Jeehun Kim1,2, Stephen E. Russek3, Karl F. Stupic3, Cassandra M. Stoffer3, Kathryn E. Keenan3, David Rutkowski4, Jeff Kammerman4, Jean H. Brittain4, and Xiaojuan Li1
1Program of Advanced Musculoskeletal Imaging (PAMI), Cleveland Clinic, Cleveland, OH, United States, 2Department of Electrical, Computer, and Systems Engineering, Cleveland ClinicCase Western Reserve University, Cleveland, OH, United States, 3National Institute of Standards and Technology (NIST), Boulder, CO, United States, 4Calimetrix, Madison, WI, United States
Synopsis
Keywords: Phantoms, Phantoms
Motivation: There is not a commercially available musculoskeletal (MSK) relaxometry phantom.
Goal(s): Develop a dedicated MSK relaxometry phantom for T1, T2, and T1rho measurement quality assurance.
Approach: A cylindrical phantom with 12 vials containing modulated T1 and T2 samples was created, with an MR-visible thermometer for temperature tracking. Phantom stands were used for consistent positioning. MRI/NMR measurements were collected for longitudinal stability and temperature dependence.
Results: MRI measurements were in good agreement between two vendors with CVs<3% and demonstrated longitudinal stability with CVs<3% over a 3-month period. NMR measurements showed clear changes in T1, T2, and T1rho with changing temperature.
Impact: A stable MSK relaxometry phantom prototype was successfully
developed and characterized, including changes with temperature. With
harmonized measurement protocols, this phantom will facilitate the use of
quantitative relaxometry MRI in large-scale multi-site multi-vendor trials.
Introduction
Quantitative MRI (qMRI) relaxometry is widely investigated to
probe tissue compositional changes at early stages of diseases, which are difficult
to capture with conventional morphological imaging. Particularly, T1, T2, and
T1rho have been studied to detect early cartilage degeneration in
osteoarthritis.1,2 Due to its quantitative nature, standardization
of measurement and quality assurance of qMRI are crucial when interpreting findings
from patients, especially in multi-site and multi-vendor trials. Quantitative
phantoms are essential to this process.3
However, no dedicated musculoskeletal (MSK) relaxometry phantoms are commercially
available.
The goal of the study was to develop a dedicated MSK
relaxometry phantom that can be used for quality assurance of articular
cartilage T1, T2 and T1rho measures, with a focus on the knee joint. Methods
MSK Relaxometry Phantom Design
An MSK relaxometry phantom prototype was designed to have
the following attributes: 1) materials that mimic T1, T2, and T1rho relaxation
times characteristic of articular cartilage; 2) compatibility with commonly
used knee coils; and 3) the ability to measure temperature for possible
temperature dependency of phantom material. The Special-Order MSK Relaxometry
Phantom (Calimetrix, Madison, WI) is composed of a cylindrical phantom housing filled
with a doped water solution. It contains 12 vials filled with gels that simultaneously modulate
T1 and T2 to mimic knee cartilage (Figure 1).
A NIST-traceable analog MR-visible thermometer with marks on 15, 30, and 35⁰C is
embedded within the phantom to allow precise temperature monitoring from both
visual observation of the thermometer through the phantom wall and by MR
images. Finally, four custom external positioners (“stands”) for commonly used
MRI knee coils are provided to ensure repeatable phantom positioning parallel
to the patient table (QED 1Tx/28Rx or 15Rx knee coils, Philips 1Tx/8Rx or 16Rx knee
coils).
MRI T1, T2, and T1rho Measurement
MRI measurements were performed using two 3T scanners (Siemens
Prisma with QED 1Tx/15Rx knee coil and GE Healthcare SIGNA Premier with AIR
Anterior array coil) for inter-vendor comparison using coefficients-of-variation
(CV). T1 was measured using 2D inversion-recovery spin-echo (IR SE) sequence. T2
was measured using spin-echo (SE) and MLEV-16 (phase modulated CPMG) prepared
MAPSS sequence.4,5 T1rho was measured using a spin-lock-prepared
MAPSS sequence4,5
(Table 1).
