Nastaren Abad1, Eric Fiveland1, Seung-Kyun Lee1, Yihe Hua1, Shengzhen Tao2, Joshua Trazasko3, Matt A. Bernstein3, and Thomas K.F Foo1
1GE Research, Niskayuna, NY, United States, 2Mayo Clinic, Jacksonville, FL, United States, 3Mayo Clinic, Rochester, MN, United States
Synopsis
Keywords: System Imperfections: Measurement & Correction, System Imperfections: Measurement & Correction, Gradient Non-Linearity; Spatial Fidelity; High Performance Gradients
Spatial encoding in MR-systems is subject to gradient non-linearity.
If ignored or inadequately calibrated during system install, non-linear
gradient fields manifest as spatial distortions impacting image quality, diminish
accuracy for applications requiring MR guided intervention and introduce systematic
errors in quantitative imaging such as diffusion MRI. High-performance, high-efficiency
gradient systems, such as MAGNUS, require accurate calibration for imaging
precision. In this study, a fiducial phantom was utilized to characterize and
correct for residual distortions after standard gradient calibrations. Results highlight
that distortion due to gradient non-linearity can be successfully reduced by phantom-based
calibration for improved accuracy inline with QC metrics.
Introduction
Spatial
encoding fields in MR systems are inherently non-linear, with certain
design/engineering trade-offs requiring the constraints for gradient non-linearity (GNL) to be
intentionally relaxed in-lieu of higher gradient amplitudes and slew rates,
reduced eddy currents and/or higher PNS thresholds1,2. Typically,
in whole-body systems with larger design field-of-view (FOV), GNL artifacts appear subtle, and distortions manifest largely near
the edge of the FOV. Here, odd-order spherical harmonic terms up to the 5th order are sufficient to
parametrize gradient fields. However, systems with high-performance
asymmetric, transverse gradients such as the compact 3T3
and specialized head-only gradients4–7 exhibit
complex non-linear gradient fields and require higher-order, and both odd-
and even-order terms to adequately model the gradient field8. Furthermore, inserts such as the MAGNUS4
have high gradient amplitude efficiency (~0.32 mT/m/A) and require high precision
in calibration of the gradient driver settings. In such systems, standard systems
calibrations are insufficient as they do not have sufficient degrees of freedom
to capture spatially-varying non-linear fields.
Notably,
with higher gradient amplitudes, non-linear spatially-varying fields introduce
systematic inaccuracies in contrasts such as in diffusion MRI, where $$$b\ \propto\ G^2$$$, resulting in spatially varying
b-values, that can be corrected if gradient non-linearity is well-characterized9. Spatial
distortions from GNL are of concern for various applications such as MR-guided interventions, radiation therapy and
neurosurgical planning, as well as detection
of subtle changes in brain morphology/volumetry associated with disease
progression in longitudinal studies. In clinical representations such as
Alzheimer’s disease, small volume changes are especially susceptible as the patient
position may vary in longitudinal studies10–12.
Spatial
fidelity is a key performance criterion for MR systems, and previous studies
have improved relative distortions by either characterizing field decomposition or by using design-based gradient field
maps for pixel-wise distortion correction. In this study, a dedicated fiducial
phantom was utilized to explicitly characterize
and correct for absolute residual GNL by re-calibrating the
gradient coil gain. Systematically offsetting the phantom allowed for
displacement in marker position, a standard metric for quantitative image
accuracy, to be determined over 26-cm design FOV. Methods
Fiducial
Phantom: Distortions from GNL were characterized using the EMR128(Phantom Lab,Greenwich,NY)11, an 18-cm sphere containing 221fiducials at fixed positions.
MR
Acquisition: All acquisitions utilized a 3.0 T GE MRI (GE HealthCare, Waukesha, WI, USA), equipped with a MAGNUS insert (GE Research, Niskayuna, NY), using a single transmit-receive birdcage
coil. The phantom was scanned with and without Gradwarp, a
standard, vendor-supplied GNL correction, using a fast spoiled gradient-echo (3D FSPGR) sequence. The acquisition
parameters were: receiver bandwidth=±62.5 kHz, readout direction=S/I, matrix=256x256,
1.0-mm isotropic resolution, TR/TE=4.1/1.5 ms,
flip angle=11˚, 252 slices. To cover the entire 26-cm diameter
spherical volume (DSV) for the gradient coil, the 18-cm phantom was
systematically re-positioned using a custom designed rig which allowed for ±5
cm directional offsets (from isocenter) in the RL/AP/SI directions. An American
College of Radiology (ACR) phantom was additionally scanned using the ACR test
protocol with default and fiducial-based gradient calibrations.
Image
Analysis: Associated
software tool (SMARI, Phantom Lab) was utilized to track fiducial positions by establishing a local coordinate space, and template
based correlation. MATLAB was used to generate displacement maps with the spatial distortion summarized as a displacement of
measured versus known/actual fixed marker positions. Optimal gradient
calibration factors can be determined by a simple linear regression to minimize
residuals. Results & Discussion
Phantom fiducial displacement with
and without GNL correction is shown in Figure 2. It was noted that using default systems calibrations alone was not sufficient to
account for the complex non-linear fields, with residual distortion fields exhibiting linear spatial
dependence on all three axes.
