John P. Mugler III1
1University of Virginia, United States
Synopsis
Keywords: Image acquisition: Sequences
This
presentation provides an in-depth discussion of key features that make turbo/fast
spin echo a robust imaging method, including configuration of the phase-encoding
gradients, selection of the RF-pulse phases, effects of signal decay during the
echo train, and flip angles other than 180°
for the refocusing RF pulses.
The pulse sequence commonly
referred to as turbo spin echo (TSE) or fast spin echo (FSE) is widely used for
applications from head to toe. It is
truly a “bread & butter” technique of clinical MRI. TSE/FSE pulse sequences provide a wide range
of clinically-useful contrast behaviors, such as T1, T2 and proton-density weighting,
fat-signal suppression (“STIR”) and fluid-signal suppression (“FLAIR”). First described in 1986 by Hennig and
colleagues1, subsequent research on optimization and refinement of
this method, especially in the late 1980s and early 1990s, led to the
techniques widely used in clinical practice today.
This
presentation will take a “deep dive” into the TSE/FSE pulse sequence, providing
an in-depth discussion of key features that make TSE/FSE a robust imaging
method. The specific aspects that will
be explored include: configuration of the phase-encoding gradients, selection
of RF-pulse phases, impact of the signal decay/evolution during the echo train
on image contrast and artifacts, and the use of flip angles other than 180° for the refocusing RF pulses, including
continuously varying flip angles.Acknowledgements
No acknowledgement found.References
1. Hennig J, Nauerth A, Friedburg H. RARE imaging: a fast
imaging method for clinical MR. Magn Reson Med. 1986; 3:823-833.