Rita Schmidt1 and Amir Seginer2
1Department of Brain Sciences, Weizmann Institute of Science, Rehovot, Israel, 2Life Sciences Core Facilities, Weizmann Institute of Science, Rehovot, Israel
Synopsis
Keywords: fMRI Acquisition, fMRI (task based), Auditory stimuli
Motivation: With improved scanning methods and SNR, auditory response studies in functional MRI require well-defined stimuli but are encumbered by the acoustic noise from the acquisition gradients and by the temporal inaccuracy of the external audio source.
Goal(s): Our goal was to use the acoustic noise generated by the MRI gradients to circumvent an external source for auditory fMRI.
Approach: We implemented BEBOP (BEeping BOld Pulse sequence) EPI which enables varying the echo spacing, and thus pitch, per slice and per repetition and so provides a platform for auditory fMRI.
Results: An fMRI feasibility study at 7T was successfully implemented, measuring auditory-motor responses.
Impact: We demonstrated a new approach that simultaneously generates auditory stimuli and measures their BOLD response. The new approach offers
high temporal accuracy of the auditory tasks due to the MRI gradients’ high
temporal precision.
Introduction
One of the first auditory fMRI experiments performed1 used as
stimuli the acoustics generated by gradients played (or not) before the
excitation. This resulted in a simple experiment with high temporal accuracy.
Later studies turned to external stimuli, enabling great versatility of the
audio stimuli. However, the latter is confounded by acoustic noise from the
scan itself2,3 and by the limited synchronization accuracy of the
external audio source. With improved scanning methods and SNR, especially at
ultra-high field, auditory fMRI studies require a well-defined stimuli. Suggested solutions include attenuation of
the acoustic noise4 and sparse fMRI with embedded breaks for the
stimuli2,3. We implemented a new approach employing the inherent
acquisition acoustics to overcome the above disadvantages. Our BEBOP (BEeping
BOld Pulse sequence) EPI utilizes controllable varying echo spacings (ESPs)
that change the audio pitch per acquisition (changing with slice and
repetition) and so provides a platform for planning auditory tasks. A feasibility study with a block-design auditory
response was performed at 7T, using blocks of fixed and randomized ESPs as
reference and “task” blocks, respectively. Methods
Pulse sequence with varying auditory pitch. Controllable varying ESPs were
implemented by enabling delays around the EPI readout gradients (while maintaining
the readout bandwidth).
Stimuli and task block-design. Two blocks were defined - a
‘fixation’ block with a fixed pitch/ESP and a ‘task’ block with a preset
randomization of pitches/ESPs. A ‘fixation’/‘task’ block-design of either 8/6
or 16/16 seconds was executed for ~6 minutes with a rate of 12 slices/2 sec.
During some experiments, the volunteers were asked to tap with their 4 fingers
but only when hearing the randomized-pitch.
Varying ESPs and signal variation. Preliminary testing on a phantom
showed signal variation when varying ESPs. Three possible factors are i)
changes in image distortions, ii) differing ghost artifacts, and iii) eddy
currents effects. Ghost artifact were found to be minimal when the navigators’
ESPs changed with the acquisitions’. To circumvent undesired signal changes in
an actual fMRI experiment, a subset of slices of interest were forced to
maintain the ‘fixed-pitch’ ESP even in the ‘randomized-pitch’ block (see
Fig.1). The slices of interest were chosen as slices through the
auditory cortex and optionally through the motor cortex.
Scan parameters. Scans of healthy volunteers were acquired with a 7T Terra
scanner (Siemens). TR=2 sec, TE=22 ms, BW/Pixel=2480 Hz/pixel, x3 acceleration,
in-plane resolution 2.1x2.1 mm2, 12 slices with 2.2 mm slice
thickness, fixed ESP=0.53 ms, random ESP values range 0.53-0.8 ms. Two sets of
scans were performed: Set A, a scan centered around the primary auditory
cortex having 4 central slices (5 to 8) with fixed ESP (no gap between slices),
and Set B, a scan covering both the primary auditory and the motor
cortices (150% gap between slices, slices 3,4,10,11,12 with fixed ESP). Results
Fig.1a-b show the time varying ESPs, per slice, for Set A
and for Set B. Fig.1c shows the main expected
sound frequency due to the ESPs (freq=1/2ESP) as function of time.
Fig.1d shows the recorded audio amplitude during 18 seconds of a Set A scan.
Fig.2 shows image distortions for different ESPs. Only minor
distortions were observed.
Fig.3 shows the standardized signal changes (ΔS) during a BEBOP
scan. Fig.3a and 3b show ΔS in slices with fixed ESPs for voxels with t-test>2.5,
while Fig.3c and 3d show ΔS of a set of voxels in a slice with
varying ESPs. The plots show ΔS along the 6 min scan and summed over block repetitions. The
signals are compared to a simulated hemodynamic response (HRF) of a matching
“OFF/ON” block design. ΔS and the simulated HRF in Fig.3b are
highly correlated (r=0.7) while in Fig.3d the signal is in correlation to the
variation of the ESPs and not in correlation to the simulated HRF response.
Fig.4 shows the
t-test overlaid on EPI images for the slices of interest. Set A exhibits
high scores at the primary auditory cortex. Set B shows high t-test
scores for both auditory and motor regions.
Fig.5 compares t-test
maps for listening-only and for combined auditory and motor tasks. High
activation is shown in the basal ganglia in Fig.5b and in the motor and
supplementary motor areas in Fig.5c.
Conclusions
We demonstrated a new approach that simultaneously generates
auditory stimuli and measures their BOLD response. The new approach offers
high temporal accuracy of the auditory tasks due to the MRI gradients’ high
temporal precision. A block design
feasibility study was successfully implemented, measuring auditory-motor responses.Acknowledgements
Special thanks to Santosh Kumar Maurya for artistic visualization of the BEBOP scan.References
1.
Bandettini,
P. A., Jesmanowicz, A., Van Kylen, J., Birn, R. M., & Hyde, J. S. (1998).
Functional MRI of brain activation induced by scanner acoustic noise. Magnetic resonance in medicine, 39(3), 410-416.
2.
Peelle,
J. E. (2014). Methodological challenges and solutions in auditory functional
magnetic resonance imaging. Frontiers in neuroscience, 8, 253.
3.
Gonzalez-Castillo,
J., Olulade, O. A., & Talavage, T. M. (2012). Using functional MRI to study
auditory comprehension. Imaging in Medicine, 4(1), 137.
4.
Seifritz,
E., Di Salle, F., Esposito, F., Herdener, M., Neuhoff, J. G., & Scheffler,
K. (2006). Enhancing BOLD response in the auditory system by
neurophysiologically tuned fMRI sequence. Neuroimage, 29(3), 1013-1022.