Calibration
of the gradients, by minimization of residuals, captured this first-order
spatial distortion, highlighting that the linear scaling coefficients can be
accurately estimated (Figure 2c).
Table1. shows the mean and maximum
distortions and RMSE at a 10-cm and 20-cm DSV, with the systems default calibration (residual RMSE ≤0.41 mm), and fiducial-based calibration (residual RMSE ≤0.3 mm). The residuals after additional
calibration demonstrated spatial accuracy in line with conventional whole-body
gradients11 (~0.08
mT/m/A) and with the asymmetric compact 3T gradient coil8 (0.13
mT/m/A), which has lower amplitude efficiency than MAGNUS due to the proximity
of its primary and shield layers. The max-distortion values satisfy QC criteria established in the ADNI study.
Data
from the ACR phantom pre- and post-calibration highlighted marginal
improvements in sharpening of contrast at the edges and at grid junctions. Preliminary work with the fiducial phantom and
the MAGNUS gradient coil focused on refining linear scaling gradient coefficients.
For additional accuracy, simulated spatial distortion coefficients can also be
evaluated on a per-system basis using an iterative approach as previously
proposed by Tao, et al.8, by
inclusion of higher azimuthal orders, or by utilizing a pixel-by-pixel fit over
the distortion field13,14Conclusion
In this work, we demonstrated that geometrical distortions due to GNL can be improved by more precise gradient calibration with a fiducial-phantom in line with the
published QC criteria. While this was established on an asymmetric head-only
gradient with more complex non-linear fields, a similar method can be utilized
for whole-body configurations that have similar high-efficiency. Additionally fiducial-based characterization can potentially
identify system specific errors outside of manufacturing and assembly
tolerances.Acknowledgements
Grant funding from NIH
U01EB028976, NIH U01EB024450, CDMRP W81XWH-16-2-0054. References
1. Alsop, D. C. & Connick, T. J. Optimization of torque-balanced asymmetric head gradient coils. Magnetic Resonance in Medicine 35, 875–886 (1996).
2. Doty, F. D. MRI Gradient Coil Optimization. in Spatially Resolved Magnetic Resonance 647–674 (John Wiley & Sons, Ltd, 1998). doi:10.1002/9783527611843.ch60.
3. Foo, T. K. F. et al. Lightweight, compact, and high-performance 3T MR system for imaging the brain and extremities. Magnetic Resonance in Medicine 80, 2232–2245 (2018).
4. Foo, T. K. F. et al. Highly efficient head-only magnetic field insert gradient coil for achieving simultaneous high gradient amplitude and slew rate at 3.0T (MAGNUS) for brain microstructure imaging. Magnetic Resonance in Medicine 83, 2356–2369 (2020).
5. Myers, C. & Roemer, P. B. Highly linear asymmetric transverse gradient coil design for head imaging. in Proceedings of 10th Annual Meeting SMRM (1991).
6. Huang, S. Y. et al. Connectome 2.0: Developing the next-generation ultra-high gradient strength human MRI scanner for bridging studies of the micro-, meso- and macro-connectome. Neuroimage 243, 118530 (2021).
7. Feinberg, D. A. et al. Design and Development of a Next-Generation 7T human brain scanner with high-performance gradient coil and dense RF arrays. Proceedings of the 29th Annual Meeting of the ISMRM, Virtual Meeting (2021) (2021).
8. Tao, S. et al. Gradient nonlinearity calibration and correction for a compact, asymmetric magnetic resonance imaging gradient system. Phys Med Biol 62, N18–N31 (2017).
9. Tao, A. T. et al. Improving apparent diffusion coefficient accuracy on a compact 3T MRI scanner using gradient nonlinearity correction. J Magn Reson Imaging 48, 1498–1507 (2018).
10. Ching, C. R. K. et al. Phantom-based MRI corrections and power to track brain change. in 2012 9th IEEE International Symposium on Biomedical Imaging (ISBI) 1172–1175 (2012). doi:10.1109/ISBI.2012.6235769.
11. Gunter, J. L. et al. Measurement of MRI scanner performance with the ADNI phantom: Measurement ADNI phantom. Med. Phys. 36, 2193–2205 (2009).
12. Jack, C. R. et al. The Alzheimer’s Disease Neuroimaging Initiative (ADNI): MRI Methods. J Magn Reson Imaging 27, 685–691 (2008).
13. Weavers, P. T. et al. Image-based gradient non-linearity characterization to determine higher-order spherical harmonic coefficients for improved spatial position accuracy in magnetic resonance imaging. Magn Reson Imaging 38, 54–62 (2017).
14. Tao, S., Trzasko, J. D., Shu, Y., Huston, J. & Bernstein, M. A. Integrated Image Reconstruction and Gradient Nonlinearity Correction. Magn Reson Med 74, 1019–1031 (2015).