Lukas Hingerl1, Bernhard Strasser1, Gilbert Hangel2,3, Stanislav Motyka1, Fabian Niess1, Eva Niess1, Alexandra Lipka1, Dario Goranovic1, Philipp Lazen1, Stephan Gruber2, Siegfried Trattnig2,4, and Wolfgang Bogner1
1Department of Biomedical Imaging and Image-guided Therapy, Radiology and Nuclear Medicine, Medical University of Vienna, HFMR Centre, Vienna, Austria, 2HFMR Centre, Vienna, Austria, 3Department of Neurosurgery, Medical University of Vienna, Vienna, Austria, 4Institute for Clinical Molecular MRI in Musculoskeletal System, Karl Landsteiner Society, Vienna, Austria
Synopsis
Keywords: Shims, Magnets (B0)
Metamaterials enabling magnetic
replicators for $$$B_0$$$ shimming as well as for transcranial magnetic
stimulation may overcome traditional methods especially
concerning locality of field manipulation, field enhancement or cancellation in the distance in real-time by current tuning.
Introduction
Metamaterials
are assembled materials exhibiting uncommon physical behavior such as super-lenses1 or magnetic cloaks or concentrators2,3 and described by
negative electric/magnetic permittivity/permeability. While the
first can be manufactured, the features of the second example
can be only emulated, as described below. Rosa Mach-Batlle4,5 demonstrated experimentally
that by
emulating a negative µ-material the magnetic
field in free space can be manipulated (e.g. creating a magnetic dipole field) – without
violation of Maxwell’s equations or Earnshaw's Theorem.
We propose that this ‘magnetic replicator’ can be applied in
MRI, where $$$B_0$$$-inhomogeneities
(Figure 1) hamper the imaging quality. $$$B_0$$$-shimming
reduces such inhomogeneities, but doesn't change fields only locally. Hence, to goal is to
investigate the feasibility of a coil-insert which can generate a $$$B_0$$$
field by mimicking a 'virtual' source inside the object
under investigation.Theory and Method
Using
Pendry’s transformation optics6,7, for a very long cylindrical, shell
like geometry with µ=-1 (Figure 2) Mach-Batlle showed that the magnetic
field of an inner, source current inside the shell can be
replicated outside of the shell in a specific
distance d which is proportional to the shell’s inner/outer radii
ratio. A further equivalency was shown: A µ=-1 material can be
physically realized by an arrangement of DC-currents spatially oscillating their direction and positioned on the inner and outer surfaces of the shell
given by two surface current densities. A second equivalency showed that only one surface current density is
necessary, instead of two.
To be
applied for MRI we addressed the following
preliminary questions:
1) For
a $$$B_0$$$-inhomogeneity of ~200Hz (e.g. at $$$B_0$$$ = 3T), corresponding
to 5mT, can the required magnetic flux density be achieved?
2) What
is the optimal geometry of the device for brain MRI? Long shell like
structures as
previously
proposed are unpractical since they only manipulate the
field orthogonal along $$$B_{x,y}$$$, but $$$B_z$$$ needs to be changed. Is it
possible to revert the direction of the field replication to outside-to-inside to fit a human head?
3) What
is the influence of the static magnetic field $$$B_0$$$?
All
simulations were performed with COMSOL Multiphysics (magnetostatics module) based on finite-element
calculations.Results
As
described in the literature, magnetic field divergencies are created up to the coordinate point were the target field is replicated. This obeys Earnshaw's
theorem (no local field maximum in free space). Figure 3
shows exemplary $$$B_y$$$ fields of (from top to bottom): A single
current, µ=1 for wire and air; a theoretical
µ=-1 shell with an right-centered inner current and an outer $$$B_y$$$
field is replicated on the same position as in the example above; a
µ=1 shell with the surface currents as derived in4 and the same inner
current mimicking a µ=1 material; a µ=-1 material with an outer
current on the right which replicates the $$$B_y$$$ field to the inner shell
domain to the left; a µ=1 material with mimicking current
replicating the field twice from outer-right to inner-right
(weakened) as well as inner-left (weakened) which replicates to
outer-left (enhanced) and to the inner shell domain also. The
currents for these last case were derived.
