Danielle Kara1, Yuchi Liu2, Shi Chen1, Thomas Garrett1, Xiaoming Bi3, Deborah Kwon4, and Christopher T Nguyen1,4,5,6
1Cardiac Innovation Research Center, Heart Vascular and Thoracic Institute, Cleveland Clinic, Cleveland, OH, United States, 2Siemens Medical Solutions USA, Cleveland, OH, United States, 3Siemens Medical Solutions USA, Los Angeles, CA, United States, 4Cardiovascular Medicine, Heart Vascular and Thoracic Institute, Cleveland Clinic, Cleveland, OH, United States, 5Biomedical Engineering, Case Western Reserve and Cleveland Clinic, Cleveland, OH, United States, 6Imaging Institute, Cleveland Clinic, Cleveland, OH, United States
Synopsis
Keywords: DWI/DTI/DKI, Diffusion Tensor Imaging
Motivation: SNR and parameter map accuracy in cardiac DTI are limited by maximum gradient strength related to motion-compensation and diffusion encoding time, precluding evaluation of helical cardiomyocyte structure.
Goal(s): Our goal was to improve SNR and cardiac DTI tissue microstructure characterization using an MR system capable of 200mT/m maximum gradient strength.
Approach: DTI was performed in human and swine subjects using standard (40mT/m), performance (80mT/m), and ultra-high-performance (200mT/m) maximum gradient strengths, with zeroth, first, and second-order motion compensating gradients.
Results: SNR and DTI tissue characterization were improved with ultra-high-performance gradients, however second-order motion compensation continued to be required to prevent motion artifacts.
Impact: Ultra-high performance 200mT/m gradients enable high SNR
cardiac DTI with improved characterization of helical cardiomyocytes, potentially
addressing the clinical need for noninvasive cardiac microstructure evaluation.
Purpose
Breathing and cardiac motion pose significant challenges in
diffusion tensor cardiac magnetic resonance (DT-CMR). While diffusion weighted
(DW) acquisitions employing traditional (M0) and first-order (M1)
motion-compensating diffusion gradients suffer from motion-related signal
dropout, second-order (M2) motion-compensating diffusion gradients enable
quality spin echo DT-CMR1,2.
However, increasing motion compensation results in increased diffusion time and
TE, corresponding to increased opportunity for motion during diffusion encoding
and decreased signal to noise ratio (SNR). With the introduction of an MR system
capable of 200mT/m maximum gradient strength, reduction in TE is possible,
resulting in restored SNR and improved DT-CMR. Reduced diffusion time is also
possible, which may enable the use of lower order motion compensation. In this
work, DT-CMR data acquired using clinical standard (STD), performance (P), and
ultra-high-performance (UHP) maximum gradient strengths are compared as are the
effects of zeroth (M0), first (M1), and second (M2) order motion compensation
at UHP gradient strengths.Methods
DT-CMR was performed in 15 healthy volunteers, 1 pericarditis
patient, and 1 swine subject 8 weeks post-myocardial infarction (MI) under IRB and
IACUC approved protocols on a 3T MR system (MAGNETOM Cima.X, Siemens Healthineers
AG, Erlangen, Germany) capable of 200mT/m maximum gradient strength
(free-breathing 2DRF zoomed diffusion prepared spin echo2,
350mm FOV, 128x48 matrix, TR=500ms, 12 diffusion directions, b0=50s/mm2,
b=500s/mm2, 8 averages, end systole, human subjects: five 8-mm
slices, swine subject: six 5-mm slices). Human subjects were scanned using five
DTI protocols: STD-M2 (Gmax=40mT/m, TE=122ms), P-M2 (Gmax=80mT/m, TE=79ms),
UHP-M2 (Gmax=200mT/m, TE=59ms), UHP-M0 (Gmax=200mT/m, TE=40ms), UHP-M1
(Gmax=200mT/m, TE=54ms). The swine subject was scanned using three DTI protocols
(STD-M2, P-M2, and UHP-M2) and slice-matched late gadolinium enhancement (LGE)
imaging was performed.
Respiratory motion correction of DW images was achieved with
MT-MOCO2.
DTI analysis yielding mean diffusivity (MD), fractional anisotropy (FA), and
helix angle (HA) maps was performed using a custom python library. Manual
segmentations of the left ventricle (LV) were used to calculate mean MD, FA,
and helix angle transmurality (HAT). SNR in DW images was calculated in the LV
using a modified NEMA method 13.
