Felix Spreter1, Simon Reiss1, Johannes Fischer1, Ali Caglar Özen1, Alexander Maier2, and Michael Bock1
1Division of Medical Physics, Dept. of Diagnostic and Interventional Radiology, University Medical Center Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, 2Dept. of Cardiology, University Medical Center Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany
Synopsis
Keywords: Interventional Devices, Interventional Devices, Arterial Spin Labeling
Motivation: Intra-Arterial-Spin Labeling (iASL) could offer an alternative for myocardial perfusion measurements without the use of contrast agents by labeling blood with a catheter transmit RF-coil.
Goal(s): To study the parameters influencing the iASL labeling efficiency inside the artery to improve the SNR.
Approach: The complex iASL process is investigated with simulations and in-vitro measurements for different coil geometries (solenoid, loop), driving currents, coil orientations relative to the magnetic field, and blood flow velocities.
Results: The simulations were in good agreement with measurements, show a maximal mean flip angle created by the labeling pulse of 105°, and a threshold behavior for the transmit current.
Impact: The improvement of the labeling process inside
coronary arteries started in this study, is an important step towards making
ASL a viable method for myocardial perfusion quantification.
Introduction
Regional perfusion assessment in the myocardium is a
crucial diagnostic tool for ischemic heart disease. In interventional MRI underperfused
ischemic areas supplied by an individual artery can be identified by injection
of Gadolinium-based contrast agents1,2; however, injections cannot be
repeated very often due to the accumulation of the contrast agent. Arterial
spin labeling (ASL) is an alternative for perfusion quantification: here, the
blood itself is labeled by an RF pulse eliminating the need for potentially
nephrotoxic contrast agent injections3. Recently
we proposed intra-Arterial Spin Labeling (iASL), where the labeling pulse is
applied from inside the artery by transmitting the pulse via a small transmit
coil at the catheter tip4. In
this study, the complex iASL process inside the artery is simulated, and
in-vitro measurements are performed to find the best coil geometry and to optimize
the RF pulse shape. Methods
The labeling process was simulated in MATLAB. First,
the B1-field of a coil at
the catheter tip was calculated using Biot-Savarts law (i.e., the coil dimensions
are assumed to be much smaller than the RF wavelength). Then, the dynamic
change of the magnetization M flowing through this B1-field was computed. In each voxel, the RF-induced rotation
angle of M were calculated from the local transverse B1-component of the RF field,
the flow velocity and the voxel size: $$\Delta \vec{\phi}(x,y,z)=\gamma \cdot \vec{B}_1(x,y,z) \cdot \frac{dz}{v(x,y)}$$
The flow velocity around the catheter (Ø 2 mm) inside the vessel (Ø 6 mm) was calculated using the
Navier-Stokes equation. The flow was assumed to be laminar. A schematic of the
simulation is shown in Figure 1. Inside
a vessel cross section - downstream of the coil - an effective flip angle map was
calculated for the magnetization, and the mean flip angle $$$\phi_{mean}$$$ across the vessel was used as a measure of the
labeling efficiency. Two types of coils were simulated: a rectangular single
loop (length: 20 mm, width: 2.1 mm) and
a solenoid coil (length: 5 mm, diameter: 2.1 mm, number of windings: 7). The
current inside the coil, the flow velocity around the coil and the orientation
of the coil relative to B0
were varied to investigate their influence on the labeling efficiency.
The simulation results were compared to phantom
measurements at a 3T clinical MRI system (Siemens Prisma Fit). Therefore, both
coil geometries were built from enameled copper wire (Ø 0.15 mm), connected with a micro-coaxial cable via
a tuning/matching circuit to an RF signal generator (KEYSIGHT EXG Analog Signal
Generator N5171B). The coils were mounted at the center of a tube (Ø 6 mm) on a 3D-printed holder with
outer dimensions of a 6F catheter. A constant water flow of 2 ml/s was created
by gravitational pressure difference. The signal in a cross section 3 cm
downstream of the coil was measured for different RF currents in the coil and
compared to the simulation results. The coil was positioned at a 70° angle to B0,
which corresponds to the typical position of a coronary guiding catheter in the
left coronary ostium.Results
The mean simulated flip angle for the 2 coil geometries is shown in Figure
2 as a function of the RF current. A maximal angle of 105° is reached for the solenoid
coil whereas the maximum for the rectangular coil was 92°. The flip angle dependency
on the orientation and the flow velocity are shown in Figure 3. Interestingly, only
small differences are present between the coil geometries even though they
create substantially different B1
profiles. For the rectangular coil the measured signal pattern in a downstream
vessel cross section agrees well with the simulation (Fig. 4). The signal
reduction between labeled and reference image is shown in Fig. 5.Discussion & Conclusion
Based in this study the labeling efficiency of common catheter coil
geometries is limited to a mean flip angle of about 105°, which is reached for
a solenoid coil with 7 windings stretched over 5 mm mounted on a 6F catheter and
an applied current of 10 mA. This can be explained by the inhomogeneous
labeling over the cross section of the artery, the flow velocity profile and the
confined geometry of catheter coils. The phantom measurements agree well with
the simulations, and the small differences can be attributed to the limitations
of the flow model and inaccuracies of the coil positioning. The simulation
framework will thus be used to optimize coil geometries and transmit pulse
shapes for in-vivo iASL experiments in a porcine model.Acknowledgements
Grant
support by the German Research Foundation (DFG) under RE 4876/1-1 is gratefully
acknowledged. This study is part of SFB1425
project P15, funded by the Deutsche Forschungsgemeinschaft (DFG, German
Research Foundation #422681845).References
1. Heidt, T. et al. Real-time magnetic resonance
imaging – guided coronary intervention in a porcine model. Sci Rep 9,
8663 (2019).
2. Heidt, T. et al.
Magnetic resonance imaging for pathobiological assessment and interventional
treatment of the coronary arteries. Eur Heart J Suppl 22, C46–C56
(2020).
3. Kober, F., Jao, T.,
Troalen, T. & Nayak, K. S. Myocardial arterial spin labeling. Journal of
Cardiovascular Magnetic Resonance 18, 22 (2016).
4. Reiss, S. et al.
Quantifying myocardial perfusion during MR-guided interventions without
exogenous contrast agents: intra-arterial spin labeling. Zeitschrift für
Medizinische Physik S0939388923000028 (2023)
doi:10.1016/j.zemedi.2023.01.002.