Zijian Gao1, Ziqiang Yu1, Ziqin Zhou1,2, Jian Hou1, Baiyan Jiang1,3, Michael Tim-yun ONG 4, and Weitian Chen1
1Department of Imaging and Interventional Radiology, The Chinese University of Hong Kong, Hong Kong, Hong Kong, 2MR Collaboration, Siemens Healthineers Ltd., Hong Kong, Hong Kong, Hong Kong, 3Illuminatio Medical Technology Limited, Hong Kong, Hong Kong, Hong Kong, 4Department of Orthopaedics and Traumatology, The Chinese University of Hong Kong, Hong Kong, Hong Kong
Synopsis
Keywords: Magnetization Transfer, Quantitative Imaging, Macromolecular proton fraction
Motivation: The residual dipolar coupling (RDC) can lead to the orientation-dependent measurements in ordered tissues in MRI, potentially confounding their clinical applications.
Goal(s): We demonstrate the potential confounding effect from tissue orientation in quantitative magnetization transfer can be suppressed by using a new technique Macromolecular Proton Fraction Mapping based on Spin-Lock (MPFSL).
Approach: Applying MPFSL, we can adjust both the resonance frequency offset and the amplitude of spin-lock radiofrequency pulse to achieve a strong effective spin-lock field to suppress RDC, eliminating orientation-dependency of MPF measurement. Human knee specimen experiments conducted verified this finding.
Results: The MPF measured using MPFSL shows insensitivity to tissue orientations.
Impact: Spin-lock based quantitative
magnetization transfer imaging can achieve orientation-independent
quantification, thus having potential applications in characterization of highly-ordered
tissues such as cartilage and myelin.
Introduction
Orientational anisotropy of tissues with ordered structures often confounds quantification
of tissue parameters. In highly-ordered tissues like
cartilage and myelin, dipole-dipole interactions between tissue structures and
the static magnetic field $$$B_0$$$ result in orientation-dependency of MRI signal, leading to well-known "magic angle effect"1,2, as commonly observed in T2 maps3–5.
Quantification of magnetization transfer (MT) parameters such as
macromolecular proton fraction (MPF) provides valuable insights into the
biochemical composition and molecular properties of biological tissues 6–8. Several works reported orientation-dependency of
MT parameters 9–11. wherein quantitative MT is commonly performed based
on saturation radiofrequency (RF) pulses.
Recently, a new technique called Macromolecular Proton Fraction mapping based on Spin-Lock
(MPFSL) has been reported for quantitative MT based on off-resonance spin-lock
RF pulses12,13.
Note the spin-lock RF pulses can suppress residual dipolar coupling
(RDC) when spin-lock field is sufficiently high and thus quench orientation-dependency of MRI signal14,15. For on-resonance
spin-lock, the maximum spin-lock magnetic field is limited by RF power and
specific absorption rate (SAR). Consequently, orientation-dependency is often
observed in on-resonance T1rho imaging with spin-lock< 1000Hz16,17. In contrast, for
off-resonance spin-lock, the spin-lock field is a combination of B1 field from
RF pulse and resonance frequency offset which allows us to achieve a strong spin-lock
field without violating the limit of SAR and RF power. Thus we can use off-resonance spin-lock techniques to achieve orientation independent quantitative MT imaging. This study investigated the orientational independence of MPFSL based
on this mechanism with human knee specimen experiments.Methods
Theory
In highly-ordered
tissue, the relaxation rate $$$R_{1\rho}$$$ at off-resonance
spin-lock can be expressed as:
$$
R_{1\rho}=R_w^i+R_2^a(\theta)+R_{mt}(\Delta\omega,\omega_1) \tag{1}
$$
where $$$R_w^i $$$ and $$$R_2^a(\theta)$$$ are the
relaxation rates of the water pool, corresponding to isotropic and anisotropic
water molecular relaxation, respectively; $$$R_{mt}$$$ is the relaxation rate due to MT effect 18; $$$\Delta\omega$$$ is frequency offset (FO) and $$$\omega_1$$$ is frequency of
spin-lock (FSL); $$$\theta$$$ is the orientation of ordered tissue with
respect to $$$B_0$$$. Note the chemical exchange influence can be ignored in Eq. (1) using MPFSL13.