Paired t-tests were performed to test for significant differences between
acquisition methods, with significance defined as p<0.05. Results
Figure 1 shows DW images from a representative volunteer. UHP-M2
images are observed to have superior signal intensity, while STD-M2 and UHP-M0
images have significantly reduced signal due to long TE and motion-induced
dropout, respectively. UHP-M1 images also exhibit motion-induced dropout, as
indicated by red arrows.
Figure 2 shows MD, FA, and HA maps from a representative healthy
volunteer. STD-M2 and UHP-M0 maps are extremely poor quality with almost no
helix structure. Signal dropout in UHP-M1 data results in deviations from the
expected right-to-left-handed helical structure of the LV in the HA map. While
P-M2 and UHP-M2 maps demonstrate the expected decreasing transition in HA from
endocardium to epicardium, the transition is smoother and better defined for
UHP-M2 data.
Results of paired t-tests for average SNR, MD, FA, and HAT in
the LV from healthy volunteer data are presented in Figure 3. UHP-M2 images
have significantly higher SNR than STD-M2 (p<0.0001) and P-M2 (p<0.01)
images. While mean MD (p<0.01), FA (p>0.01), and HAT (p>0.01)
from P-M2 and UHP-M2 data are comparable and within expected physiological
ranges, UHP-M2 data yields reduced variability and decreased mean HAT, as expected
for healthy volunteers4.
Figure 4 shows DTI maps obtained from patient data. STD-M2,
UHP-M0, and UHP-M1 maps exhibit poor quality, suggesting impaired clinical
utility. While P-M2 and UHP-M2 maps are comparable, disruptions of the
expected, smoothly varying helical structure of the LV are observed in the
basal slices of the P-M2 HA maps that are shown to be imaging artifacts due to
their absence in the UHP-M2 data (indicated with arrows).
LGE images and DTI maps obtained from the swine subject are
presented in Figure 5, with a magnified view of the scar region (gray box). While
the regions of scar exhibit elevated MD, reduced FA, and reduced HA in P-M2 and
UHP-M2 datasets, the boundaries of the scar are more clearly defined and show
improved agreement with enhanced regions on LGE images in UHP-M2 maps. Conclusion
While high-quality DT-CMR is possible on systems achieving
Gmax=80mT/m, the use of UHP gradient systems (Gmax=200mT/m) enables short-TE
acquisitions yielding significantly higher SNR DW images (p<0.01) and
consequently improved DT-CMR. Despite shortened diffusion time possible with the
UHP system, M2 compensation continues to be required to prevent motion-induced
signal dropout in cardiac DW images. In swine and patient data, UHP gradients improved
visualization of MI scar and helical cardiomyocyte structure, particularly in
the endocardial border regions. Acknowledgements
This
work was supported by NIHLBI (R01
HL151704, R01 HL159010) and NIBIB (R01 EB033853). We
thank the Imaging Institute and the Cardiovascular Innovation Research Center
for their support.References
- Nguyen C, Fan Z, Xie Y, et al. In vivo
diffusion-tensor MRI of the human heart on a 3 tesla clinical scanner: An
optimized second order (M2) motion compensated diffusion-preparation approach:
Optimized in vivo M2 DT-CMR for a 3T clinical scanner. Magn Reson Med.
2016;76(5):1354-1363. doi:10.1002/mrm.26380
- Nguyen
CT, Christodoulou AG, Coll‐Font J, et al. Free‐breathing diffusion tensor MRI
of the whole left ventricle using second‐order motion compensation and
multitasking respiratory motion correction. Magn Reson Med.
2021;85(5):2634-2648. doi:10.1002/mrm.28611
- Goerner
FL, Clarke GD. Measuring signal‐to‐noise ratio in partially parallel imaging
MRI. Med Phys. 2011;38(9):5049-5057. doi:10.1118/1.3618730
- Nguyen
CT, Dawkins J, Bi X, Marbán E, Li D. Diffusion Tensor Cardiac Magnetic
Resonance Reveals Exosomes From Cardiosphere-Derived Cells Preserve Myocardial
Fiber Architecture After Myocardial Infarction. JACC Basic Transl Sci.
2018;3(1):97-109. doi:10.1016/j.jacbts.2017.09.005