Using the Siemens scanner, six measurements over three months were compared for
longitudinal CV.
NMR Measurement
NMR spectroscopy measures were performed at NIST using a 3T
system (Tecmag Redstone with Doty Scientific 5mm probe) on single material
sample tubes. IR T1s with composite inversion pulses were measured with 20
logarithmically-spaced inversion times. T2s were measured using three
sequences: CPMG sequences with echo timings of 2ms and 5.8ms, respectively, and
a spin-echo sequence. All T2 sequences contained 20 linearly-spaced echo times.
T1rho sequences used a spin-lock-prepared sequence with 20 times of spinlock.
Each sequence was adjusted to ensure full coverage of the relaxation range over
the timing array. All measurements were allowed for sufficient recovery (TR>5T1).
T1 and T2s were measured at six degrees ranging 16°C-26°C, with an increment of 2°C.Results
IR T1 and SE T2 demonstrated good
agreement between the two vendors, achieving CVs<3% in all vials except for vial
1, which had the shortest T2 (Table
2).
SE T2, MAPSS T2, and MAPSS T1rho showed similar values in all vials except for
vials 1 and 2. NMR measurement for vial 1 confirmed this trend and showed T1rho
dispersion (Figure 2).
Longitudinal measurements of all T1 and T2s had CVs<3% over
the 3-month period (Table
2).
In vial 1, NMR-measured T1
increased with increasing temperature, while NMR-measured T2 and T1rho
decreased with increasing temperature (Figure 3).Discussion
We developed a phantom that can be used for MSK applications
with T1, T2, and T1rho in the range of knee cartilage. MRI measurements using
IR T1 and SE T2 were reproducible between the two vendors. Such agreement is
essential to conduct a multi-vendor quantitative relaxometry study. Vial 1 had discrepancy
between T2s with different spacing between refocusing pulses, verified by the
NMR measurement. This finding shows the necessity of reporting specific
sequence parameters when reporting T2. NMR measurements also verified the T1
and T2 dependency on temperature, and these could be used to correct phantom measurements
conducted on MRI systems with different ambient temperatures. The phantom
measurements were stable over 3-months. These measurements will be continuously
collected for long-term stability evaluation. NIST NMR measurements will be
performed on all vials to provide reference values.Conclusion
A stable MSK relaxometry phantom prototype was successfully
developed and characterized, including changes with temperature. With
harmonized measurement protocols, this phantom will facilitate the use of
quantitative relaxometry MRI in large-scale multi-site, multi-vendor trials.Acknowledgements
The study was supported by NIH/NIAMS R01AR077452.References
1. de Mello, R., Ma, Y., Ji, Y., Du, J., &
Chang, E. Y. (2019). Quantitative MRI MSK techniques: An update. AJR. American
journal of roentgenology, 213(3), 524.
2. Eck, B. L., Yang, M., Elias, J. J., et al. (2023). Quantitative MRI for
evaluation of musculoskeletal disease: cartilage and muscle composition, joint
inflammation, and biomechanics in osteoarthritis. Investigative radiology,
58(1), 60-75.
3. Keenan, K. E., Ainslie, M., Barker, A. J., et al. (2018). Quantitative
magnetic resonance imaging phantoms: a review and the need for a system
phantom. Magnetic resonance in medicine, 79(1), 48-61.
4.
Li, X., Pedoia, V., Kumar, D., et al. (2015). Cartilage T1ρ and T2
relaxation times: longitudinal reproducibility and variations using different
coils, MR systems and sites. Osteoarthritis and cartilage, 23(12), 2214-2223.
5. Kim, J., Mamoto, K., Lartey, R., et al. (2020). Multi-vendor multi-site T1ρ
and T2 quantification of knee cartilage. Osteoarthritis and cartilage, 28(12),
1539-1550.