Thomas L Chenevert1, Yuxi Pang1, Debosmita Biswas2, Ramesh Paudyal3, Amaresh Konar3, Jiachao Liang4, Lisa J Wilmes4, Nastaren Abad5, Luca Marinelli5, Humera Tariq1, Ajit Devaraj6, Dallas Turley7, Johannes M Peeters8, Nola M Hylton4, David C Newitt4, Savannah C Partridge2, Amita Shukla-Dave3,9, and Dariya Malyarenko1
1Radiology, University of Michigan Health System, Ann Arbor, MI, United States, 2Radiology, University of Washington, Seattle, WA, United States, 3Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 4Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA, United States, 5GE Research Center, Niskayuna, NY, United States, 6Clinical Science, Philips Healthcare, Highland Heights, OH, United States, 7Philips Healthcare, Bothell, WA, United States, 8Clinical Science, Philips, Best, Netherlands, 9Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
Keywords: Cancer, Diffusion/other diffusion imaging techniques, ADC measurement accuracy, system gradient nonlinearity correction, multi-center oncology imaging trials
Gradient
nonlinearity (GNL) induces spatial bias in diffusion b-value that confounds apparent
diffusion coefficient (ADC) measurements for anatomy offset from MRI scanner
isocenter. For emerging vendor-provided GNL correction (GNC) a standardized quality
control (QC) procedure is desired to streamline GNC application for multi-site imaging
trials that utilize ADC for tumor monitoring and therapy response assessment.
This QC procedure was developed and tested on four MRI scanner systems with
vendor-provided on-line ADC GNC for trial-specific phantoms and patient scans
for head-and-neck, breast, and myelofibrosis cancers.
Introduction
Ongoing
clinical trials evaluate apparent diffusion coefficient (ADC) for tumor
treatment response assessment and monitoring disease progression1,2. Gradient
nonlinearity (GNL) bias in b-value confounds ADC measurements for anatomy
offset from scanner isocenter3,4. Previous work has demonstrated the stability of
system GNL characteristics5 that warrants on-scanner GNL correction (GNC) for ADC based on vendor
specific gradient design parameters6. Recently emerging on-scanner ADC GNC
implementations require quality control (QC) procedures practical in clinical
trial settings for timely performance evaluation. This work demonstrates
the implementation of a standardized QC procedure for ADC GNC on clinical MRI
scanners across four cancer imaging trials (in breast, bone marrow and head
& neck) that use anatomy-specific DWI QC phantoms.Methods
On-scanner
GNC for clinical trials: ADC GNC7,8 was implemented by vendors on four clinical 3T MRI
scanners with distinct gradient models (Table 1). These scanners are used by Academic-Industrial
Partnership (AIP) imaging centers for their respective clinical trials that evaluate
ADC for: head & neck (H&N, cancer
therapy response)9,
myelofibrosis (MF, bone marrow cancer monitoring)10, breast cancer (contrast-free
diagnosis and therapy response assessment)1,2. The sites scanned their respective DWI
QC phantoms based on ice-water and ambient-temperature water in sodium polyacrylate gel (SPAAG,
“flood” phantom) and in polyvinylpyrrolidone (PVP, breast-phantom by CaliberMRI, Boulder,
CO) materials. The patient DWI scans were performed according to trial-specific
DWI protocols that included b-values=0,100,300,800s/mm2 with
single-shot (ss) or multi-shot (ms) EPI acquisitions.
ADC GNC
analysis:
ADC maps were generated on scanner consoles with and without GNC for phantoms
and trial subjects. The QC assessment tools were shared with the sites as
Matlab (R2019b, Mathworks, Natick, MA) p-code libraries. Sites calculated
fractional-bias maps, %ADC bias = 100% (ADC– ADCGNC) / ADCGNC
and b-value correction maps in scan DICOM coordinates8 in ITK meta-image
header (MHD) format. Qualitative comparisons were performed between fractional
bias and b-value correction maps to QC consistent GNC patterns. The constructed
ADC and gradient model-bias maps (MHDs) were shared with a central analysis site for
quantitative evaluation. To assess absolute accuracy, corrected ADCGNC
was compared to ground-truth ADC values11,12 for phantoms. For patients, the
fractional bias maps were compared to gradient-model b-value correction maps to confirm
adequate GNC performance.Results and Discussion
The fractional bias maps
illustrate the extent and spatial pattern of ADC non-uniformity bias corrected
by GNC for phantoms (Figure
1) and trial subjects (Figure 2). Adequate GNC deployment was ensured by the consistency
of fractional bias (Figures
1,2, middle rows) and system GNL model correction maps (Figures 1,2, bottom rows),
as well as small differences (<1% voxel-wise) between vendor-provided and
vendor-agnostic AIP GNC (not shown). GNC restored spatial uniformity for the phantom
ADC maps (Figure 1)
of ice-water (tubes), “flood” gel and fibro-glandular (FG) tissue-mimic (breast
PVP phantom background material). The GNL bias correction also improved cross-platform
reproducibility for the ice-water phantoms and accuracy of measured absolute diffusion
values for the phantom materials (ADCGNC±0.03)μm2/ms: ADCice-water(@0ºC)
= 1.11 μm2/ms;
ADCSPAAG(@isocenter) = 2.13 μm2/ms; and ADCPVP(%PVP0,
10, 25, 40) = (1.92, 1.58, 0.98, 0.60)μm2/ms in the PVP-DWI breast
phantom.