The
anisotropic water molecular relaxation rate $$$R_w^a(\theta)$$$ is given by:
$$
R_w^a(\theta) = \frac{R_2^a(\theta)}{1+4\omega_{eff}^2\tau_b^2} \tag{2}
$$
Where $$$\omega_{eff}$$$ is effective spin-lock RF strength19, which equal to$$$\sqrt{\Delta\omega^2+\omega_1^2}$$$ , and $$$R_2^a(\theta)$$$ is $$$R_2^a(3cos^2\theta-1)^2/4$$$ , $$$R_2^a$$$ is the maximum of $$$R_2^a(\theta)$$$ 20.
In MPFSL, relaxation
rates in two off-resonance spin-lock fields with different spin-lock amplitude
but along the same direction are obtained13 to remove water pool signal, resulting in a relaxation rate termed $$$R_{mpfsl}$$$, which is specific to the MT pool for efficient
MPF maps. Combine this condition with
Eq. (1) and (2), we have:
$$
R_{mpfsl}=R_{1\rho}^{(2)}-R_{1\rho}^{(1)}=\Delta(\frac{R_2^a(\theta)}{1+4\omega_{eff}^2\tau_b^2})+\Delta R_{mt} \tag{3}
$$
The
element $$$\Delta(\frac{R_2^a(\theta)}{1+4\omega_{eff}^2\tau_b^2})$$$ can be minimized using
a stronger $$$\omega_{eff}$$$,resulting in a very low sensitivity of $$$R_{mpfsl}$$$ with respect to orientations.
Human
knee specimen preparation
A human knee specimen from a total knee replacement surgery. All experiments were conducted under the approval from the Institutional Review Board. To ensure stability and proper positioning during imaging, the
specimen was affixed to a sealed plastic square container using
ethyl-2-cyanoacrylate adhesive (Henkel Ltd, Germany). The container, accommodating the
specimen, was filled with phosphate-buffered saline (PBS) at room temperature. The container was attached to a hand-made device
which allows precise orientational control in scanner. Figure 1
illustrates the experiment setup.
MR
data acquisition and analysis
All
specimen studies were performed using a 3T MRI Scanner (Prisma, Siemens
Healthcare, Germany) with a Tx/Rx Knee Coil. The parameters
of MPFSL protocol include FSL1/FSL2=100/500Hz, FO1/FO2=1000/5000Hz and
time of spin-lock (TSL)=60ms. Additionally, PD-weighted (PDW), T2 maps, T1
maps, and on-resonance T1rho mapping (FSL=500Hz) were
also acquired. The images were acquired at orientations 0°, 15°, 30°, 45°, 60°,
75°, and 90° with respect to $$$B_0$$$, respectively. Further details of MRI
imaging protocol are shown in Table 1. The data analysis
was performed using MATLAB (MathWorks,USA). Results
Figure 2 displays PDW images and the maps of R1, R2, R1rho and $$$R_{mpfsl}$$$ at different orientations. Note R2 and R1rho vary with orientation. Conversely, $$$R_{mpfsl}$$$, like R1, is independence of orientation.
Figure 3 shows plots of the relaxation rates with varying orientations. Two ROIs was chosen in the superficial zone (SZ) and deep zone
(DZ), respectively. R2 and R1rho show significant variations at different orientations. $$$R_{mpfsl}$$$ and R1 display minimal sensitivity to orientations.Discussion and Conclusion
We demonstrated that quantitative MT
based on spin-lock approach can suppress RDC and eliminate confounding effect from
tissue orientation when measuring MPF. This can be achieved with a typical
protocol used in MPFSL without violating the limit of RF power or SAR. The orientation independent MPF
mapping is desirable for ordered tissue's characterization. Acknowledgements
This study was supported by a grant from the Research
Grants Council of the Hong Kong SAR (Project GRF 14201721), and a grant from
the Innovation and Technology Commission of the Hong Kong SAR (Project
No.MRP/046/20x). References
1. Erickson
SJ, Prost RW, Timins ME. The “magic angle” effect: background physics and
clinical relevance. Radiology. 1993;188(1):23-25.
doi:10.1148/radiology.188.1.7685531
2. Xia Y. Magic-angle effect in
magnetic resonance imaging of articular cartilage: a review. Invest Radiol.
2000;35(10):602-621. doi:10.1097/00004424-200010000-00007
3. Mosher TJ, Smith H, Dardzinski BJ,
Schmithorst VJ, Smith MB. MR imaging and T2 mapping of femoral cartilage: in
vivo determination of the magic angle effect. American Journal of
Roentgenology. 2001;177(3):665-669.
