Afis Ajala1, David Mills1, Eric Fiveland1, Heather Chan1, Seung-Kyun Lee1, Thomas KF Foo1, Jia Xu2, Bryan Bednarz3, Diego Hernando3, Shane Wells4, James H. Holmes2, and Desmond Yeo1
1GE HealthCare, Niskayuna, NY, United States, 2University of Iowa, Iowa City, IA, United States, 3University of Wisconsin-Madison, Madison, WI, United States, 4University of Michigan, Ann Arbor, MI, United States
Synopsis
Keywords: MR-Guided Interventions, MR-Guided Interventions
Motivation: Simultaneous ultrasound (U/S) and MR imaging has been reported on whole-body MRI systems with conventional gradients (CVG) but remains unexplored on high-performance gradient (HPG) platforms.
Goal(s): We assessed the MR and U/S image quality, alongside thermal impact on an MR-compatible U/S probe at CVG and HPG configurations.
Approach: Fast spoiled gradient echo imaging, echo planar imaging (EPI), B0-field map and U/S probe temperature were measured.
Results: Moderate-to-severe susceptibility-induced signal loss (due to U/S probe) extended ~8 mm into phantom at gradient isocenter with no visible U/S image artifacts. A substantial temperature rise was observed on U/S probe during EPI at HPG configuration.
Impact: This study evaluates and compares image
quality and thermal risks of simultaneous MR and U/S imaging on a
high-performance gradient system versus a conventional clinical system. Our
findings can help optimize the protocols for image-guided U/S intervention
using high-performance gradients.
Introduction
Using
an MRI-compatible research ultrasound (U/S) probe on a whole-body MRI
system, simultaneous MR
and U/S imaging has been demonstrated
for the purpose of mitigating the effects of motion on the accuracy of image-guided
therapy1,2. MRI scanners
equipped with high-performance
gradients (HPG) combine high gradient amplitudes (Gmax) and high
maximum slew rates (SRmax) to enable high-resolution MRI with
reduced susceptibility artifacts3–5. For example, the MAGNUS gradient
system can simultaneously deliver a Gmax and SRmax
of 300 mT/m and 750 T/m/s respectively3. However, multimodal MR and U/S imaging
on an HPG MRI system remains unexplored. Here, we assess MR and U/S image
quality, and the temperature rise in an MRI-compatible U/S probe at conventional
gradient (CVG) and HPG configurations.Materials and Methods
A multi-tissue U/S phantom (Model
040GSE, Zerdine® Hydrogel) was used for all MR and U/S imaging.
Ultrasound Imaging
All U/S imaging was carried out on
a 4D system (Vivid E95, GE Healthcare) equipped with a hands-free MR-compatible
E4D research probe1. 4D U/S images of the phantom were acquired in harmonic
imaging mode (1.7/3.3 MHz) as shown in Figure 1A.
Temperature Measurement
The U/S probe surface temperature
was measured during multimodality imaging using fiber optic temperature sensors
placed on the surface of the probe (Figure 1B). A 10-minute wait time was
allowed between each temperature measurement to permit convective cooling of
the U/S probe. All temperature measurements were normalized to start at the ambient
scanner room temperature (~21oC).
Magnetic Resonance Imaging
All MRI HPG experiments were
carried out on a MAGNUS gradient system (GE Healthcare, Waukesha, WI) at Gmax=300
mT/m and SRmax=750 T/m/s. Imaging at CVG configuration was achieved
by derating the MAGNUS system to operate at Gmax and SRmax
of 100 mT/m and 200 T/m/s respectively. RF signal transmit and reception were achieved
using the integrated body coil and a 20-channel AIR anterior coil array (GE
Healthcare) respectively (Figure 1C). To assess the B0 inhomogeneity
induced by the U/S probe, a 2D B0-field map was obtained during U/S
imaging using the following parameters: FOV=25 cm, TE/TR=3.4 ms/100 ms, voxel size=1.95x1.95x5
mm3, slices=40, average(s)=1. A control B0
map without the U/S probe was also acquired with identical parameters. For U/S probe
surface temperature measurement and image quality assessment, multi-phase fast
spoiled gradient echo (FSPGR) images were obtained at CVG and HPG configurations for a duration
of 5 mins with parameters: FOV=250 mm, voxel size=1.95x1.95x5 mm3, slices=10,
TE/TR=0.9 ms/3.0 ms, bandwidth=±31.25 kHz, temporal resolution=4 s. EPI
scans with identical parameters were obtained at bandwidth=±500 kHz using HPG (TE/TR=8.9
ms/32.0 ms, temporal resolution=0.32 s and echo spacing=0.3 ms) and CVG (TE/TR=14
ms/44.0 ms, temporal resolution=0.44 s/phase and echo spacing=0.6 ms) configurations.Results and Discussion
Representative
slices from B0 maps and their corresponding magnitude images show
regions of higher B0 inhomogeneity around the location of the U/S probe (Figure
2A and 2B). The regions of higher inhomogeneity are not visible in the control
images as the U/S probe was removed from the phantom. The depth of the B0
inhomogeneity on the phantom was maximum (19.5 mm) at 28
mm from the gradient isocenter (Figure 2C). FSPGR and EPI MRI images obtained
during temperature mapping on the surface of the U/S probe at CVG and HPG configurations
are shown in Figure 3. There was no visible difference between the FSPGR images
obtained at CVG and
HPG configurations (Figure 3A). However, EPI distortion artifacts in the acquisitions
at CVG configuration were more severe compared to HPG configuration due to
the shorter ESP enabled by the latter (Figure 3B). There is also no significant
difference in the U/S images obtained during multimodal imaging at CVG and HPG configurations
(Figure 4). The temperature changes during the multimodal scans stayed below 28
oC except for the EPI scans carried out at HPG configuration (Figure 5). The U/S probe surface temperature reached ~38 oC at the
sides of the probe. This increased temperature can be associated with higher gradient-induced
eddy currents on metallic components in the U/S probe at higher SRmax. The eddy current heating at HPG configurations needs to be
addressed in future work to fully utilize the higher SRmax of HPGs for EPI acquisitions during simultaneous MR-U/S imaging. Conclusion
High-performance gradient systems can be utilized for rapid
simultaneous MR-U/S FSPGR and EPI imaging with lesser distortion compared to conventional gradient configurations. However, there is a need for mitigation of eddy current induced
heating in EPI at High-performance gradient configuration.Acknowledgements
Research reported in this work was supported by the National Institute of Health under award numbers R01CA266879
and R01CA190298.References
1. Bednarz,
B. P. et al. First-in-human Imaging Using a MR-compatible e4D Ultrasound
Probe for Motion Management of Radiotherapy. Phys Med 88, 104–110
(2021).
2. Lee, W. et al. A Magnetic
Resonance Compatible E4D Ultrasound Probe for Motion Management of Radiation
Therapy. IEEE Netw 2017, 10.1109/ULTSYM.2017.8092223 (2017).
3. Foo, T. K. F. et al. Highly
efficient head-only magnetic field insert gradient coil for achieving
simultaneous high gradient amplitude and slew rate at 3.0T (MAGNUS) for brain
microstructure imaging. Magnetic Resonance in Medicine 83,
2356–2369 (2020).
4. Setsompop, K. et al. Pushing the
limits of in vivo diffusion MRI for the Human Connectome Project. Neuroimage
80, 220–233 (2013).
5. Weiger, M. et al. A
high-performance gradient insert for rapid and short-T2 imaging at full duty
cycle. Magn Reson Med 79, 3256–3266 (2018).