Anne Josset1, Ounay Ishak1, Jonathan Vappou1, and Élodie Breton1
1Université de Strasbourg, CNRS, ICube, UMR7357, Strasbourg, France
Synopsis
Keywords: Thermometry/Thermotherapy, Thermometry
Motivation: Monitoring thermal ablations with PRFS thermometry ensures therapy safety and efficiency. Spectral fat saturation cannot be 100% efficient: the remaining fat signal may affect PRFS temperature measurements.
Goal(s): This work experimentally evaluates the residual signal after spectral fat saturation, and the error it causes in PRFS thermometry.
Approach: The IDEAL algorithm is used to quantify fat signal after spectral fat saturation. The error in PRFS thermometry with spectral fat saturation is evaluated in fat-water phantoms.
Results: The residual fat signal after spectral fat saturation leads to errors in PRFS thermometry, that increase with the fat content, and oscillate with TE and absolute temperature.
Impact: This work demonstrates
experimentally that PRFS temperature errors can be significant during hyperthermia
despite fat suppression methods. Careful selection of the TE according to
spectral fat saturation strategy can mitigate such errors and improve the
accuracy of PRFS thermometry.
Introduction
MRI-PRFS
(Proton Resonance Frequency Shift)1 is the gold standard for the
monitoring of MR-guided thermal ablations. PRFS relies on the variation of the
resonance frequency of hydrogen in water molecules. In comparison, the
frequency of hydrogen in fat molecules does not change with temperature,
resulting in errors in temperature estimates in fat-containing tissues. Spectral
fat saturation is hence employed to suppress the signal from fat in PRFS
thermometry. Spectral fat saturation is commonly optimized for minimizing the
signal of fat in magnitude images, even though the 2 fat peaks close to water
cannot be saturated. This leads to the choice of saturation pulse flip angles
differing from 90° in combination with recommendations regarding TE (in/out
of-phase)2,3. Currently, the optimal TE for PRFS is chosen close to
the T2*4 of the tissue, the link between TE and the effectiveness of
spectral fat saturation is not considered, and fat signal is considered fully
attenuated. In this work, we quantify the remaining fat signal after fat
saturation, and experimentally evaluate the impact of the residual fat signal on
PRFS temperature monitoring.Methods
Experiments were conducted on a 1.5T MRI system (MAGNETOM
Sola, Siemens Healthcare, Germany).
The first experiment aimed at evaluating the
effectiveness of clinical spectral fat saturation in pure oil. Peanut oil was
chosen due to the resemblance of its NMR spectrum to the one of triglycerides5.
A multiecho 2D spoiled gradient-echo (GRE) sequence was used to acquire images
with 12 different TEs in each echo train. From those images, the fraction of
each fat peak was estimated using the IDEAL algorithm5 assuming no
water. Data were acquired with and without spectral fat saturation. The
spectral fat saturation pulse corresponds to the vendor “Standard” fat
saturation that is used in both diagnostic MRI and PRFS thermometry (Gaussian, bandwidth
197Hz, flip angle 110°). The fat spectrum is generally described as 6 main peaks6;
the main peak, methylene, represents 70% of the fat signal (Fig.
1). The diacyl peak is neglected (amplitude <1%).
The second experiment focused on evaluating
temperature errors in fat-suppressed PRFS-thermometry. Water-fat phantoms
(peanut oil)7 were elaborated in vials (diameter 3cm) with proton
density fat fraction (PDFF) of 10, 20 and 30 %, and a control sample with only
agar hydrogel. Reference temperature values were measured by fiber optic
temperature probes (Rugged Monitoring, Canada) placed in the center of each
vial. Phantoms were heated to 60°C in a water bath, then placed in a cane sugar
filled container in the MRI for the cooling-down experiment (Fig.
2).
The multiecho spoiled GRE sequence with spectral fat saturation was acquired
during 45 min providing images at different TEs (9.3 to 21.5 ms) from which PRFS
thermometry was obtained. B0-drift was corrected from two room-temperature agar
tubes placed on the sides. Because PRFS determines temperature variations, the
phase image stack can be inverted to mimic the process of heating8. Temperature
errors were calculated between PRFS measurements and optical probe data.