4. Wang L, Regatte RR. Investigation of
regional influence of magic-angle effect on T2 in human articular cartilage
with osteoarthritis at 3 T. Academic Radiology. 2015;22(1):87-92.
5. Leskinen HPP, Hänninen NE, Nissi MJ.
T 2 orientation anisotropy mapping of articular cartilage using
qMRI. Phys Med Biol. 2023;68(8):085004. doi:10.1088/1361-6560/acc169
6. Fast macromolecular proton fraction
mapping from a single off‐resonance magnetization transfer measurement -
Yarnykh - 2012 - Magnetic Resonance in Medicine - Wiley Online Library.
https://onlinelibrary.wiley.com/doi/full/10.1002/mrm.23224. Accessed October
31, 2023.
7. Kisel AA, Naumova AV, Yarnykh VL.
Macromolecular proton fraction as a myelin biomarker: principles, validation,
and applications. Frontiers in neuroscience. 2022;16:819912.
8. Eckstein F, Burstein D, Link TM.
Quantitative MRI of cartilage and bone: degenerative changes in osteoarthritis.
NMR in Biomedicine. 2006;19(7):822-854. doi:10.1002/nbm.1063
9. Joseph T. Myelin water,
magnetization transfer, and inhomogeneous magnetization transfer imaging
orientation dependence : observations and thoughts. 2023.
doi:10.14288/1.0435676
10. Karan P, Edde M, Gilbert G, Barakovic
M, Magon S, Descoteaux M. Data-driven characterization and correction of the
orientation dependence of magnetization transfer measures using diffusion MRI.
October 2023:2023.10.05.561088. doi:10.1101/2023.10.05.561088
11. Pampel A, Müller DK, Anwander A,
Marschner H, Möller HE. Orientation dependence of magnetization transfer
parameters in human white matter. NeuroImage. 2015;114:136-146.
doi:10.1016/j.neuroimage.2015.03.068
12. Hou J, Wong VWS, Wong GLH, et al.
Macromolecular proton fraction mapping based on spin-lock for the non-invasive
diagnosis of early stage liver fibrosis. Int Soc Magn Reson Med,
International Society for Magnetic Resonance in Medicine. 2021.
https://index.mirasmart.com/ISMRM2021/PDFfiles/0314.html. Accessed October 31,
2023.
13. Hou J, Wong VW, Jiang B, et al.
Macromolecular proton fraction mapping based on spin‐lock magnetic resonance
imaging. Magn Reson Med. 2020;84(6):3157-3171. doi:10.1002/mrm.28362
14. Rautiainen J, Nissi MJ, Liimatainen T,
Herzog W, Korhonen RK, Nieminen MT. Adiabatic rotating frame relaxation of MRI
reveals early cartilage degeneration in a rabbit model of anterior cruciate
ligament transection. Osteoarthritis and Cartilage.
2014;22(10):1444-1452. doi:10.1016/j.joca.2014.04.023
15. Casula V, Nissi MJ, Podlipská J, et al.
Elevated adiabatic T1ρ and T2ρ in articular cartilage are associated with
cartilage and bone lesions in early osteoarthritis: A preliminary study. J
Magn Reson Imaging. 2017;46(3):678-689. doi:10.1002/jmri.25616
16. Hänninen N, Rautiainen J, Rieppo L,
Saarakkala S, Nissi MJ. Orientation anisotropy of quantitative MRI relaxation
parameters in ordered tissue. Sci Rep. 2017;7(1):9606.
doi:10.1038/s41598-017-10053-2
17. Hänninen NE, Nykänen O, Prakash M,
Hanni M, Nieminen MT, Nissi MJ. Orientation anisotropy of quantitative MRI
parameters in degenerated human articular cartilage. Journal Orthopaedic
Research. 2021;39(4):861-870. doi:10.1002/jor.24778
18. Zaiss M, Zu Z, Xu J, et al. A combined
analytical solution for chemical exchange saturation transfer and semi-solid
magnetization transfer. NMR Biomed. 2015;28(2):217-230.
doi:10.1002/nbm.3237
19. Wáng YXJ, Zhang Q, Li X, Chen W, Ahuja
A, Yuan J. T1ρ magnetic resonance: basic physics principles and applications in
knee and intervertebral disc imaging. Quantitative imaging in medicine and
surgery. 2015;5(6):858.
20. Pang Y. An order parameter without
magic angle effect (OPTIMA) derived from dispersion in ordered tissue. Magnetic
Resonance in Med. 2020;83(5):1783-1795. doi:10.1002/mrm.28045