Nicolas Boulant1, Caroline Le Ster1, Alexis Amadon1, Guy Aubert2, Alexander Beckett3,4, Jean Belorgey2, Cédric Bonnelye1, Dario Bosch5,6, David Otto Brunner7, Guillaume Dilasser2, Olivier Dubois2, Philipp Ehses8, David Feinberg3,9, Sajjad Feizollah10, Vincent Gras1, Simon Gross7, Quentin Guihard2, Hervé Lannou2, Denis Le Bihan1, Franck Mauconduit1, Frédéric Molinié2, François Nunio2, Klaas Pruessmann11,12, Lionel Quettier2, Klaus Scheffler5,13, Tony Stöcker8, Christine Tardif10, Kamil Ugurbil14, Alexandre Vignaud1, An Vu15,16, and Xiaoping Wu14
1NeuroSpin, CEA, Gif sur Yvette, France, 2Irfu, CEA, Gif sur Yvette, France, 3Advanced MRI technologies, Sebastopol, CA, United States, 4Helen Wills Neuroscience institute, University of California, Berkeley, CA, United States, 5Department for biomedical Magnetic Resonance, University of Tübingen, Tuebingen, Germany, 6High-field MR center, Max Planck Institute for biological cybernetics, Tuebingen, Germany, 7Skope MRT, Zürich, Switzerland, 8DZNE, Bonn, Germany, 9Helen Wills Neuroscience institute, Universiy of California, Berkeley, CA, United States, 10Montreal Neurological Institute-Hospital, McGill University, Montreal, QC, Canada, 11ETH Zürich, Zürich, Switzerland, 12University of Zürich, Zürich, Switzerland, 13High field MR center, Max Planck Institute for biological cybernetics, Tuebingen, Germany, 14Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States, 15University of California, San Francisco, CA, United States, 16San Francisco VA Health Care System, San Francisco, CA, United States
Synopsis
Keywords: Gradients, Gradients
Motivation: Third order shim coils can impact gradient-magnet interactions with consequences on image quality and magnet safety.
Goal(s): To demonstrate the influence of third order shim coils on the SC72 gradient coil using field and vibration measurements.
Approach: The gradient transfer function was measured at 11.7T (Iseult) and at 7T (Terra) with and without connection of the 3rd order shim coils. Vibration measurements were carried out on Iseult from 0 to 11.7T in the two configurations as well.
Results: The data demonstrate a drastic influence of the 3rd order shim coils and their circuits.
Impact: The work suggests caution when using third
order shim coils at 7T and above. It also paves the way for further
investigations to improve gradient waveform fidelity.
Introduction
Higher
magnetic fields and stronger gradients intensify their interaction and put
higher demands on the equipment1,2. Besides hardware damage,
which is of tremendous importance, these interactions can also lead to field
perturbations that can affect data quality. Iseult is a 11.7T whole-body
scanner equipped with the SC72 whole-body gradient coil (Siemens Healthcare,
Erlangen, Germany), which embeds second and third order shim coils. In a long
test-campaign aimed at characterizing the gradient-magnet interactions on that
system, it was observed that the connection of the third order shim coils all
the way to their shim amplifiers had a significant impact on the results. To
rule out possible causes related to magnet design, it was decided to repeat
some of the measurements on other systems with the identical gradient coil.Methods
Gradient
Transfer Functions (GTF) were characterized using a field camera (Skope MRT,
Zürich, Switzerland) on the SC72 at 11.7T (Iseult) and at 7T (Terra). Both
systems have the same versions of the electronic chain and scanner software
electronic chain generation. The measurements were performed with and without
the 3rd order shim coils connected to the filter panel inside the Faraday cage.
For further confirmation, the same GTF measurements were carried out on other
7T, 9.4T and 10.5T systems with the same SC72 gradient coil, all with their 3rd
order shim coils disconnected. Vibrations were also measured with mono-axial
accelerometers (B&K, Naerum, Denmark) glued to the flange of the gradient
coil in Iseult from 0T to 11.7T, in steps of 1T, and in the two 3rd order shim
configurations (connected and disconnected). Results
Figure 1 reports the GTF (magnitude and phase)
at 11.7T (Iseult) and 7T (Terra) around the most intense resonance peak at 1350
Hz on the gradient Z axis, with and without connection of the 3rd order shim
coils, revealing drastic differences. Figure 2 likewise shows the cross terms
(Z->3rd order spherical harmonics) characterized by a dynamic field camera
at 11.7T, indicating vibration-induced currents circulating in some 3rd order
shim coils when running currents in the Z gradient coil, as confirmed by
current measurements on the individual shim coils on Figure 3. Figure 4 reports
the GTF self-term characterizations on the other 7T, 9.4T and 10.5T systems with
3rd order shim coils disconnected. Finally Figure 5.a shows the acceleration spectrum
for the gradient Z axis at 11.7T with 3rd order shim coils connected and disconnected.
Figure 5.b reports the normalized height of the main peak acceleration at 11.7T
with respect to B0, again showing drastic differences regarding the
3rd order shim coil configurations.Discussion
In this work, we demonstrate that the 3rd order
shim coils and their associated circuits greatly impact gradient-magnet
interactions and that these effects were consistently observed for various
magnet designs (7T Magnetom and Terra) and field strengths (7, 9, 10.5 &
11.7 T). In addition to the potential hardware damage, peaks and dips in the
GTF have been shown to affect gradient waveform fidelity in MRI operation3,4 and thus can deteriorate image quality. Although the phenomenon to date is
still not completely understood, the connection of the 3rd order shim coils
presumably provides a path for the current generated by vibrations to circulate
and create field disturbances. Disconnecting the 3rd order shim coils greatly
improved the quality of the gradient response. Surprisingly, more vibrations of
the gradient coil (Figure 5) could be observed when the 3rd order shim coils
were disconnected. When connected, vibrations reached a plateau at around 8T,
indicating increased damping. The data suggests that this saturation behavior
reflects more interactions between the gradient and the third order shim coils
and that vibrations proportional to B0 may in fact be desirable to
indicate less interactions.Conclusion
For the setups investigated in this work, 3rd
order shim coils and their associated circuits were shown to have great impact
on gradient-magnet interactions and gradient waveform fidelity. Although this
work focused on the Z gradient axis, similar behaviors were observed for the X
and Y axes.Acknowledgements
AROMA
H2020 FET-Open (885876). ANR-21-ESRE-0006.
ERPT equipment program of the Leducq Foundation. NIH grants NIBIB P41 EB027061, U01 EB025144, and S10 RR029672.References
[1] Boulant N, Quettier L et al. Commissioning of the Iseult CEA 11.7
T whole-body MRI: current status, gradient-magnet interaction tests and first
imaging experience. MAGMA 2023 Apr;36(2):175-189. doi:
10.1007/s10334-023-01063-5.
[2] Winkler S.A,
Schmitt F, Landes H, DeBever J, Wade T, Alejski A, Rutt B.K. Gradient and shim
technologies for ultra-high field MRI. NeuroImage 2018;168:59-70.
[3] Vannesjo SJ, Haeberlin M, Kasper L, Pavan
M, Wilm BJ, Barmet C, Pruessmann KP. Gradient system characterization by
impulse response measurements with a dynamic field camera. Magn Reson Med 2013;
69:583-593.
[4] Vannesjo SJ, Graedel NN, Kasper L, Gross S,
Busch J, Haeberlin M, Barmet C, Pruessmann KP. Image reconstruction using a
gradient impulse response model for trajectory prediction. Magn Reson
Med 2016; 76:45-58.