James L. Kent1, Ladislav Valkovič2,3, Iulius Dragonu4, Mark Chiew1,5,6, and Aaron T. Hess1
1Wellcome Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom, 2Oxford Centre for Clinical Magnetic Resonance Research, John Radcliffe Hospital, University of Oxford, Oxford, United Kingdom, 3Department of Imaging Methods, Institute of Measurement Science, Slovak Academy of Sciences, Bratislava, Slovakia, 4Research & Collaborations GB&I, Siemens Healthcare Ltd, Camberley, United Kingdom, 5Department of Medical Biophysics, University of Toronto, Toronto, ON, Canada, 6Physical Sciences, Sunnybrook Research Institute, Toronto, ON, Canada
Synopsis
Keywords: Motion Correction, Motion Correction
RF
sensor methods for motion correction require no (or minimal) additional
hardware, are sequence independent and have high temporal resolution. Several
RF sensor-based methods for rigid-body head motion detection have been
demonstrated but which method offers the most sensitivity to motion is yet
unknown. We aim to compare the sensitivity of PT and pTxS methods by simultaneously
measuring these signals during continuous motion, training a linear model from EPI
registered images and analysing the ability to predict rigid head positions. Currently, we see little difference in their ability to predict rigid head motion but further investigation is needed.
Introduction
RF
sensor methods for motion detection require no (or minimal) additional
hardware, are sequence independent and have high temporal resolution. Several RF
sensor-based methods for rigid-body head motion detection have been
demonstrated1,2 but which method offers the most sensitivity to
motion is yet unknown. Pilot tone (PT) uses an external ~90kHz off-resonance radiofrequency
source at a single frequency and is measured on the receive channels during
data acquisition and hence the signal appears in the oversampled region of
k-space. Parallel transmit scattering (pTxS) differs in that it reflects the electrical coupling between transmit channels during transmission
and is measured by the directional couplers (typically used for SAR
monitoring). In this work we compare the accuracy of PT and pTxS in predicting rigid-body head motion with a multivariate
linear regression model.Methods
Two subjects were scanned in a 7 Tesla MRI (VB17, Siemens Healthcare) equipped with
parallel transmit and imaged using an 8Tx/32Rx head coil (Nova Medical). Subject
A was guided via a projector to perform circular, up/down, and left/right motion
paradigms. Subject B performed similar head motions as subject A but with a
larger amplitude and additional roll motion. A segmented CAIPI 3D EPI
sequence was used to obtain 1200 volumes (TRvol = 256 ms). TE/TR = 6/16 ms, FA = 15°, BW = 2894 Hz/px, FOV = 250×250×125 mm3, slice thickness = 3.9 mm, scan matrix = 64×49×32
(including 75% partial Fourier), total 3D acceleration factor 8. PT and pTxS
signals were simultaneously measured. A water excitation pulse was overlaid
with a frequency-division multiplexed scattering matrix measurement3
(100 kHz offset; 10 kHz spacing; 15% pulse amplitude), giving 16 pTxS
measurements per image volume. PT was implemented by transmitting a pure sine
wave (90 kHz off resonance) using an external RF signal generator (Signal
Hound, VSG25A) with a nominal output power of -30 dBm attached to a dipole
antenna placed at the end of the patient table.
