2186

Non-Linear Echo Combination Allows for Reduced Scan Times in T2* Weighted Imaging at 1.5T
Brian L Burns1
1GE Healthcare, Menlo Park, CA, United States

Synopsis

Keywords: Blood Vessels, Susceptibility, echo combination, multi-echo

Motivation: Scan times for multi-echo T2*w sequences are too long at 1.5T.

Goal(s): Shorten scan times by reducing the number of echoes needed for a specific contrast.

Approach: During this ISMRM, we introduced a novel echo combination method that increases contrast over the standard Root Sum-of-Squares (RSOS) technique. We applied this method to 7 and 6 echo 3D SPGR protocols and compared the results to a 9 echo protocol that used RSOS.

Results: Contrast was comparable or better in the 6 and 7 echo protocols using our technique, compared to the 9 echo protocol with RSOS, resulting in a scan time reduction of 23-26%.

Impact: Our echo combination technique allows fewer echoes in multi-echo T2*w imaging to match the contrast of higher echo protocols resulting in a 23-26% scan time reduction. This alleviates one of the primary concerns clinicians have using T2*w imaging at 1.5T.

Introduction

T2* weighted contrast has become a core part of brain MRI exams and is routinely used by clinicians to aid in the diagnosis of diseases such as MS1 or Alzheimer’s2, and it is a recommended part of a standard acute ischemic stroke triage protocol3. However, because T2* values in the brain are 50-80% longer at 1.5T than they are at 3T4, scan times at 1.5T can be prohibitively long to acquire sufficient contrast. This results in compromises that must be made on spatial resolution, brain coverage, or high parallel acceleration rates to bring the scan times down to clinically acceptable levels.
In this work, we apply a novel echo combination method introduced at this ISMRM5 to reduce the scan time of multi-echo T2* weighted acquisitions by reducing the number of echoes needed to achieve the same level of contrast as a longer echo protocol that uses standard Root-Sum-of-Squares (RSOS) echo combination. We show that scan times can be reduced by 23-26% using this echo combination strategy without any loss in contrast or vessel conspicuity.

Methods

We scanned one healthy volunteer on a 1.5T MR450 system (GE Healthcare, Chicago, Il). Images were acquired using a fully-flow compensated 3D multi-echo SPGR sequence with unipolar echoes for 9, 7, and 6 echo protocols at different readout bandwidths (rBW). Data was acquired according to the table below:
Num Echoes
rBW (Hz/pixel)
TEs(ms)
Scan Time
9
416
26,30.2,34.3,38.5,42.7,46.8,51.0,55.2,59.3
7:57
7
416
26,30.2,34.3,38.5,42.7,46.8,51.0
6:20
7
275
26,31.2,36.4,41.6,46.8,52.0,57.2
6:04
6
416
23.9,28.1,32.3,36.4,40.6,44.7
5:32
6
233
23.1,28.9,34.7,40.5,46.2,52.0
5:52
Flip angle was adjusted to be the Ernest angle for a given TR and T1=840ms. The spatial resolution and brain coverage was matched across all acquisitions. Magnitude and phase images were reconstructed separately using the vendor reconstruction pipeline which was modified to combine magnitude echoes using the following two methods5:
$$RSOS(echo):=\sqrt{\sum (echo_n\cdot echo_n)}\qquad\quad(1)$$
$$TE\cdot ln(echo):=exp^{\sum\big(\frac{TE_n}{\sum TE_n}\cdot ln(echo_n)\big)}\qquad\quad(2)$$
SWI images were created from these echo combined magnitude images5.

