Leyre Garcia-Ruiz1,2, Rebeca Echeverria-Chasco1,2, Veronica Aramendia-Vidaurreta1,2, Jose María Mora-Gutiérrez3, Nuria Garcia-Fernandez2,3, Marta Vidorreta4, Gorka Bastarrika1, and María A. Fernández-Seara1,2
1Department of Radiology, Clínica Universidad de Navarra, Pamplona, Navarra, Spain, 2Instituto de Investigación Sanitaria de Navarra, IDISNA, Pamplona, Navarra, Spain, 3Department of Nephrology, Clínica Universidad de Navarra, Pamplona, Navarra, Spain, 4Siemens Healthcare, Madrid, Spain
Synopsis
Keywords: Arterial Spin Labelling, Perfusion, T1 mappinng
Motivation: Pseudo-continuous Arterial Spin Labeling (PCASL) MRI is a promising technique for assessing renal perfusion. Previous studies have been carried out at 1.5T and 3T field strengths, but a comparison of data acquired at both field strengths has never been performed.
Goal(s): The aim of this work was to assess the effects of field strength on the measured perfusion signal and to evaluate the reproducibility of renal PCASL at both fields
Approach: Healthy volunteers underwent repeated scans on both scanners, spaced a week apart.
Results: The results showed excellent reproducibility and minimal differences in RBF, ATT and T1 values between visits and field strengths.
Impact: The similarity in the study results across both field
strengths underscore the potential for expanding PCASL-based renal perfusion
assessment, increasing the technique accessibility, and widening diagnostic
capabilities.
INTRODUCTION
Arterial
Spin Labeling (ASL) is a promising technique for renal perfusion assessment. Different imaging techniques at independent magnetic filed strengths (1.5T and 3T) have been used to assess renal perfusion1. However, there is a notable absence of comparative research at both field strengths. Therefore, this study aims to compare the outcomes and reproducibility of renal
ASL at both 1.5T and 3T. These findings might contribute to ASL reliability in assessing renal perfusion, supporting its potential integration into clinical practice. METHODS
Subjects and Study
Design:
Written informed consent was obtained from subjects. Urine and blood were obtained to measure the
estimated glomerular filtration rate (eGFR) with the CKD-EPI equation2.
Eight healthy volunteers (eGFR>90ml/min/1.73m2, mean age±standard
deviation (SD)=32±7.9 years) were scanned twice within the same day on 1.5T
Aera and 3T Skyra MRI scanners (Siemens Healthcare), using 18-channel
body and spine coils. The same protocol was repeated with a median of seven days (range=1-17 days).
MRI Protocol:
Pseudo-continuous ASL (PCASL) sequence employing
six different post-labeling delays (PLDs), with background suppression (BS) and
SE-EPI readout. Pre-saturation pulses were applied before the labeling pulses. BS
was achieved using a slice-selective FOCI pulse before labeling and two non-selective
tangent/hyperbolic tangent pulses after labeling. Pulse timings
were optimized for each PLD to suppress static tissue signal to 15%. This
optimization was performed for each field strength considering the expected
kidney T1 values. PCASL were acquired with and without BS pulses
for the PLD=1.3s, to estimate the efficiency of the BS pulses.
T1 Inversion Recovery sequence
employing 16 inversion times and SE-EPI readout. Other sequence parameters are described in Table 1.
Image processing:
All the images underwent registration in Elastix3
using a groupwise registration method. Subsequent analysis was conducted
using custom scripts in Matlab (Mathworks). T1 maps were generated by fitting
the signal to the inversion-recovery equation [1]:
$$S_k=M_0(1-2e^{-TI_k/T1)}+M_0e^{-TR/T1})$$
To compute Renal Blood Flow (RBF) and Arterial
Transit Time (ATT) maps, a voxel-by-voxel approach was employed, fitting the
data to the one-compartment Buxton4 model, with the assumption that
by the time of image acquisition, labeled spins remained in the blood [2]:
$$(SI_C-SI_L)_t=\begin{cases}0&\text{if }0<\tau_t+\text{PLD}_t<\text{ATT}\\\frac{2\alpha\text{T1}_b}{6000\lambda}M_0RBFe^{-\frac{\text{ATT}_t}{\text{T1}_b}}(1-e^{-\frac{(\tau_t+\text{PLD}_t-\text{ATT})}{\text{T1}_b}})&\text{if }\text{ATT}<\tau_t+\text{PLD}_t<\tau_t+\text{ATT}\\\frac{2\alpha\text{T1}_b}{6000\lambda}M_0RBFe^{-\frac{\text{PLD}_t}{\text{T1}_b}}(1-e^{-\frac{\tau_t}{\text{T1}_b}})&\text{if } \tau_t+\text{ATT}<\tau_t+\text{PLD}_t \end{cases}$$
where $$$(SI_C - SI_L)$$$ is the signal difference between control and label images for each PLD, $$$\alpha=0.75\times0.9\times0.9$$$ is the
labeling efficiency, considering PCASL efficiency and the effects of 2 BS
pulses, $$$\lambda=0.9$$$mL/g is the
tissue-blood partition coefficient, $$${T1_{b}}=1.48$$$s and $$$1.65$$$s are the
longitudinal relaxation times of blood at 1.5T and 3T, respectively and $$$\tau$$$ is the labelling duration. Regions of Interest (ROIs) in the cortex and
medulla were manually drawn in the T1 maps and used to obtain mean values by averaging
both kidneys.