Figure
4 shows the proposed device in a 2D-z-axisymmetric coordinate system. Only a single surface current
density is needed and no outer/inner current was used as
proposed in the shell case (last equivalency). Plotted
are the $$$B_z$$$ and $$$B_r$$$ fields ranging within the desired order of $$$10^1$$$mT.
The replicated field (~1cm apart from the surface) resembles the
field of a simple current quite well and similar as the field given in
Figure 1.
The
proposed spherical device geometry (Figure 5) consists of currents along rings with alternating
direction, with a surface current density distribution J given
by
$$$\vec{J}_{\phi}(z)=I/\pi/R\sum_{n=1}^N(d/R)^n
cos(n z) \vec{e}_{\phi}$$$
d=-21cm (controlling parameter proportional to replica position);
R=20cm
the radius of the spherical device;
I=1A
the current of a hypothetical centered wire;
z vertical coordinate (normalized);
$$$\vec{e}_{\phi}$$$ the azimuthal unit vector;
N=180
(the higher the more the field resembles a single
current)
A
simulation with a static background field of 3T showed no change of
simulation results and all expected effects were adding up.Discussion
Negative
µ-materials represent new
pathways of focusing magnetic fields in the distance and are of
extraordinary interest for medical purposes. Only a handful
experimental validations exist, and some showed that magnetic
fields can be manipulated inside inaccessibly
volumes. The proposed variant of the magnetic replicator
ensures spherical symmetry and allows for positioning of the patient
head inside the replicator contrary to the original case where the
replicator would need to be positioned tangential to the head
and orthogonal to the scanner’s z-axis. The parameters I, d and N drastically influence the simulation outcome
and various shapes of fields can be created and the ratio (d/R) and N restrict the current maximum. The
field divergencies unfortunately limit practical purposes of
magnetic replicators and would need to be taken into account. Unexplored is still the dependence of the $$$B_z$$$ field
on spatially linear and time-varying gradients.Conclusion
Magnetic
replicators for B0 shimming as well as for transcranial magnetic
stimulation may overcome traditional methods especially
concerning the locality of field manipulation, real-time field
manipulation by current tuning and patient
safety due to distant surface currents.Acknowledgements
We want
to thank the Austrian Science Fund (FWF): Project Number TAI-676
(1000 Ideas Program)References
1.) Milton Graeme W,
Nicorovici Nicolae-Alexandru P,
McPhedran Ross C and
Podolskiy Viktor A
2005A
proof of superlensing in the quasistatic regime, and limitations of
superlenses in this regime due to anomalous localized resonance Proc. R.
Soc. A.4613999–4034http://doi.org/10.1098/rspa.2005.1570
2.) Navau C, Prat-Camps J, Sanchez A. Magnetic energy harvesting and
concentration at a distance by transformation optics. Phys Rev Lett.
2012;109(26):263903. doi:10.1103/PhysRevLett.109.263903
3.) Navau, C., Mach-Batlle, R., Parra, A. et al. Enhancing the sensitivity of magnetic sensors by 3D metamaterial shells.
Sci Rep 7, 44762 (2017). https://doi.org/10.1038/srep44762
4.) Mach-Batlle R, Bason MG, Del-Valle N, Prat-Camps J. Tailoring Magnetic Fields in Inaccessible Regions. Phys Rev Lett. 2020;125(17):177204. doi:10.1103/PhysRevLett.125.177204
5.) Mach-Batlle, Rosa, Albert Parra, Jordi Prat-Camps, Sergi Laut, Carles
Navau, and Alvaro Sanchez. "Negative permeability in magnetostatics and
its experimental demonstration." Physical Review B 96, no. 9 (2017): 094422.
6.) A. J. Ward & J. B. Pendry
(1996)
Refraction and geometry in Maxwell's equations,
Journal of Modern Optics,
43:4,
773-793,
DOI: 10.1080/09500349608232782
7.) Pendry JB, Schurig D, Smith DR. Controlling electromagnetic fields. Science. 2006;312(5781):1780-1782. doi:10.1126/science.1125907