Brian Rutt1, Peter Roemer2, Andrew Alejski3, Trevor Wade3, Matthew Bester3, Koray Ertan4, Alexander Bratch5, Gregor Adriany5, and Kamil Ugurbil5
1Radiology, Stanford University, Stanford, CA, United States, 2Roemer Consulting, Lutz, FL, United States, 3Robarts Research Institute, University of Western Ontario, London, ON, Canada, 4Stanford University, Stanford, CA, United States, 5CMRR, University of Minnesota, Minneapolis, MN, United States
Synopsis
Keywords: Gradients, Gradients, Peripheral nerve stimulation, magnet heating
Motivation: Two interactions between gradient coils and their environment severely limit imaging performance: stimulation of peripheral nerves in human subjects and deposition of heat energy into the superconducting main magnet leading to helium loss or quench. This heating increases significantly for higher B0 and gradient switching frequencies.
Goal(s): To address the above limitations by recasting the gradient design problem in a unique way to simultaneously minimize both of these interactions.
Approach: Using these new methods, we designed and modeled a head gradient coil: H12.
Results: We achieved >1.4 fold PNS improvement and >6 fold lower magnet heating compared to existing state-of-the-art head gradient coils.
Impact: Our method allows the design and
real-world use of stronger and faster-switching gradients than any in existence,
while minimizing peripheral nerve stimulation and excessive magnet heating. The
potential impact is especially high for head gradients operating at ultra high
field.
Introduction
As a result of the push for ever-higher gradient performance, there are
two problematic interactions that now limit operational performance: peripheral
nerves stimulation (which can be painful), and excessive power deposition into
the superconducting main magnet). These problems are a particular concern for
head gradients operating at high switching frequencies and ultra high B0.
The gradient system operating limits can be adjusted to avoid these problematic
interactions, at the expense of imaging performance. Our goal was to address
these problems more fundamentally, by recasting the gradient design problem to minimize
these interactions and to allow for the creation of optimal gradient designs
that constrain both peripheral nerve stimulation and magnet heating.Methods
Our minimum-electric-field gradient design method1 minimizes the peak electric
field magnitude (Emax) on the surface of simplified body models
positioned in the gradient coil. Emax per unit slew rate is an
accepted predictor of PNS thresholds according to regulatory standards2 and our own work3.
Our method can be used to optimize gradient designs over a large and diverse
set of body models, thereby achieving global population-wide PNS optimization.
Our magnet-heating-minimized gradient design concept is
summarized as follows. We represent the gradient fringe field by a current
density pattern on an equivalent surface, which can be decomposed into a set of
basis functions. Each basis function is then propagated through a multiphysics gradient-cryostat-interaction
(GCI) model that couples the mechanical vibrations with the induced currents
that result in frequency-dependent power deposition into the cryostat. These
basis functions and associated power deposition matrices are then incorporated
into our gradient design code.
Figure 1 illustrates these new design concepts, which serve
to optimize the gradient coil design to meet desired gradient performance
constraints while simultaneously minimizing stored magnetic energy in the
gradient coil (slew rate), peak electric field impinging on the body (PNS), and
power deposited into the magnet (GCI).
We employed the above novel concepts to design
the H12 head gradient. We set the following design targets: inner diameter
400mm, outer diameter 616mm, length 1050mm, shoulder corner to isocenter
distance 170mm, imaging region diameter 240mm, target gain (sensitivity) ≥220
µT/m/A, target inductance ≤440 µH, target Emax ≤7mV/m per T/m/s,
target 10.5T magnet heating ≤6W for 40mT/m gradient amplitude, target gradient
cooling ≥40kW. These targets would produce gradient performance parameters of Gmax
264mT/m at 1200A, Smax 1000T/m/s at 2000V, with PNS thresholds ≥1.4
fold higher and magnet heating ≥5 fold lower than existing head gradients.Results
Figure 2 shows the optimized H12 X, Y and Z wire patterns,
with primary windings in red and shield windings in yellow.
Figure 3A shows H12 X, Y and Z wire patterns superimposed on
a body model which displays surface electric fields. Emax values for
X, Y, and Z gradient axes are 6.5, 6.0, 7.0 mV/m, respectively. Figure 3B shows
PNS thresholds plotted in ∆Gstim vs ∆t space4, indicating the higher
predicted PNS thresholds for H12 compared to three other head gradients and one
body gradient3,5.
Solid lines indicate measured thresholds whereas dotted lines represent Emax-predicted
thresholds, showing the close match to measured thresholds whenever both were
available.
Figure 4 shows the calculated power deposition into the
magnet for 40mT/m sinusoidal gradient amplitude, spanning frequencies from 500
to 3000 Hz, for a head gradient with design parameters equivalent to H12 but
without magnet-heating minimization (solid curves), compared to the magnet-heating-minimized
H12 design (dashed curves). Large reductions of power deposition are shown: ~10-fold
for X and Y and ~6-fold for Z.
Figure 5A shows the construction concept in CAD
design view, with a foundation layer representing the inner shell (cylindrical
main section plus conical shoulder section) plus bent rungs that locate the
X-primary wires. Y and Z primary layers are added via the placement of similar
bent rungs on top of the previously wound layer. Shield windings are laid into straight
rungs. All support parts are 3D printed, and all conductors are doubly-insulated
6.5mm OD hollow copper conductor. Figure 5B shows the early construction phase
of H12, in which the inner shell including X-primary rungs has been printed and
X-primary wires have been wound.Discussion and Conclusion
We have designed, modeled, and begun to build the H12 head gradient
coil, which is intended for use in the 10.5T magnet at CMRR. Gradient
performance is among the highest of any demonstrated to date, with significant
improvements in strength, slew rate, cooling, PNS and GCI performance over
other gradients; these are the result of our ability to globally optimize
gradient design subject to Emax and magnet-heating constraints.Acknowledgements
The authors gratefully acknowledge research support from NIH U01
EB025144 and NIH R01 EB025131. We also acknowledge support from the Sim4Science
program at ZurichMedTech.References
1. Roemer, P.B. & Rutt, B.K. Minimum
electric-field gradient coil design: Theoretical limits and practical
guidelines. Magn Reson Med 86, 569-580 (2021).
2. IEC. Medical electrical equipment –
Part 2-33: Particular requirements for the basic safety and essential
performance of magnetic resonance equipment for medical diagnosi. International Electrotechnical Commissioin 60601-2-33 Edition 3.2(2015).
3. Roemer, P.B., Wade, T., Alejski, A.,
McKenzie, C.A. & Rutt, B.K. Electric field calculation and peripheral nerve
stimulation prediction for head and body gradient coils. Magn Reson Med 86,
2301-2315 (2021).
4. Chronik, B.A. & Rutt, B.K. Simple
linear formulation for magnetostimulation specific to MRI gradient coils. Magn Reson Med 45, 916-919 (2001).
5. Davids,
M., et al. Peripheral nerve
stimulation informed design of a high-performance asymmetric head gradient
coil. Magn Reson Med 90, 784-801 (2023).