Reconstruction was performed using a SENSE reconstruction
in the BART toolbox4. Brain extraction and 3D rigid-body
registration were performed against the first image volume using FSL’s BET and
FLIRT tools5 to obtain the relative head position from the transformation
matrices. These were then lowpass filtered to remove sporadic erroneous measurements from registration. Eddy
current related phase shifts and +/- readout frequency changes in the PT signal
were corrected by multiplying each data point with the conjugate phase from a reference
(summation across all receive channels). PT signals were averaged for each EPI
shot to obtain 16 measurements per image volume. For pTxS,
the real and imaginary components of the (8×8) S-matrix were reshaped as a
vector (1×128) and used for training and evaluation of the linear model in
MATLAB (R2021a, MathWorks). For PT, the absolute value of the PT signal results in a (1×32) vector. A multivariate linear regression model was formed by fitting the first 400 EPI registrations (600
for Subject B) to the RF signals and the remaining 800 volumes were used to
test the model. The six degrees of motion were collapsed to a single global
RMSE6 score with an assumed head radius of 80mm. To prevent
overfitting, PT and pTxS signals were standardised, whitened using a singular
value decomposition (SVD), and truncated to reduce their dimensionality to rank
12. A combination of signals was also evaluated (1×160).Results
When
instructed to move, subjects A/B moved with an average velocity of 2.5/4.8mms-1
with translations and rotations up to 20/30mm and 7/10°. During a stationary
20s window, the RMSE for subjects A/B was 1.0/0.7mm for PT and 0.6/0.5mm for
pTxS. During motion, the ‘ground truth’ registration precision for subjects A/B
was estimated to be 0.5mm/0.9mm for translations and 0.2°/0.3° for rotations (determined
as the first order derivative/). The cumulative energy in the
first 12 singular values was 97% for pTxS, 99% for PT, and 97% for combined
signals. The RMSE for each degree of freedom is shown in Table 1 as well as the
average global RMSE which was
1.5/2.3mm for PT, 1.4/2.1mm for pTxS and 1.3/2.1mm for a combination of
these for Subjects A/B. The predicted motion traces from PT, pTxS and a combination of both are
shown in Figure 1 for each degree of freedom and correlate well with the ground
truth.Discussion
Both
PT and pTxS show an increased global RMSE for subject B. This is likely due to changes
in coil loading related to rigid motion; this is especially evident towards the end of the test
scans. Motion predictions using all methods
agree with EPI registered motion and where they do not, they tend to agree well
with each other, indicating true changes not captured by the rigid-body motion
model. Further work is required to interpret non-rigid body system
changes and refine the motion model to achieve the accuracy required for
in-vivo motion tracking. Going forward, we will also explore the sensitivity to
smaller amplitudes of motion.Conclusion
There
is high sensitivity to rigid body head motion in both PT and pTxS signals. Both
methods currently achieve a similar degree of accuracy, however, more investigation
is required to determine the sensitivity for smaller amplitudes of motion.Acknowledgements
JK acknowledges the support of EPSRC through an iCASE award
in collaboration with Siemens Healthcare Ltd.
The Wellcome Centre for Integrative Neuroimaging is supported
by core funding from the Wellcome Trust (203139/Z/16/Z). LV is funded by a Sir
Henry Dale Fellowship awarded jointly by the Wellcome Trust and the Royal
Society (221805/Z/20/Z) and also acknowledges the support of the Slovak Grant
Agencies VEGA (2/0003/20) and APVV (#19–0032). MC
is supported by the Canada Research Chairs Program.References
- Papp D, Jaeschke
S, Rieger SW, Clare S, Hess AT. Simultaneous detection of cardiac, respiratory
and rigid body head motion using the scattering of a parallel transmit RF coil
at 7T. In: Proc. Intl. Soc. Mag. Reson. Med. 26. 2018. p. 1168.
- Speier P, Fenchel
M, Rehner R. PT-Nav: a novel respiratory navigation method for continuous
acquisitions based on modulation of a pilot tone in the MR-receiver. In: Proc.
ESMRMB. Vol. 28. 2015. pp. 97–98.
- Jaeschke SHF,
Robson MD, Hess AT. Scattering matrix imaging pulse design for real-time
respiration and cardiac motion monitoring. Magn Reson Med 2019. 82:2169–2177.
- Uecker M, et al.
mrirecon/bart: version 0.7.00 (v0.7.00). Zenodo.
- Jenkinson M,
Bannister P, Brady M, Smith S. Improved Optimization for the Robust and
Accurate Linear Registration and Motion Correction of Brain Images. Neuroimage
2002;17:825–841.
- Jenkinson M. Measuring Transformation Error by RMS
Deviation. Intern Tech Rep TR99MJ1, FMRIB Centre, Univ Oxford 1999.