Sarah Garrow1, Rasim Boyacioglu1, Kathryn E Keenan2, Mark Griswold1, and William Grissom1
1Case Western Reserve University, Cleveland, OH, United States, 2National Institute of Standards and Technology, Boulder, CO, United States
Synopsis
Keywords: Thermometry/Thermotherapy, Thermometry
Motivation: Proton resonance frequency (PRF)-shift thermometry is the current standard for MR-temperature monitoring in interventional procedures. However, the long TE required for phase contrast induces signal dropout and increases sensitivity to metal and motion artifacts.
Goal(s): Use quadratic RF phase (qRF) MR fingerprinting to image off-resonance frequency for thermometry at a real time frame rate.
Approach: Because PRF change is much more sensitive to temperature change in aqueous tissue than T1/T2, we propose a “lightweight” constant low-flip-angle MRF sequence optimized for 3s or less frame rate to measure temperature from the PRF shift.
Results: We implemented qRF-MRF thermometry with high spatiotemporal resolution.
Impact: A
thermometry method with high temporal resolution that does not suffer from
signal dropout at high temperatures can enable more accurate temperature
monitoring for MR-guided interventional procedures. Additionally, it has
potential to be more robust to motion artifacts.
Introduction
Proton resonance
frequency shift (PRFS) thermometry is the current standard for MR-based
temperature monitoring in interventional procedures. It measures phase changes
due to frequency changes with temperature [1]. However, the relatively long TE
required for the PRFS method leads to signal dropout in heated regions as T2
shortens and T1 lengthens with heating. Additionally, PRFS acquisitions
can have poor SNR efficiency and high sensitivity to physiological and other
spurious motion. Quadratic RF phase MR fingerprinting (qRF-MRF) can also quantify
off-resonance using much shorter TEs and has a very high SNR efficiency, and
thus has the potential to generate high-precision temperature maps [2]–[4]. Here we report a dynamic qRF-MRF
pulse sequence and model-based image and continuous sliding window temperature
map reconstruction for high temporal resolution and demonstrate its precision
in vivo and its ability to monitor focused ultrasound heating. Methods
Pulse Sequence Mechanism
The qRF-MRF thermometry
sequence works by continuously sweeping the RF excitation phase of a constant
low-flip-angle as illustrated in Figure 1. The resonance frequency continuously sweeps
from -1/TR to +1/TR Hertz in a short-TE/short-TR bSSFP sequence. Fig 1 shows
that voxels with different temperatures or off-resonance frequencies will then generate
distinct signal time courses, which can be matched to a Bloch-simulated
dictionary to quantify off-resonance. We used Monte Carlo simulations with random
noise added to optimize the sequence parameters, including the ramp of the
quadratic RF phase and flip angle.
Scanner Implementation
The qRF-MRF thermometry sequence was
implemented on a 3T (Vida, Siemens Healthcare) scanner, using a 16-channel Rx
coil and a spiral-out readout. A constant TR = 10 ms, TE = 2.5 ms, and flip
angle = 10° were used with a quadratically increasing 4.2n2 RF phase
to continuously cycle the sequence’s resonance frequency from -50 to +50 Hz
range. Conventional 2DFT PRF temperature maps were also acquired using a 12 ms
TE gradient echo sequence, generating a dynamic image every 2.2 seconds. Both
sequences had 256x256 mm2 FOVs and 2 x 2 x 5 mm3 resolution.
Temperature Reconstruction
A dictionary was
constructed for a single brain tissue-like T1/T2 pair (825
ms, 70 ms), a range of off-resonances (-50 Hz to 50 Hz), and 21 Gaussian linewidths
(1 to 51 Hz). Temperature maps were reconstructed dynamically over 1-3 second-long
sliding time windows. Images within each window were jointly reconstructed
using a
dictionary-constrained algorithm that solved for coefficient maps in an SVD-compressed-dictionary
subspace; the forward model is illustrated in Figure 2A. A temperature
map was generated from a window’s frequency map by synthesizing a GRE image
with TE equal to the qRF-MRF TR to match phase wraps, and then applying the
hybrid referenceless multibaseline subtraction thermometry method Figure 2B [5].
Experiments
In vivo healthy subject scans without
heating were conducted in accordance with local IRB,
acquiring two slices covering the hippocampus and thalamus in two subjects,
measured over several minutes to enable comparison of 2DFT and qRF-MRF
temperature precision (Figure 3).
In a phantom, a heating experiment was performed
using a focused ultrasound transducer (H115MR, Sonic Concepts) at 250 mV pp for
2 minutes. The phantom was a 1% agar weight by volume of water for the coupling
medium and a 1% agar / 4 % graphite weight by volume for the main body of the
phantom. The qRF-MRF data were reconstructed with 300 TRs per dynamic time
point, updated every 0.5 seconds (Figure 4)Results and Discussion
Figure 3 shows data for 2 different slices in 2 subjects and a
comparison of precision between 2DFT and MRF, measured as standard deviation
over time; the qRF-MRF has improved temperature precision in nearly every case,
for the same scan time. Figure 4 compares temperature maps of the
heated phantom, both showing similar values for peak heating. The qRF-MRF
method provides temperature maps with much higher (>4x here) temporal
resolution due to the sliding window reconstruction. Figure 5A shows a comparison of temperature error and dictionary match correlations between the 2DFT, gridded qRF-MRF, and iterative recon qRF-MRF; the iterative recon yields improved temperature precision and greater certainty in frequency matching. Figure 5B shows maps of Gaussian linewidths and through-time standard deviations from a qRF-MRF reconstruction; areas with wider linewidths correspond to areas with higher temperature uncertainty. Also shown clearly in the linewidth map are high-iron regions such as the putamen. Conclusion
qRF-MRF
temperature mapping enables high spatiotemporal resolution thermometry with
improved precision compared to 2DFT acquisitions. Future work will focus on
evaluating relative sensitivity to spurious artifacts such as from water bath
motion in FUS ablation. Acknowledgements
This work was supported by NIH grants R01 NS120518 and R01
EB028773. References
[1] V. Rieke and K.
Butts Pauly, “MR thermometry,” J. Magn. Reson. Imaging, vol. 27, no. 2,
pp. 376–390, Feb. 2008, doi: 10.1002/jmri.21265.
[2] R.
Boyacioglu, M. Poorman, K. Keenan, and M. Griswold, “Magnetic Resonance
Fingerprinting with Quadratic RF Phase for Continuous Temperature Monitoring in
Aqueous Tissues,” Proc. Intl. Soc. Mag. Reson. Med. 28 (2020).
[3] K.
Keenan, M. A. Griswold, M. E. Poorman, and R. Boyacioglu, “Magnetic Resonance
Fingerprinting Thermometry,” US Patent No. 11519987 B2
[4] C.
Y. Wang, S. Coppo, B. B. Mehta, N. Seiberlich, X. Yu, and M. A. Griswold,
“Magnetic resonance fingerprinting with quadratic RF phase for measurement of
T2* simultaneously with δf, T1, and T2,” Magnetic Resonance in Medicine,
vol. 81, no. 3, pp. 1849–1862, 2019, doi: 10.1002/mrm.27543.
[5] W.
A. Grissom et al., “Hybrid referenceless and multibaseline subtraction
MR thermometry for monitoring thermal therapies in moving organs: Hybrid MR
thermometry in moving organs,” Med. Phys., vol. 37, no. 9, pp.
5014–5026, Aug. 2010, doi: 10.1118/1.3475943.