Statistical analysis:
Differences between field strengths, were tested
by paired t-test. Reproducibility was assessed by Bland-Altman plots and by the inter-session within subject coefficient
of variation (wsCV), calculated as the root mean square of the squared ratio of the SD
to the mean for each subject of the two repeated measurements5.RESULTS AND DISCUSSION
RBF, ATT, and T1 maps for a
representative subject are depicted in Figure 1.Table
2 provides averaged cortical and medullary values across the group, for each visit and field strength. T1
measurements aligned with literature values6, and as expected, T1
values were longer at 3T compared to 1.5T. Inter-session wsCVs for T1 in the cortex and medulla were low, indicating
excellent reproducibility (<2% for 1.5T and <1.2% for 3T). RBF and ATT measurements obtained during both
visits and at both field strengths yielded remarkably similar results,
consistent with the literature1,6-7, with no statistically
significant differences between field strengths, except for the RBF values in
the medulla (p<0.001), which were higher at 1.5T. It is worth noting that
cortical RBF values were also 10% higher at 1.5T (although the differences did
not reach statistical significance). These data were also reproducible, with
cortical and medullary wsCVs<11% for RBF and ATT at both 1.5T and 3T, except
for the medulla at 3T (wsCV=18.3%). Bland-Altman plots (Figure 2) showed good
reproducibility between visits for both the cortex and medulla. Perfusion weighted signal (PWS), temporal
Signal-to-Noise Ratio (tSNR), and BS efficiency for PLD=1.3s were
also obtained (Table 3). Results showed a higher BS efficiency in the 1.5T
scans, which could be attributed to a more uniform magnetic field, which partly explain the observed differences in RBF, that could
also be due to a higher labeling efficiency, at 1.5T compared to 3T. Although
this parameter was not measured.CONCLUSION
Similarity in results across both field
strengths highlights the potential for wider adoption of PCASL-based renal
perfusion assessment and support its use in multicenter clinical studies.Acknowledgements
Leyre Garcia-Ruiz received PhD grant support from Asociación de Amigos de la Universidad de Navarra. This project was supported by the
Government of Navarra under the frame of ERA PerMed (ERAPERMED2020-326 -
RESPECT) and under grant: PC181-182 RM-RENAL.
References
1. Odudu A, Nery F, Harteveld AA,
Evans RG, Pendse D, Buchanan CE, Francis ST, Fernández-Seara MA. Arterial spin
labelling MRI to measure renal perfusion: a systematic review and statement
paper. Nephrol Dial Transplant. 2018 Sep 1;33(suppl_2):ii15-ii21. doi:
10.1093/ndt/gfy180.
2. Levey AS, Stevens LA, Schmid
CH, Zhang YL, Castro AF 3rd, Feldman HI, Kusek JW, Eggers P, Van Lente F,
Greene T, Coresh J; CKD-EPI (Chronic Kidney Disease Epidemiology
Collaboration). A new equation to estimate glomerular filtration rate. Ann
Intern Med. 2009 May 5;150(9):604-12. doi:
10.7326/0003-4819-150-9-200905050-00006.
3. Klein S, Staring M, Murphy K, Viergever MA,
Pluim JPW: elastix: A Toolbox for Intensity-Based Medical Image Registration. IEEE Trans Med Imaging 2010; 29:196–205.
4. Buxton RB, Frank LR, Wong EC,
Siewert B, Warach S, Edelman RR. A general kinetic model for quantitative
perfusion imaging with arterial spin labeling. Magn Reson Med. 1998 Sep;40(3):383-96. doi: 10.1002/mrm.1910400308.
5. Bland JM, Altman DG.
Measurement error. BMJ. 1996 Sep 21;313(7059):744.
6. Cox EF, Buchanan CE, Bradley
CR, Prestwich B, Mahmoud H, Taal M, Selby NM, Francis ST. Multiparametric Renal
Magnetic Resonance Imaging: Validation, Interventions, and Alterations in
Chronic Kidney Disease. Front Physiol. 2017 Sep 14;8:696. doi: 10.3389/fphys.2017.00696.
7. Kim DW, Shim WH, Yoon SK, Oh
JY, Kim JK, Jung H, Matsuda T, Kim D. Measurement of arterial transit time and
renal blood flow using pseudocontinuous ASL MRI with multiple post-labeling
delays: Feasibility, reproducibility, and variation. J Magn Reson Imaging. 2017
Sep;46(3):813-819. doi: 10.1002/jmri.25634.