Results

Figure 1 shows a 10mm MinIP of echo combined magnitude images using either RSOS or the proposed TE*ln(echo) combination method for the 9 and 7 echo protocols. Comparing the RSOS results for the 9 echo protocol in Fig 1A to the 7 echo protocol in Fig 1B we see that reducing the number of echoes at the same rBW results in an expected loss of contrast and small vessel conspicuity. The 7 echo protocol with reduced rBW in Fig 1C shows some of the SNR and contrast recovered, but smaller vessels are still less conspicuous than the 9 echo protocol in Fig 1A. Figs 1D and 1E show the results for the TE*ln(echo) combined 7 echo protocols. The proposed echo combination method produces more contrast with higher vessel conspicuity than RSOS even when it is applied to a higher echo protocol.
Figures 2 and 3 show 10mm MinIP magnitude image comparisons between RSOS and TE*ln(echo) combined 9, 7, and 6 echo protocols, with Figure 3 showing a zoomed in view of Figure 2. As can be seen, the 7 and 6 echo protocols in Figs 2D and 2E reconstructed with TE*ln(echo) echo combination have superior vessel conspicuity and overall contrast compared to the RSOS echo combined data sets in Figs 2A, 2B, and 2C.
Figure 4 shows the 10 mm MinIP SWI images generated from the magnitude images in Figs 2 and 3. The contrast benefits from the TE*ln(echo) echo combination are still evident in Figs 4D and 4E, albeit not as pronounced as the magnitude image results.
Line profiles across the two internal cerebral veins in the 10mm MinIP SWI from Fig 4 are shown in Fig 5. Comparing the width and depth of the TE*ln(echo) echo combined line profiles to RSOS, we can see the depth and width are increased by ~10-15% because of the increased influence on contrast from later echoes, where increased T2* signal dephasing makes large veins wider. This change in line profile translates to blacker cerebral veins with sharper edges using our method.

Discussion/Conclusion

These results confirm that reducing the rBW and number of echoes in a 1.5T multi-echo acquisition can improve SNR and CNR. However, the contrast and vessel conspicuity will not match a higher echo protocol if RSOS echo combination is used. Our proposed echo combination method can recover, if not surpass, the contrast lost by shortening the echo train, resulting in shortened scan times at 1.5T of up to 23-26%.

Acknowledgements

The author would like to thank Dr Janine Lupo and Dr Suchandrima Banerjee for their valuable feedback.

References

  1. Sati P, et al. The central vein sign and its clinical evaluation for the diagnosis of multiple sclerosis: a consensus statement from the North American Imaging in Multiple Sclerosis Cooperative. Nat Rev Neurol. 2016;12:714-722
  2. Cogswell PM, et al. Amyloid-Related Imaging Abnormalities with Emerging Alzheimer Disease Therapeutics: Detection and Reporting Recommendations for Clinical Practice. AJNR. 2022; DOI: 10.3174/ajnr.A7586
  3. Alberts M, et al. Recommendations for Comprehensive Stroke Centers: A Consensus Statement From the Brain Attack Coalition. Stroke. 2005;36:1597–1616
  4. Peters A, et al. T2* measurements in human brain at 1.5, 3 and 7 T. Mag Reason Imag. 2007;25:748-753
  5. Burns B, Improved Susceptibility Contrast in Multi-Echo T2* Imaging Through Non-Linear Echo Combination, Proc. Intl. Soc. Mag. Reson. Med. 2024
  6. Haacke EM, et al. Susceptibility Weighted Imaging (SWI). Mag Reson Med. 2004;52:612–618

Figures

Figure 1: Comparison of the 10mm MinIP of echo combined magnitude images using either RSOS (A-C) or the proposed TE*ln(echo) (D-E) combination method for the 9 and 7 echo protocols at different rBW. (A,B,D) Used 416Hz/pixel and (C,E) Used 275 Hz/pixel. The 275Hz/pixel 7 echo image from the TE*ln(echo) method have equal to or better contrast than the 416Hz/pixel 9 echo image that used RSOS.

Figure 2: Comparison of the 10mm MinIP of echo combined magnitude images using either RSOS (A-C) or the proposed TE*ln(echo) (D-E) combination method for the 9, 7, and 6 echo protocols where the rBW was reduced in the 6 and 7 echo protocols to recover SNR. Both of the reduced echo images using the TE*ln(echo) method have equal to or better contrast than the 9 echo image that used RSOS.

Figure 3: Zoomed in view of the same images from Figure 2, showing the improved vessel conspicuity of the TE*ln(echo) method (D,E) in the reduced echo protocols compared to the 9 echo protocol that used RSOS (A-C).

Figure 4: SWI 10mm MinIPs of the echo combined magnitude images in Figure 2, showing the improved vessel conspicuity from the TE*ln(echo) method in D and E compared to RSOS in A-C.

Figure 5: Zoomed in view of the same SWI images from Figure 4, showing the line profiles across the two internal cerebral veins in each image. The 7 and 6 echo protocols using the TE*ln(echo) method (dotted lines) have wider, deeper line profiles than the other images, indicating an increase in T2* contrast.

Proc. Intl. Soc. Mag. Reson. Med. 32 (2024)
2186
DOI: https://doi.org/10.58530/2024/2186