YASUO TAKATSU1,2, Masafumi Nakamura2,3, Yuichi Suzuki4, and Tosiaki Miyati2
1Molecular Imaging, School of Medical Sciences, Fujita Health University, Toyoake, Japan, 2Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, Kanazawa, Japan, 3Department of Radiology, Otsu City Hospital, Otsu, Japan, 4Department of Radiology, The University of Tokyo Hospital, Tokyo, Japan
Synopsis
The position dependence of the ADC
in magnetic resonance imaging (MRI) by EPI- and TSE-DWI was assessed using
phantoms based on the relationship between SNR and B1 maps. ADC decreased with
distance from the center of the magnetic field for both EPI-DWI and TSE-DWI. The Pearson correlation coefficient between ADC and FA
was strong and more significant than between
ADC and SNR. ADC depends on the slice position and decreases with an
increase in distance from the magnetic field center. Caution should be taken
when comparing and quantitatively evaluating the ADC at sites shifted in the
long-axis direction.
Background
Single-shot echo-planar imaging (EPI)
is widely used in diffusion-weighted imaging (DWI); however, switching the
diffusion gradient on and off to induce the currents (eddy currents) of the
magnetic field that is changing with time, resulting in a non-uniform magnetic
field [1] that causes artifacts, which affect the apparent diffusion
coefficient (ADC) [1]. Hence, DWI with turbo spin echo (TSE), which is superior
to magnetic field inhomogeneity, is used as an alternative when artifacts
appear with EPI [2]. There were no significant differences between EPI and TSE
in terms of the calculated ADC [3]. On the other hand, ADC has been reported to
reduce with a decrease in signal-to-noise ratio (SNR) [4, 5]. In
other words, when quantitatively assessing the brain regions with ADCs, it is
necessary to recognize the influence of the position. When set up from the
lower cerebellum to the parietal region, as in a typical MRI scan of the brain,
ADC values may vary depending on the slice position; however, the extent of
this variation has not been clarified. If the ADC changes due to slicing, it
may interfere with quantitative assessment and site-by-site comparison.Purpose
The position dependence of the ADC
in magnetic resonance imaging (MRI) by EPI- and TSE-DWI was assessed using
phantoms based on the relationship between SNR and B1 maps.Methods
Five pure water-filled containers
were placed parallel to the direction of the static magnetic field from the center of the magnetic field to the foot
direction and a cross-section image was scanned at the center of each
container using a 1.5-T MRI scanner. Five
slice positions were set for each container (from the center of the static magnetic field and receiving coil "position A" to the direction of the foot "position E", the slice interval is appoximately 33.5 mm.). Diffusion times
for both EPI- and FSE-DWI were matched. The slice thickness was adjusted to
match the SNR at the center of the magnetic field for both sequences. A B1
map was analyzed. Pearson correlation
coefficients between ADC and SNR and between ADC and flip angle (FA) were
calculated.
Results
ADC decreased with distance from
the center of the magnetic field for both EPI-DWI and TSE-DWI. The Pearson correlation coefficient between ADC and
SNR was 0.77 in EPI-DWI and 0.54 in TSE-DWI, there was no significant difference between ADC and SNR in
EPI-DWI (P = 0.07) and TSE-DWI (P
= 0.21), whereas that
between ADC and FA was 0.99 in EPI-DWI and 0.96 in TSE-DWI, there were significant
differences (P <0.01). The FA decreased by 0.047 degrees per mm from the center of the
magnetic field to foot direction.Discussion
The measured SNR in the present study was affected by
the uniformity of the static magnetic field and the sensitivity of the
receiving coil. The ADC is expected to
be involved in the linearity of the static magnetic field or the gradient magnetic field. From these results, we
considered that SNRs may be involved in ADC changes but not necessarily in a
strong way.Conclusion
ADC depends on the slice position
and decreases with an increase in distance from the magnetic field center.
Caution should be taken when comparing and quantitatively evaluating the ADC at
sites shifted in the long-axis direction. The effect of B1 inhomogeneity
was more significant than SNR.Acknowledgements
No acknowledgement found.References
[1] Le Bihan D, Poupon C, Amadon A, Lethimonnier F. Artifacts
and pitfalls in diffusion
MRI. J Magn Reson Imaging. 2006;24:478–88.
[2] Baltzer PA, Renz DM, Herrmann KH, Dietzel M, Krumbein I, Gajda M, et al. Diffusion-weighted imaging (DWI) in MR mammography (MRM):
clinical comparison of echo planar imaging (EPI)
and half-Fourier single-shot turbo spin echo (HASTE) diffusion
techniques. Eur Radiol. 2009;19:1612–20.
[3] Mikayama R, Yabuuchi H, Sonoda S, Kobayashi K,
Nagatomo K, Kimura M, et al. Comparison of intravoxel incoherent motion diffusion-weighted imaging between turbo spin-echo and echo-planar imaging of the head and neck. Eur Radiol. 2018;28:316–24.
[4] Mazaheri Y, Vargas HA, Nyman G, Shukla-Dave A, Akin O, Hricak H. Diffusion-weighted MRI of the prostate at 3.0 T:
comparison of endorectal
coil (ERC) MRI and phased-array coil (PAC) MRI-The impact of SNR on ADC measurement. Eur J Radiol. 2013;82:e515–20.
[5] Alsop DC. Phase
insensitive preparation of single-shot RARE: application to diffusion imaging
in humans. Magn Reson Med 1997;38:527–33.