Results
In
the peanut-oil phantom, as expected, the fat peaks circling the water peak
(Olefinic, Glycerol) are weakly attenuated by spectral fat saturation (Table
1).
Between 14 and 26% of the 3 main fat peaks (methylene, α-olefinic and methyl)
persist despite spectral fat saturation.
Temperature
errors in the 4 water-fat phantoms are plotted against absolute temperature and
TE for the reversed cooling-down experiment (Figure
3).
In the pure agar sample, temperature errors vary between -1°C and +1.6°C. For
fat-water phantoms, the maximum positive temperature deviations recorded occur for
TE 9.3ms, and increase with PDFF (3.3°C, 6.3°C, and 9.9°C errors in phantoms
10%, 20%, and 30% oil, respectively). Similarly, maximum negative temperature
deviations occurred at TE 21.5ms (-1.8°C, -3.4°C, and -3.8°C observed for
phantoms 10%, 20%, and 30% oil, respectively). Minimum temperature errors were
found for TE 11.9ms (methylene peak, out-of-phase 37°C).Discussion
Residual
fat signal after spectral fat saturation leads to temperature errors during
thermal ablations depending on TE, temperature variation and fat fraction. Spectral
fat saturation targets the primary methylene fat peak and the 2 peaks
surrounding it. Failure to carefully select the TE according to the fat
saturation flip angle leads to residual fat signal, when the magnitude from the
2 fat peaks adjacent to water are not balanced by the voluntary remaining
signal from the 3 main fat peaks.Conclusion
Spectral
fat saturation cannot be 100% efficient, and the remaining fat signal affects fat
suppressed PRFS thermometry depending on fat fraction, TE and temperature
variation. Carefully selecting TE according to the spectral fat saturation
strategy mitigates this bias in fat-containing tissues.Acknowledgements
This
work has benefitted from funding of the FUI (Fonds Unique Interministériel, BPI
France) for the UFOGUIDE project, and the ANR (Agence Nationale de la
Recherche) French national program ‘Investissements d’Avenir’ for the
LABEX-CAMI (ANR-11-LABX-0004), the TechnoFUS joint laboratory project
(ANR-21-LCV3-0007-01), and the IHU Strasbourg (Institute of Image Guided
Surgery, ANR-10-IAHU-0002).References
1.
Ishihara
Y, Calderon A, Watanabe H et al. A Precise and Fast Temperature Mapping Using
Water Proton Chemical Shift. Magn. Reson. Med. 1995;34(6):814–823.
2.
Kuroda
K, Oshio K, Mulkern R. V et al. Optimization of Chemical Shift Selective
Suppression of fat. Magn. Reson. Med. 1998;40(4):505-510.
3.
Mao
J, Yan H, Brey W. W et al. Fat tissue and fat suppression. Magn Reson. Imaging.
1993;11(3):385-393.
4.
Chung
A. H, Hynynen K, Colucci V et al. Optimization of spoiled gradient-echo phase
imaging for in vivo localization of a focused ultrasound beam. Mag Reson Med.
1996;36(5):745-52.
5.
Yu
H, Shimakawa A, McKenzie C. A et al. Multiecho Water-Fat Separation and
Simultaneous R2* Estimation with Multifrequency Fat Spectrum Modeling. Magn. Reson.
Med. 2008;60(5):1122–1134.
6.
Hamilton G, Yokoo T, Bydder M et al. In vivo characterization of the liver fat 1H MR spectrum.
NMR Biomed. 2011;24(7).
7.
Bush
E. C, Gifford A, Coolbaugh C. L et al. Fat-Water Phantoms for Magnetic
Resonance Imaging Validation: A Flexible and Scalable Protocol. JoVE. 2018;No.
139, 57704.
8.
Rieke
V, Pauly K. B. MR thermometry. Journal of Magnetic Resonance Imaging. 2008; 27(2):376 –390.