Dario Bosch1,2, Georgiy Solomakha1, Martin Freudensprung3, Felix Glang1, Nikolai Avdievich1, and Klaus Scheffler1,2
1MPI for Biological Cybernetics, Tübingen, Germany, 2University Hospital Tübingen, Tübingen, Germany, 3Institute of Neuroradiology, Universitätsklinikum Erlangen, Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen, Germany
Synopsis
Keywords: Parallel Transmit & Multiband, High-Field MRI
Motivation: Dipole RF elements with electronically switchable B1 field patterns have been shown to improve receive performance. The question whether they are beneficial for transmission is still open.
Goal(s): Improve flip angle homogeneity by modulating the Tx sensitivity of a dipole during excitation.
Approach: An RF coil built from eight dipoles with electronically switchable sensitivities was constructed. Achievable flip angle homogeneity with 2-kT-points pulses was evaluated in simulations and in experiments.
Results: Flip angle homogeneity could be increased by electronically switching the Tx sensitivity during the course of the RF pulse.
Impact: RF elements with switchable transmit sensitivities offer a novel
degree of freedom for excitation that promises improved flip angle
homogeneity. This addresses one of the most pressing problems in
ultra-high field MRI.
Introduction
The concept of rapidly reconfigurable radio frequency (RF) receiver
elements for ultrahigh-field MRI offers improved parallel imaging
performance1. In this study, a first attempt is made to use fast
switching coaxial dipoles for transmission.
In contrast to the receive application, the use of fast switching
transmit coils aims at improving flip angle (FA) homogeneity, which
is a major challenge in ultrahigh-field MRI. In a previous study 2,
we evaluated fast switching dipoles in simulations (Figure 4). Here
we present first experimental results, both in a phantom and in-vivo.Coil design
Dipole elements with a circuit that
allows switching
between inductive
and capacitive impedance
were designed similar to1, closely following3.
The current distribution in the dipoles is biased toward the
inductive end, resulting in a spatial shift of the B1+
sensitivity in the same direction (Figure
1).
In
a simulation study, four
different head array
coils
with eight dipoles each were evaluated, varying the dipole
length between 17 cm and 19 cm, and
including or omitting a reflector above the dipoles.
Simulations were performed for 400 MHz in CST Studio using the Duke
voxel model and resulted in 2 sets of sensitivity profiles for each
coil: “up” and “down”, where all sensitivity profiles were
shifted in the cranial/caudal direction (Figure 2). Based
on the simulation results, a 17 cm array with reflector was
constructed. A custom CMOS
driver was used for switching the dipoles between up and down
configuration.Pulse design and MRI measurements
Experiments were performed on a
Siemens 9.4 T whole-body MRI scanner. Switching between the up
and down state was controlled
from the sequence via an optical trigger signal. All MR sequences
were written in a modified version of pulseq that
allows pTx RF-shimming for
each RF pulse4.
Single-channel B1+
and B0
maps were acquired using a presaturated turbo-flash sequence with
weighted hybrid mapping5.
B1+
and B0
mapping was performed both in a head-shaped
agar phantom and in a healthy male subject.
A homogeneous FA distribution of 10° throughout the brain was
chosen as the optimization target. The normalized root mean square
error (nRMSE) of the obtained FA distribution was used as a quality
metric, excluding outliers. The optimization was performed similarly
to the spatial domain method6, using the variable exchange
method7 to solve the magnitude-least-squares problem with 20
iterations on a Tikhonov-regularized minimization. Optimization was
performed for the entire volume in the phantom and for the brain only
in-vivo.
For
excitation, 2-kT-points pulses
were designed.
Fast
switching during the pulse was implemented by changing the coil
profiles from up to
down configuration
after half the excitation time. In this case, magnitude and phase
were optimized for each subpulse.
The k-space location of the
first kT-point was chosen by randomly generating 25,000
k-space positions in the range 28 m-1 ≤ kx,y,z ≤ 28 m-1
and selecting the one
that gave the best performance for each
scenario. The second kT-point was positioned in the center of
k-space. For
comparison, a similar pulse was designed for the static
“down”-configuration without
fast switching.
For
evaluating excitation homogeneity experimentally
a gradient echo sequence (FA=2°,
TR=11 ms, TE=1.5 ms)
was used on the phantom.Results
In the agar phantom, optimizing the 2-kT-point pulses resulted in an
nRMSE of 17.12% and 15.61% for the static and the fast-switching
scenario, compared to 80.82% in CP-mode.
The pulses were
played out successfully in a gradient echo sequence and the resulting
images matched the simulated FA distributions well (Figure 3).
When optimizing
pulses for the in-vivo data, nRMSE was reduced to 12.37% and 11.59%
for static and fast-switching, compared to 22.35% in CP-mode (Figure
5).Discussion and Conclusion
In simulations (Figure 4), the introduction of transmit elements with
rapidly switchable B1+
sensitivities allows for a significant reduction in FA inhomogeneity
in ultrahigh-field MRI. Compared to the simulation results the effect
of fast switching stays well below expectations both in-vivo and in
the phantom. Compared to the previously published simulation results
the difference of up and down sensitivities is much lower under
experimental conditions, which needs to be investigated next. We
were, however, able to demonstrate that the concept works not only in
simulations but also experimentally.
Similar to the
observations made with reconfigurable Rx elements, switching between
different Tx sensitivity patterns within a single element can be seen
as a way to effectively emulate a larger number of independent
virtual Tx elements. Considering the cost and technical
complexity of increasing the number of Tx channels, the use of
rapidly switchable B1+
sensitivities in transmit elements should be further investigated.Acknowledgements
Funding by the Deutsche Forschungsgemeinschaft (DFG - German Research Foundation) under the Reinhart Koselleck Programme (DFG SCHE 658/12) and by the European Research Council (ERC Advanced Grant No 834940, SpreadMRI) is gratefully acknowledged.References
1. G.A. Solomakha et al 2023; ISMRM #5070
2. D. Bosch et al
2023; CMRR Workshop on UHF
Imaging
3. G.A. Solomakha et al 2023; ISMRM #1061
4.
M. Freudensprung et al 2023; ESMRMB #LT53
5. D. Bosch et al 2022; ISMRM #2865
6. W. Grissom et al 2006; Magn. Reson. Med., 56: 620-629
7.
K. Setsompop et al 2008; Magn. Reson. Med., 59: 908-915