Tingou Liang1, Junqi Yang2, Bing Keong Li3, Erping Li4, Wenwei Yu2, and Shao Ying Huang1
1Singapore University of Technology and Design, Singapore, Singapore, 2Chiba University, Chiba, Japan, 3Jiangsu LiCi Medical Device Co. Ltd, Lianyungang, China, 4Zhejiang University, Hangzhou, China
Synopsis
Keywords: Low-Field MRI, Magnets (B0)
Motivation: The current optimizations of permanent magnet array (PMA) designs are guided by checking field properties, not reflecting the quality of reconstructed images.
Goal(s): We aim to propose indicating parameters for the encoding capability of spatial encoding magnetic fields (SEMs) to guide PMA designs.
Approach: Local k-spaces are pushed to be point-wise. The coverage area of the point-wise k-spaces between the spokes of the maximum and minimum angles is calculated to indicate the encoding capability.
Results: The proposed method is fast, enlarging the optimization space, resulting in SEMs having superior encoding capability with stronger field, higher gradient, and lower NRMSE of the resultant images.
Impact: The proposed point-wise k-space evaluation enlarges the solution space
for the PMA optimizations thus significantly improves the performances of the
outcomes, i.e., SEMs having superior
encoding capability with higher field strength and higher gradient compared to
that from conventional approachs.
Introduction
Permanent magnet array (PMA) designs are important for
portable magnetic resonance imaging (MRI) due to low power consumption. The
current PMA designs with built-in gradients are guided by checking the quality
of simulated reconstruction images1. This time-consuming process is
usually simplified to checking target field properties (strength and patterns)
to save time1,2. However, field properties and image quality may
not always be correlated. Here, we proposed point-wise k-space, an intermediate
parameter to indicate the encoding capability of a spatial encoding magnetic
field (SEM) to guide PMA designs. It saves time with significantly improved
correlation to image quality.Methods
The local k-space is the spatial gradient of the accumulated
phase at a location3. Using SEMs, acquisitions are taken at the n-th
SEMs with $$$n_t$$$ time steps and a step size of $$$\Delta t$$$. The local k-space during the time
$$$t=n_t\cdot\Delta t$$$ is expressed as $$\vec{k}(\vec{r},t,n)=\gamma\sum_0^{n_t}{\nabla\vec{B}_{SEM}(\vec{r},n)\cdot\Delta t} (1)$$ where $$$\gamma$$$ is the gyromagnetic constant, $$$\nabla\vec{B}_{SEM}(\vec{r},n)$$$ is the n-th SEM. When rotational SEMs (rSEMs) are used, the n-th SEM corresponds to an
angle $$$\theta$$$, $$$\vec{k}(\vec{r},t,n)=\vec{k}(\vec{r},t,\theta)$$$.
Assuming a fixed $$$\Delta t$$$, we can deduce the relationship
between the local k-space and the SEM as follows: $$\vec{k}(\vec{r},t,\theta)\propto\nabla\vec{B}_{SEM}(\vec{r},\theta)\cdot\Delta t(2)$$ Therefore, the gradient $$$\vec{G}(\vec{r},\theta)=\nabla\vec{B}_{SEM}(\vec{r},\theta)$$$
is proportional to the local k-space value. For evaluation using local k-space, usually the field-of-view (FOV) is
split into a few blocks, and the local k-space at the center is plotted to
represent the whole block3. Fig.1(b) shows 3-by-3 local k-spaces of the SEM in Fig.1(a) when it rotates. For
each plot, a spoke contains the signal points at one angle. Larger coverage of
the spokes (higher gradients and wider spreading) indicates better encoding
capability. This approach can only tell the encoding capability of an SEM when it
changes slowly, i.e., the block shares similar fields as the center.
To evaluate SEMs accurately and efficiently, local k-spaces are pushed
to be point-wise. An intermediate parameter, $$$k_p$$$, is proposed to evaluate the
encoding capability of the field. For rSEMs, $$$k_p$$$ is defined as the multiplication of the arc-shape
area covered by the spokes (Fig.1(c)) and a penalty for gradient ununiformity: $$k_p(\vec{r})=\gamma\Delta t A_\text{arc}(1-\frac{std(G(\vec{r}))}{\bar{G}(\vec{r})}) (3)$$ where $$$A_\text{arc}=\frac{1}{2}(\varphi_{max}^k(\vec{r})-\varphi_{min}^k(\vec{r}))\bar{G}(\vec{r})$$$, $$$\varphi_{max}^k$$$ and $$$\varphi_{min}^k$$$ are the extrema
k-space angles among all rotation angles, $$$\bar{G}$$$ and $$$std(G)$$$ are the average and standard deviation of the
gradients over different angles at $$$\vec{r}$$$. Furthermore, for an
rSEM, the point-wise k-space is axial-symmetric. Therefore, only those points
along a radius are evaluated, further accelerating the evaluation process.
