Dinesh K Deelchand1 and Gülin Öz1
1Radiology, University of Minnesota, Minneapolis, MN, United States
Synopsis
Keywords: Spectroscopy, Spectroscopy, LCModel
Motivation: The actual B0 resonance field of 3T scanners is different between the three major MR vendors.
Goal(s): Here, we report the effect on metabolite quantification of using basis sets at different B0 fields than the B0 frequency of the measured in vivo MRS data.
Approach: Basis sets were simulated at thirteen different B0 fields. Semi-LASER MRS data measured from the posterior cingulate cortex at 3T were fitted using LCModel.
Results: Results show that biases in metabolite concentrations were within 1% when the basis set B0 frequencies were within ±0.5 MHz from the actual scanner frequency.
Impact: A single sequence-specific
basis set can be used to analyze harmonized spectroscopic data collected on
clinical 3T scanners from different vendors that operate at different
frequencies.
Introduction
Following the international
magnetic resonance spectroscopy (MRS) consensus group recommendations1-2,
the proton semi-LASER sequence was successfully standardized across the three
major MR vendors3 (GE, Philips, and Siemens). All localization RF
pulses patterns, durations and inter-pulse delays were matched, in addition to utilizing
the same VAPOR technique for water suppression with interleaved OVS pulses.
This has resulted in comparable spectral pattern between vendors in several
brain regions at multiple sites. This should potentially allow pooling of
multi-site/multi-vendor spectroscopic data. However, the actual B0
resonance field of 3T scanners is different between the three major MR vendors
and varies between 123.2 to 127.8 MHz. As such, previous multiple-vendor and
multi-sites studies3-5 have used simulated basis sets specific to
each scanner when quantifying brain metabolites.
So far, it is unclear
whether we need basis sets simulated at slightly different frequencies for each
specific magnetic field strength where the standardized semi-LASER data was collected.
Therefore, the aim of the current study was to determine the effect on metabolite
quantification of using basis sets at different B0 fields from the
measured in vivo data.Methods
Five healthy volunteers
participated in the study after giving written informed consent approved by the
Institutional Review Board at the University of Minnesota. Studies were
performed with a 3T whole-body Siemens scanner. The standard body RF coil was
used for radiofrequency transmission and the 32-channel phased-array Siemens
head coil was used for signal reception. T1-weighted MPRAGE images
were used to position the VOI in the posterior cingulate cortex (2×2×2 cm3).
After FAST(EST)MAP shimming, metabolite and water reference spectra were
acquired using semi-LASER (TR/TE = 5000/28 ms, 64
transients). Note that the 3T MRS data used in this study were collected for a
study6 focused on test-retest reproducibility of spectroscopic data.
The basis spectra
consist of 19 metabolites: alanine, ascorbate, aspartate, creatine (Cr),
γ-aminobutyric acid, glucose, glutamate (Glu), glutamine (Gln), glutathione,
glycerophosphorylcholine, myo-inositol
(Ins), scyllo-inositol, lactate, N-acetylaspartate (NAA), N-acetylaspartylglutamate (NAAG),
phosphocreatine (PCr), phosphorylcholine, phosphorylethanolamine, and taurine. Thirteen
basis sets were simulated ranging from -1 to 10 MHz from the nominal B0
frequency of the measured MRS data which was at 123.245 MHz. A measured macromolecule
(MM) spectrum was also included. The MM data
was resampled at the different frequencies and used in the appropriate basis
sets.
All spectra were
processed with MRspa7 and resulting averaged spectra were fitted
using LCModel v6.3-1R8 with the 13 different basis sets. The fitted
region was from 0.5 to 4.2 ppm and no baseline correction or filtering was applied
to the processed data. Differences in concentrations relative to the fitting
done with the basis set generated at the actual scanner frequency (123.245 MHz)
are reported for the six major metabolites, i.e. tNAA(=NAA+NAAG), tCr(=Cr+PCr), tCho, Glu, Ins and Glx(=Glu+Gln).Results
Figure
1 compares the simulated spectra of glutamate and myo-inositol at three different B0 frequencies. When the
difference in B0 was within 1 MHz, the spectral difference was negligible.
However, small differences in spectral pattern started to appear when the
difference was 10MHz.
High-quality
sLASER data were measured in all subjects with no lipid or baseline issues (Figure
2). LCModel outputs from one subject using basis set at -1, 0 and 10 MHz from
the actual B0 frequency of the MRS data are illustrated in Figure 3.
No obvious difference in the fitted results was apparent with comparable
residual and baseline in all cases even when the basis set frequency was 10 MHz
off.
Figure
4 shows the %difference in metabolite concentrations when using basis sets with
varying frequencies. When the basis set was within ±0.5MHz from the measured
baseline B0 frequency, the changes were within 1% for the six
metabolites. Above this frequency, the changes were >2%, except for tNAA.Discussion and Conclusion
This study shows that the
change in concentrations was less than 1% for the six metabolites (tNAA, tCr,
tCho, Glu, Glx and mIns) when the B0 frequencies of the basis sets were
within ±0.5
MHz of the actual B0 frequency of the measured MRS data. This
observation might be explained by the fact that the line-shape regularization function
in LCModel is very flexible8 such that this takes into account small
changes in spectral pattern. Furthermore, the changes in concentration observed
within the ±0.5
MHz are smaller than the test-retest CVs previously reported for these
metabolites6.
In conclusion, this
study shows that it is possible to use one sequence specific basis set to fit harmonized
MRS data measured at 3T scanners from different vendors since the difference in
B0 frequency between scanners is within 0.5 MHz.Acknowledgements
This work was supported by funding from the
National Institutes of Health (NIH) P41 EB027061, P30 NS076408, R01 NS080816
and R01 EB030000.References
1.
Öz et al. Clinical proton MR spectroscopy in central nervous
system disorders Radiology. 2014 Mar;270(3):658-79.
2.
Wilson et al. Methodological consensus on clinical proton MRS of
the brain: Review and recommendations Magn Reson Med. 2019 Aug;82(2):527-550.
3. Deelchand et al.
Across-vendor standardization of semi-LASER for single-voxel MRS at 3T NMR
Biomed. 2021 May;34(5).
4. Craven et al. Comparison
of seven modelling algorithms for γ-aminobutyric acid-edited proton magnetic
resonance spectroscopy NMR Biomed. 2022 Jul;35(7).
5. Joers et al.
Multi-site/Multi-vendor reproducibility of advanced MRS at 3T in a clinical
cohort Proc. Intl. Soc. Mag. Reson. Med. 28 (2020) 2920.
6. Terpstra et al.
Test-retest reproducibility of neurochemical profiles with short-echo,
single-voxel MR spectroscopy at 3T and 7T Magn Reson Med. 2016
Oct;76(4):1083-91.
7. Deelchand DK. MRspa:
Magnetic Resonance signal processing and analysis. https://www.cmrr.umn.edu/downloads/mrspa/
8.
Provencher SW. Estimation of metabolite concentrations from
localized in vivo proton NMR spectra. Magn Reson Med. 1993;30:672-679.