Ernesto Gomez Tamm1, Andreas Hodul2, Markus Ornter3, Zacharias Chalampalakis3, Quang Nguyen2, Onisim Soanca1, Vivian Janicaud1, Elmar Laistler1, Ivo Rausch3, and Roberta Frass-Kriegl1
1High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria, 2Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria, 3QIMP Team, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria
Synopsis
Keywords: Phantoms, Phantoms, Motion
Motivation: The development of motion tracking and correction methods requires suitable motion phantoms.
Goal(s): The design, fabrication, assembly and testing of a modular dynamic torso phantom.
Approach: The phantom consists of a saline-filled acrylic shell, with dimensions comparable to a human torso and three distinct interior motion modules. Two linear piezo stages are used to mimic breathing motion of the chest wall and the abdominal organs including the heart, respectively. A pneumatic system is used to move a heart mock-up to represent cardiac motion.
Results: The motion capabilities and MR compatibility of the phantom are validated in 3 T MRI experiments.
Impact: This work presents the development of a
dynamic torso phantom for MR applications, i.e. a scientific instrument that
can serve as ground truth for the investigation and characterization of novel
motion compensation and tracking methods.
Motivation
For magnetic resonance imaging (MRI) of the
torso, motion presents a challenge due to inevitable respiratory and cardiac
motion. Depending on the image acquisition technique, motion can cause
artifacts, including blurring, ghosting, signal dropouts and unwanted signal
enhancement1. Therefore, motion tracking and compensation are
subject of past and ongoing research2.
The development of such techniques requires
suitable phantoms that provide user-defined motion models with sufficient
spatial/temporal accuracy and reproducibility. Therefore, this work targets the
development of a dynamic torso motion phantom that can serve as ground truth
for the investigation and characterization of motion compensation and tracking
methods.Methods
Figure 1 shows the dynamic torso phantom
with its motion modules featuring three degrees of freedom (DOFs): (1)
breathing motion of the chest wall predominately in anteroposterior direction
(20 mm max. amplitude, 0.5 Hz max. rate), (2) breathing motion of the heart and
the abdominal organs in craniocaudal direction (50 mm max. amplitude, 0.5 Hz
max. rate), (3) simplified cardiac motion in craniocaudal direction (20 mm
amplitude, 4 Hz max. rate). The phantom position in the scanner corresponds to feet-first supine.
The motion modules are enclosed in a
double-walled acrylic shell (fabricated by Acrylstudio GmbH, Austria) filled
with saline solution (approx. 5 L deionized water, 5 g/L NaCl, 1 ml/L
Gadoteriol contrast agent) . The top front part of the shell is not covered by
acrylic glass but by a low-cost commercially available flexible gel mat (290 x
290 x 10 mm, polyurethane) representing the moveable chest wall. Breathing
motion (DOF 1 & 2) is realized with two independently controlled
non-magnetic linear piezo stages (Xeryon, Belgium) with a nominal driving force
of 3 N, maximum travel rage of 109 mm, speeds of up to 200 mm/s and spatial
accuracy in the µm range featuring customizable movement patterns.
For chest motion (DOF 1), the horizontal
movement of the piezo stage is translated into a vertical movement via a gear
box (Fig. 2a) containing a compound gear with a ratio of 1:5, providing an
increase of the driving force. This gear box holds a push head that contacts
and lifts the gel mat.
For abdominal organ motion (DOF 2), a
monorail track and a bulk shell on a wagon were constructed (Fig. 2b). In the
bulk shell, different organ phantoms can be placed (e.g., the additively
manufactured liver mock-up with main vessels shown in Fig. 2b). The heart
module is attached to the abdominal organ wagon to emulate heart movement with
the breathing cycle. Preliminary tests revealed that the piezo driving force
also needs to be increased for abdominal organ motion. Therefore, a second gear
box placed between the piezo motor box and the phantom body is currently under
construction.
To mimic heartbeat motion (DOF 3), a
sliding box holding a heart phantom (i.e., SLA-printed heart segmented from an
MRI scan filled with gel3: 250 g deionized water, 250 g ethylene
glycol, 3.5 g NaCl, 5 g Agarose, 0.25 g Sodium azide, and 1.5 ml Gadoteridol
contrast agent), and a full-plastic pneumatic movement system connected
underneath were designed (Fig. 2c). A double-action piston is used to avoid the
need for a metal spring for retraction. The pneumatic system is supplied with
compressed air (1.5 bar) from outside the scanner room via electromagnetic
valves (Festo GmbH, Germany) actuated by a MATLAB-controlled data acquisition
card (NI-USB-6001, National Instruments, USA), see Fig. 1c.Results
The phantom was assembled, placed on a
support plate, and first MR data were acquired without motion resulting in
artifact-free images, see Fig. 3. The currently operational motion DOFs (1
& 3) were tested while running a T1-weighted sequence for
approx. 1 min; resulting images with motion artifacts are shown in Fig. 4.Discussion and Conclusion
Besides adding the gear box for abdominal
organ motion, future work includes a detailed characterization of the
achievable accuracy and reproducibility. The modular design would permit us to
exchange, for example, the current simple heart model with a more advanced
version including flow.
Expected benefits in comparison to other
torso phantoms featuring both, breathing and cardiac motion4,5, are
higher accuracy thanks to the piezo stages and the possibility of clearly
separating cardiac, chest and diaphragm motion.
The phantom is mostly constructed from
low-cost components with standard plastic processing techniques (milling, 3D
printing). Exceptions are the custom-made outer shell and the piezo stages,
which could be replaced by a rectangular box and 3D printed pneumatic stepper
motors6 (reduced accuracy and speed), respectively. We plan to make the CAD
files of the phantom openly available once all modules are operational and
fully tested.Acknowledgements
This work was funded by Austrian Science
Fund (FWF) no. P37189-N and no. P35305-B, as well as the Focus Grants “T4MR”
and “NoLimit” from CMPBME, MedUni Vienna.References
1. Zaitsev, M., Maclaren, J. &
Herbst, M. Motion artifacts in MRI: A complex problem with many partial
solutions. J. Magn. Reson. Imaging 42, 887–901 (2015).
2. Runge, V. M., Richter, J. K. &
Heverhagen, J. T. Motion in Magnetic Resonance: New Paradigms for Improved
Clinical Diagnosis. Invest. Radiol. 54, 383–395 (2019).
3. Woletz, M., Roat, S., Hummer, A.,
Tik, M. & Windischberger, C. Technical Note: Human tissue-equivalent MRI
phantom preparation for 3 and 7 Tesla. Med. Phys. 48, 4387–4394
(2021).
4. Swailes, N. E., MacDonald, M. E.
& Frayne, R. Dynamic phantom with heart, lung, and blood motion for initial
validation of MRI techniques. J. Magn. Reson. Imaging 34, 941–946
(2011).
5. Fieseler, M. et al. A dynamic
thorax phantom for the assessment of cardiac and respiratory motion correction
in PET/MRI: A preliminary evaluation. Nucl. Instrum. Methods Phys. Res.
Sect. Accel. Spectrometers Detect. Assoc. Equip. 702, 59–63 (2013).
6. V. Groenhuis & S. Stramigioli.
Rapid Prototyping High-Performance MR Safe Pneumatic Stepper Motors. IEEEASME
Trans. Mechatron. 23, 1843–1853 (2018).