Stephen E Russek1, Kathryn E Keenan1, Karl F Stupic1, Cassandra M Stoffer1, Teryn S Wilkes2, and Lena Sherbakov2
1NIST, Boulder, CO, United States, 2Intermountain Neuroimaging Consortium, University of Colorado, Boulder, CO, United States
Synopsis
Keywords: System Imperfections: Measurement & Correction, Quantitative Imaging
We demonstrate the utility
of periodic use of standard phantoms to assess changes in MRI scanner
performance after unexpected major repairs. We present characterization of a 3T
scanner before and after an unexpected gradient failure and replacement using a
traceable system phantom and an isotropic diffusion phantom from the NIST/NIBIB
Medical Phantom Lending Library. No major changes were observed in gradient
calibration, geometric distortion, imaging uniformity, and resolution. Post-repair
improvements were noted in SNR, relaxation time, water diffusivity measurements,
although both pre and post-repair data were acceptable.
Introduction
MRI scanners are relied on to produce consistent
longitudinal data for clinical and research applications. Unexpected scanner
failures and changes can jeopardize the integrity of research and clinical
studies. Methods to document changes in scanner performance, either improvement
or degradation, with open and independent validation are important for the continued
reliability of image-based data. The NIST/NIBIB lending library1 provides a convenient source of phantoms that
can be used to periodically assess and document scanner performance. The
phantoms address a wide variety of measurements and range from objects that
have traceable properties to those that are just a common reference object.
Here, we look at changes due to an unexpected
gradient failure, (Fig.1a,b), which required the magnet to be ramped down, new
gradients and RF coils installed, and the scanner re-shimmed with new calibrations
such as eddy-current-corrections. The site had routinely imaged the ISMRM
system phantom2, the NIH/RSNA diffusion phantom3, as well as an fBIRN phantom4 required for ongoing studies. The analysis was
similar to that done at the same site when a major upgrade was scheduled5, though in the present case the study was ad-hoc
due to the unanticipated failure and used a more extensive analysis of the system phantom. The site has several on-going
long-term studies including a component of the 10-year Adolescent Brain
Cognitive Development (ABCD) Study, the largest study of brain development and
child health in the US, now in year 5. Methods
The system phantom was scanned using the body
coil (Fig.1c). The 56-element fiducial array was used to assess image uniformity
and geometric distortion, while the 14-element NiCl2 doped-water
array was used for relaxation time measurements. The diffusion phantom,
which consists of 13 vials with 6 different polyvinyl pyrrolidone (PVP)
concentrations (Fig.1d), was imaged using both 20-channel and 32-channel head
coils to measure the apparent diffusion coefficient (ADC) of water in the
polymer solutions. These measurands were chosen since they are the most likely
to highlight significant changes in the gradients, body coil, and associated
calibrations. Both phantoms included MRI-readable thermometers, which were
essential since phantom temperatures varied by 3°C. An fBIRN phantom, a simple
spherical gel structure, was imaged with a 32-channel head coil and the data
sent to a common processing site for the ABDC study6, with a “pass/fail” grade returned.
The imaging and analysis protocols2,3 were
the same for pre and post-repair. The
image uniformity and geometric distortion data are taken from isotropic 3d gradient
echo scan with 1mm voxels reconstructed with and without gradient nonlinearity
corrections. The T1 and T2 scans were obtained from inversion-recovery and
multi-spin-echo sequences, respectively. The diffusion scans used 4 b-value
(0,500,1000,2000s/mm2) trace-weighted 2d EPI scans. The total scan time was 2h
and image analysis time was approximately 4h using the lending library’s opensource
Python code or Windows executable7. The total scan and data processing time were
insignificant compared to the repair time and added little burden to the repair
process.Results and Discussion
The image uniformity data, taken from the
fiducial array, pre and post-repair are shown in Fig.2a,b. There is a uniform
loss of signal going from the center to the periphery with very good
agreement pre and post-repair throughout the phantom.
Fig2.c-f shows left/right geometric distortion
data, with and without corrections for nonlinear gradients. The geometric
distortion with gradient corrections, before and after repair are very similar
and small, with a slight improvement after the repairs. The accuracy of the fiducial
sphere positions is specified at < 0.1mm. The overall gradient calibration
factor was 0.997 pre-repair, and 0.998 post-repair.
Fig.3 shows relaxation time measurements,
plotted as a deviation from the NMR calibrated values with temperature corrections applied. For the central spheres, the standard deviation of T1 from the
calibration values were 3.2% pre-repair, 2.0% post-repair. The corresponding
standard deviation of T2 from the calibration values was 3.9% pre-repair, and 3.1%
post-repair. Fig.4 shows the measured ADC values and temperature-corrected
deviation from the calibration values from the 13 diffusion phantom vials. ADC deviation
from the calibrated values was 4.9% pre-repair and 2.7% for the 32-channel
head, while for the highest diffusivity samples, which have less uncertainty,
the ADC deviation was 2.1% pre-repair and 0.7% post-repair for the 32-channel
coil.
The data from the fBIRN phantom were analyzed by
the ABCD quality assurance protocol and metrics were similar post and
pre-repair, both passing . A summary table of all results is shown in Fig.5.Conclusions
The phantom data showed little change in
scanner performance after gradient failure and replacement. The data assisted
in the decision to reinitiate study scans by providing additional verification
that scanner performance had not significantly changed. The use of the phantoms
and data processing is straight forward and not time consuming. As such,
routine phantom imaging can provide an indication of scanner health that is
critical in the event of scanner failures and repairs. There is a great deal of
work remaining to understand and improve measurement repeatability to allow for
better scanner monitoring. It is possible that the gradient failure may have
been presaged by small changes in gradient performance that could have been
detected before failure with an appropriate phantom and protocol. Acknowledgements
We acknowledge funding support from NIBIB R21-686-0010.References
1. NIST.
NIST-NIBIB Medical Phantom Lending Library. Available from: https://www.nist.gov/programs-projects/nistnibib-medical-imaging-phantom-lending-library.
2. Stupic
KF, Ainslie M, Boss MA, Charles C, Dienstfrey AM, Evelhoch JL, Finn P, Gimbutas
Z, Gunter JL, Hill DLG, Jack CR, Jackson EF, Karaulanov T, Keenan KE, Liu G,
Martin MN, Prasad PV, Rentz NS, Yuan C, Russek SE. A standard system phantom
for magnetic resonance imaging. Magnetic Resonance in Medicine. 2021;n/a(n/a). doi:
https://doi.org/10.1002/mrm.28779.
3. CaliberMRI.
Available from: https://qmri.com/wp-content/uploads/2021/06/Diffusion-Phantom-Manual_Rev-G_web.pdf.
4. Friedman
L, Glover GH. Report on a multicenter fMRI quality assurance protocol. Journal
of magnetic resonance imaging : JMRI. 2006;23(6):827-39. Epub 2006/05/02. doi:
10.1002/jmri.20583. PubMed PMID: 16649196.
5. Keenan
KE, Gimbutas Z, Dienstfrey A, Stupic KF. Assessing effects of scanner upgrades
for clinical studies. Journal of Magnetic Resonance Imaging. 2019;0(0). doi:
10.1002/jmri.26785.
6. ABCD.
Available from: https://github.com/ABCD-STUDY/FIONA-QC-PHANTOM.
7. NIST.
Available from: https://github.com/MRIStandards/PhantomViewer.