The spatial patterns of fractional
bias maps qualitatively agreed with the system model predictions for phantoms (Figure 1) and patients (Figure 2). Consistent
with the phantom ADC measurements, GNL bias and GNC efficiency increased with
the increasing offset from the scanner isocenter for target anatomy and toward
the FOV edges (Figure 2).
More structured noise was evident on patient fractional bias maps (Figure 2, middle row)
compared to phantoms (Figure
1, 3rd row), likely related to differences between background
low-SNR DWI filtering with and without GNC implemented for on-scanner ADC
calculation.
The range of the
corresponding GNL bias in ADC observed for phantoms and patients is summarized
in Table 1.
Phantoms provided a representative sample of imaged off-center locations for
testing the GNC with the range of corrected spatial GNL ADC non-uniformity bias
from -30% (Figure 1,
long-tube ice-water, SPAAG-flood) to +20% (Figure 1, SPAAG-flood, breast PVP) at extreme
locations. For all target anatomies mostly positive (moderate) bias
prevailed and was corrected (average of 3-12%, Table 1) in the patient ADC maps for breast, head
and neck and bone marrow (Figure
2). Conclusion
The
developed GNC QC procedure accommodates different anatomy and DWI phantom
protocols for a variety of clinical trials. It provides timely feedback on the
ADC GNC performance without interrupting the imaging trial workflow. Preliminary
results for the trial subjects indicate the adequate performance of the
vendor-implemented ADC corrections on all four studied gradient systems.
Evaluation of the GNC effect on ADC reproducibility and accuracy for trial
subjects is ongoing.Acknowledgements
Funding support from National Institutes of Health Grants: R01CA190299, R01CA207290, R01CA248192, R01CA132870, U01CA211205, U01CA225427, U24CA237683. References
1. Partridge, S.C.; Zhang, Z.; Newitt, D.C.; Gibbs, J.E.; Chenevert,
T.L.; Rosen, M.A.; Bolan, P.J.; Marques, H.S.; Romanoff, J.; Cimino, L.; et al.
Diffusion-weighted MRI Findings Predict Pathologic Response in Neoadjuvant
Treatment of Breast Cancer: The ACRIN 6698 Multicenter Trial. Radiology 2018,
289, 618–627
2. Rahbar H, Zhang Z, Chenevert TL, et al. Utility of
diffusion weighted imaging to decrease unnecessary biopsies prompted by breast
MRI: a trial of the ECOG-ACRIN Cancer Research Group (A6702). Clin Cancer Res.
2019;25(6):1756- 1765.
3. Newitt DC, Tan
ET, Wilmes LJ, Chenevert TL, Kornak J, Marinelli L, Hylton N. Gradient
nonlinearity correction to improve apparent diffusion coefficient accuracy and
standardization in the American College of Radiology Imaging Network 6698
breast cancer trial. J Magn Reson Imaging. 2015;42(4):908-19.
4. McTavish, S.,
Van, A.T., Peeters, J.M. et al. Gradient nonlinearity correction in
liver DWI using motion-compensated diffusion encoding waveforms. Magn
Reson Mater Phy 35, 827–841 (2022)
5. Pang Y, Malyarenko DI, Wilmes LJ, Devaraj A, Tan ET,
Marinelli L, Endt AV, Peeters J, Jacobs MA, Newitt DC, Chenevert TL. Long-Term
Stability of Gradient Characteristics Warrants Model-Based Correction of
Diffusion Weighting Bias. Tomography. 2022 Feb 4;8(1):364-375.
6. Pang, Y.; Malyarenko, D.I.;
Amouzandeh, G.; Barberi, E.; Cole, M.; Vom Endt, A.; Peeters, J.; Tan, E.T.;
Chenevert, T.L. Empirical validation of gradient field models for an accurate
ADC measured on clinical 3T MR systems in body oncologic applications. Phys.
Med. 2021, 86, 113–120
7. Tan, E.T.; Marinelli, L.;
Slavens, Z.W.; King, K.F.; Hardy, C.J. Improved correction for gradient
nonlinearity effects in diffusion-weighted imaging. J. Magn. Reson. Imaging
2013, 38, 448–453.
8. Malyarenko DI,
Ross BD, Chenevert TL. Analysis and correction of gradient nonlinearity bias in
apparent diffusion coefficient measurements. Magn Reson Med.
2014;71(3):1312-23.
9. Lu Y, Jansen JF,
Stambuk HE, Gupta G, Lee N, Gonen M, Moreira A, Mazaheri Y, Patel SG, Deasy JO,
Shah JP, Shukla-Dave A. Comparing primary tumors and metastatic nodes in head
and neck cancer using intravoxel incoherent motion imaging: a preliminary
experience. J Comput Assist Tomogr. 2013;37(3):346-52
10. Schaefer J, Choi
S, Luker G, Chenevert T, Ross B, Talpaz M. Primary myelofibrosis evolving to an
aplastic appearing marrow. Clin Case Rep. 2018 May 31;6(7):1393-1395.
11. Holz M, Heil SR, and Sacco A. Temperature-dependent
self-diffusion coefficients of water and six selected molecular liquids for
calibration in accurate 1H NMR PFG measurements. Phys. Chem. Chem. Phys. 2000; 2(20): 4740–4742.
12. Amouzandeh G, Chenevert TL, Swanson SD, Ross BD, Malyarenko DI:
Technical note: Temperature and concentration dependence of water diffusion in
polyvinylpyrrolidone solutions. Med Phys 2022; 49(5): 3325-3332