technique harmonization and
repeatability of Xe-129 metrics.
Five healthy volunteers (three females and two
males; 25.2±2.8 years) and two e-cigarette users (one female and one male;
19.5±0.5 years) underwent Xe-129 MRI studies at both institutions. Each healthy
volunteer was rescanned within two months and each e-cigarette user within a
month. E-cigarette users had used e-cigarettes on a regular basis (minimum of
one pod per week) for at least 12 months.
MRI techniques for scanner calibration (repeated
acquisition of free-induction-decays), ventilation imaging (multi-slice
Cartesian gradient echo) and three-dimensional gas-exchange imaging (1-point
Dixon spectroscopic imaging) were implemented as recommended by the 129Xe MRI Clinical
Trials Consortium6. To
facilitate harmonization, each technique was implemented by an investigator at
one institution for the two MR scanners involved (Duke University: Siemens
Prisma Fit, software Numaris/X VA30A; University of Virginia: Siemens Prisma
Fit, software Numaris/4 VE11C). To
maximize consistency, the same source-code base was used for the implementation
at both sites, even though the software levels were different. Ventilation and
gas-exchange imaging post-processing, using the binning method, was performed
at Duke University. Secondary statistical analyses were done at the University
of Virginia, using paired t-test for the comparisons.
Each site polarized Xe-129 using a commercial
polarizer (model Xe9820, Polarean Inc, USA), achieving polarizations of 25-30%,
and employed identical commercially available RF vest-coils tuned to the Xe-129
frequency at 3 Tesla (Clinical MR Solutions, USA). The total volume of Xe-129
gas mixed with nitrogen was approximately 20% of the subject’s forced vital
capacity, measured by spirometry immediately prior to the MRI scan.Results
Figures 1 and 2 show ventilation and gas-exchange
results, respectively, for the same subject from both sites, including detailed
global analyses. Note the very similar values and images from both sites. The
table in Figure 3 shows the quantitative differences between global-analysis metrics
from each site for each subject (value from Duke University subtracted from
value from University of Virginia). Note the statistically insignificant
differences for each subject, for ventilation and for gas-exchange, with
signal-to-noise ratio for the ventilation scans being the only statistically
significant parameter (p=0.027) to differ between the sites. While differences between the two sites were
small, consistent bias of gas exchange parameters were noted (plot in Figure 3),
with Duke University values being higher than those from University of
Virginia. This might be explained by
differences in lung-inflation level; for 5 of 7 subjects, the lung inflation
level was higher at the University of Virginia. Conclusion
Harmonization
of Xe-129 MRI gas-exchange and ventilation imaging methodology among the two
sites was successfully achieved, opening the way for a large clinical trial
with these two institutions.Acknowledgements
The authors would like to thank all our nurses, research coordinators, MRI technologists and other staff for their work. References
1- https://www.cdc.gov/tobacco/basic_information/e-cigarettes/Quick-Facts-on-the-Risks-of-E-cigarettes-for-Kids-Teens-and-Young-Adults.html
2-Driehuys B, Martinez-Jimenez
S, Cleveland Z, et al. Safety and tolerability of hyperpolarized 129Xe MR
imaging in healthy volunteers and patients. Radiology 2012;262:279-89.
3-Joseph G. Mammarappallil, EM,
et al. Identification of gas exchange phenotypes using hyperpolarized 129Xe MRI
in patients with chronic obstructive pulmonary disease (COPD). Am J Respir Crit
Care Med 2019; 199:A1122.
4-
Guan S, Tustison N, Qing K, et al. 3D Single-Breath Chemical Shift Imaging
Hyperpolarized Xe-129 MRI of Healthy, CF, IPF, and COPD Subjects. Tomography 2022; 8(5):2574-2587.
5- Qing K, Tustison NJ, Mugler JP
III, et al. Probing changes in lung physiology in COPD using CT,
perfusion MRI, and hyperpolarized Xenon-129 MRI. Acad Radiol 2019; 26:326-34.
6- Niedbalski PJ, Hall CS, Castro M, et al. Protocols for
multi-site trials using hyperpolarized 129Xe MRI for imaging of
ventilation, alveolar-airspace size, and gas exchange: A position paper from
the 129Xe MRI clinical trials consortium. Magn Reson Med 2021;
86:2966-2986.