The proposed parameter was applied to PMA optimizations.
A cost function can be formulated as follows, $$L_k=min[\frac{std(k_p)}{\bar{k}_p}+\frac{std(B_{SEM})}{\bar{B}_{SEM}}] (4)$$In (4), both terms have a range of [0,1], with balanced emphasis
on both the encoding capability and SEM inhomogeneity.Results & Discussions
A PMA for wrist imaging (FOV: 80mm DSV) was
optimized using genetic algorithm (GA) and the cost function in (4) to show the
application of the proposed method. Another two traditional ways of defining
the cost functions were implemented for comparison. One using the properties of
magnetic field, $$$F_{field}=BW-R^2$$$, where $$$BW$$$ indicates field inhomogeneity, and $$$R^2$$$ is the determination coefficient indicating
the field linearity; the other using image quality, $$$F_{image}=\text{NRMSE}-\text{SSIM}+BW$$$, involving the
normalized root-mean-square error (NRMSE) and structural similarity index
(SSIM).
Fig.2 shows the PMA. It consists of a base array
(grey), which is an inward-outward (IO) ring4,5, and a symmetric double
offset ring (green and blue) to provide gradient in the x-direction1.
The variables for optimization are labeled in Fig.2(a)-(c) and tabulated in Fig.2(d).
Using GA, all methods have 50 iterations with a population of 50. The number of
rotation angles was N=144. Fig.3 shows the
optimized SEMs and the simulated images.
Table 2 lists the optimization
results for all methods. The proposed method and field-method are comparable
for the time, while the image-method is 68 times slower. The last two columns
show the NRMSE and SSIM of the simulated images in Fig.3. For the two fast
methods, the proposed method outputs a design with improved performances than
the field-method in terms of field strength, gradients, and the
resultant NRMSE with slightly lower SSIM. To check the consistency of the
comparison, ten trials were performed for k-space method and field-method with
performances plotted in Fig.5. Good consistency is observed. Conclusion
In this abstract, we propose a point-wise k-space
evaluation method to accelerate and improve optimizations of PMA designs
considering the SEMs encoding capability. Compared to the traditional method
guided by field properties, the proposed method enlarges the optimization
space, resulting in SEMs having superior encoding capability with stronger
field, higher gradient, and lower NRMSE of the resultant images.Acknowledgements
No acknowledgement found.References
1. Ting-Ou Liang, Yan Hao Koh, Tie Qiu, Erping Li, Wenwei Yu, Shao Ying
Huang, High-Performance Permanent Magnet Array Design by a Fast Genetic
Algorithm (GA)-based Optimization for Low-Field Portable MRI, Journal of
Magnetic Resonance, 2022, 107309, ISSN 1090-7807, https://doi.org/10.1016/j.jmr.2022.107309.
2. Z. H. Ren, J. Gong and S. Y. Huang, "An
Irregular-Shaped Inward-Outward Ring-Pair Magnet Array With a Monotonic Field
Gradient for 2D Head Imaging in Low-Field Portable MRI," in IEEE Access,
vol. 7, pp. 48715-48724, 2019, doi: 10.1109/ACCESS.2019.2909834.
3. D. Gallichan, C. A. Cocosco, A. Dewdney, G.
Schultz, A. Welz, J. Hennig, and M. Zaitsev, ‘‘Simultaneously driven linear and
nonlinear spatial
encoding fields in MRI,’’ Magn. Reson. Med., vol.
65, no. 3, pp. 702–714,
2011.
4. G. Aubert, “Permanent magnet for nuclear
magnetic resonance imaging equipment,” July 26, 1994. US Patent 5,332,971
5. G. Miyajima, Cylindrical permanent magnet
apparatus, Japanese Patent
JPS60210804A (Oct.23, formally Apr. 4, 1984 1985).