Yunkun Zhao1, Aditya Ashok Bhosale1, and Xiaoliang Zhang1,2
1Department of Biomedical Engineering, State University of New York at Buffalo, Buffalo, NY, United States, 2Department of Electrical Engineering, State University of New York at Buffalo, Buffalo, NY, United States
Synopsis
Keywords: Multimodal, Non-Array RF Coils, Antennas & Waveguides
Motivation: By the need to address the limitations of low-field MRI, such as low SNR and insensitive frequency tuning.
Goal(s): We proposed a design of multimodal surface coil that can enhance the B1 efficiency while reducing the associated electric field.
Approach: Through both electromagnetic simulation and bench testing, the performance of this novel coil design technique was evaluated and compared with the conventional surface coil and solenoid coil.
Results: The results affirmed the superior performance of the proposed multimodal surface coil design in terms of B1 efficiency and frequency tuning over the surface coil and solenoid coil at the low field of 0.5T.
Impact: The proposed multimodal
surface coil design improves B1 efficiency compared with the conventional surface
coil and solenoid coil at 0.5T. It also mitigates frequency
tuning challenges and has the potential to enhance the quality of low-field MRI in clinical diagnosis.
Introduction
Low-field MRI
offers several advantages, including lower operating costs, reduced power
requirements, and increased safety for patients with certain medical conditions
or implants. However, low-field MR suffers from an intrinsically low
signal-to-noise ratio (SNR) and thus low image resolution. It is known that MR
SNR is linearly proportional to the B1 efficiency of RF coils (1-3). In this work,
we introduce the multimodal surface coil, an innovative design that provides significantly
improved B1 fields over the conventional surface coil and has the potential to enhance
MR SNR. The design is based on a set of stacked resonators (4) that are
electromagnetically coupled and form a multimodal resonator. The proposed
multimodal surface coil design also provides a low frequency tuning capability
and alleviates the low frequency tuning challenges encountered in low field MR
imaging. The design has been validated through full-wave electromagnetic
simulations and standard RF bench tests and measurements.Methods
Figure 1A displays
the simulation model of a multimodal surface coil. The multimodal surface coil
comprised of seven coil loops or resonators, with six identical squared coils
with 10 cm side length each equipped with a tuning capacitor to tune the coil
to 42 MHz, and the middle coil contains an impedance matching circuit for driving
the multimodal surface coil. A 5 mm spacing between coils enhances mutual
inductive coupling, with the entire stack reaching a 3 cm height. Figure 1B and
1C show two comparative setups - a seven-turn solenoid and a conventional surface
coil, each defined by a 10 cm length. The solenoid has a 3 cm height and a 5 mm
spacing between each turn. All designs were built using a 6.35 mm wide copper
sheet conductor and are tuned to 21.3 MHz and impedance matched to 50 ohms. Performance
assessments of the multimodal surface coil involved analyzing scattering
parameters, B1 efficiency, E field efficiency, and B1/E field efficiency ratio
field distribution plots, with all electromagnetic field plots normalized to 1
W accepted power. All simulation results generated by EM simulation software CST
Studio Suite. Figures 2A, 2B, and 2C show the bench test model of multimodal surface
coil, solenoid, and conventional surface coils. The bench test models have the
same dimensions and resonant frequencies as the simulation model. The coils
were built with 6.35 mm wide copper tape and on a 3D printed polylactide
structure. Results of bench test models were obtained using a vector network analyzer
based 3-D positioning magnetic and electric field mapping system. Results
Figure 2A
illustrates the simulated scattering parameters against the frequency of the multimodal
surface coil. The figure reveals that strong coupling occurs between the coils,
leading to the emergence of four distinct split resonant peaks with lowest peak
at 21 MHz for imaging, while each individual coil resonates at 42 MHz. Figure
2B displays the B field efficiency maps in the Y-Z, X-Z, and X-Y planes at the
lowest frequency within the phantom, produced by the multimodal surface coils. Figure
2C shows the 1-D field strength plot along the dashed line in Figure 2B. The
simulation result shows the multimodal surface coil has stronger B field
efficiency comparing with the conventional surface coil and similar B1 field
efficiency comparing with the solenoid coil. Figure 3 shows that the simulated
E field efficiency and B1/E
field efficiency ratio in X-Z planes for all three setups. With
same amount of accepted power, multimodal surface coils show significantly
lower E field comparing with the solenoid coil, which leads to a much lower
noise in SNR calculation.
Figure 4A shows the
bench test scattering parameters for the multimodal surface coil and Figure 4B
shows the measured B field efficiency map for all three setups in different
planes. Figure 5 shows the measured E field efficiency and B1/E field efficiency ratio map
for all three setups in X-Z plane. In the bench test, the measured results are
in accordance with the simulated results, indicating that the multimodal
surface coil has advantages in B and E field efficiency when comparing with
conventional surface coil and solenoid coil. This would lead to an improved
overall SNR for low field MR imaging. Conclusion
In conclusion, the
proposed multimodal surface RF coil has demonstrated significant improvements in
B1 efficiency over conventional surface coils, ultimately leading to an
improved SNR for low-field MRI. The proposed multimodal surface coil design
also helps to achieve low frequency tuning which is technically challenging at
low magnetic fields. This multimodal surface coil technique can be possibly
used to design high frequency small animal RF coils.
Acknowledgements
This work is
supported in part by the NIH under a BRP grant U01 EB023829 and by the State
University of New York (SUNY) under SUNY Empire Innovation Professorship Award.References
1) Hoult, D.I. The
signal-to-noise ratio of the nuclear magnetic resonance experiment. J.Magn
Reson. 24, 71-85 (1976)
2) Pang, Y, et al. Parallel traveling wave MRI: A feasibility
study. Magn Reson Med. 67, 965-978 (2012)
3) Wang, C, et al. A practical multinuclear transceiver
volume coil for in vivo MRI/MRS at 7 T. Magn Reson Imag. 30, 78-84 (2012)
4) Zhao, Y, et al. A
coupled planar RF array for ultrahigh field MR imaging. ISMRM 2023, 3910