Qing Lin1, Weikun Chen1, Jian Wu1, Taishan Kang2, Xinran Chen1, Zhigang Wu3, Shuhui Cai1, and Congbo Cai1
1Department of Electronic Science, Xiamen University, Xiamen, China, 2Magnetic Resonance Center, Zhongshan Hospital Afflicated to Xiamen University, Xiamen, China, 3MSC Clinical & Technical Solutions, Philips Healthcare, Shenzhen, China
Synopsis
Keywords: Machine Learning/Artificial Intelligence, Quantitative Imaging, water/fat separation
Most of the
water/fat separation techniques need to acquire multiple images with different echo
time, which usually take long acquisition time. Multiple overlapping-echo
detachment (MOLED) imaging can shorten acquisition time by acquiring multiple
MR echo signals in the same k-space. Here, a new method for fast water/fat
separation T
2* mapping with MOLED technology was proposed, which can
obtain quantitative T
2* maps and M
0 maps of water and fat in a single
shot. In vivo experiment demonstrates that the proposed method can obtain
accurate quantitative parameter
values and accurate separation of water and fat images.
Introduction
Water/fat
separation is important for MRI applications. Additionally, T2* relaxation
time of the water and fat signals can provide valuable information for
biomedical research and clinical diagnosis of diseases.1 However,
most of the water/fat separation techniques need to acquire multiple images
with different echo time, which usually take long acquisition time, and may be
susceptible to physiological motions, such as Dixon-based methods.2,3
Multiple overlapping-echo
detachment (MOLED)4 imaging provides an innovative idea for
accelerating MR parametric mapping by acquiring multiple MR echo signals with
different signal evolutions in the same k-space. Here, we proposed a new method
for fast water/fat separation T2* mapping with MOLED technology.
With MOLED technology, water and fat signals with different T2*
weightings can be obtained in a single shot, which allows us to obtain quantitative
T2* maps and M0 maps of water and fat without additional scans. Experimental
results show that the proposed method can obtain accurate quantitative parameter values
and accurate separation of water and fat images.Methods
Pulse sequence
design:
The
MOLED sequence is shown in Figure 1. Signals
excited by four sequential RF pulses are moved to different locations in k-space
by shifting gradients (GROi,GPEi,i=1,2,3,4). Echoes with different evolution
time are acquired and overlapped in the k-space. GRE-EPI readout is used in two
echo trains. Consequently, these echoes contain different information of water
and fat, and convey T2* information with different weights, hence
allowing water/fat separation and T2* quantification. There are eight echoes in two echo trains
totally, providing enough information for multi-peak water/fat separation. The
TEs of these echoes range from 5.45 ms to 43 ms, while the T2* value of water and fat in the leg region is
generally distributed within 20-35ms. The appropriate TE range ensures the
accurate T2* quantification.
Signal model:
The nth echo Sn with echo time TEn can be expressed as:
$$S_{n}=(W\times{e^{\frac{-TE_{n}}{T_{2,w}^*}}}+\sum_{m=1}^6\alpha_{m}e^{i2\pi\delta{f_{m}}TE_{n}}\times{F}\times{e^{\frac{-TE_{n}}{T_{2,f}^*}}})e^{i2\pi\phi{TE_{n}}}$$
where W and F represent the complex water and fat signals. A six-peak
fat model is used in this work, and the relative amplitude αm of each peak as well as their frequency shift δfm are taken as a prior.5 φ(Hz) is off-resonance frequency induced by main magnetic field
inhomogeneity. $$$T_{2,w}^*$$$is the T2* of water and $$$T_{2,f}^*$$$ is the T2* of fat. Here, we assume all fat peaks have equal T2*.
Water/Fat
separation using neural
network:
The training data were generated
using synthetic data and MRiLab software, as shown in Figure 2a. The MOLED sequence was used to stimulate
water and six fat peak components to get their respective MR signals. The image
of each component was obtained and added together to get the final MOLED image.
U-Net was employed, as shown in Figure 2b. The inputs of
U-Net are the real and imaginary parts of original MOLED image as well as shifted
MOLED image to correct chemical shift artifact along the PE dimension, and the
outputs are T2* maps and
M0 maps of water and fat, respectively.
Experiments:
This
study was approved by the IRB at Zhongshan Hospital of Xiamen University. MOLED sequence was performed in the leg region on a
3T scanner (Ingenia CX, Philips Healthcare), using a 32-channel abdomen receive coil. Imaging parameters were as follows: flip angle of
excitation pulse = 30°, TR = 6000 ms, ESP = 0.664
ms, FOV = 22×22 cm2, slice thickness = 5 mm, imaging matrix
size= 128×124. The traditional mDixon-Quant
technique and GRE sequence were used to obtain reference
images. Single slice scan time of MOLED, mDixon-Quant and GRE were 65, 448 and 5514 ms, respectively.Results
Figure 3 shows the water/fat separation results of two different slices
(a and b). The second row shows the M0 mapping results. The water/fat separation results of MOLED
are accurate and show clear texture, matching well with mDixon. Besides, the chemical shift artifacts along the PE dimension are
corrected. The third row displays the T2* mapping results. Especially, the third image in the third row is the T2* map of MOLED obtained by adding the T2* maps of water and fat. The fourth row shows
the reference T2* map
obtained by GRE sequence and the T2* error maps. In slice #1
(Figure 3a), the average errors of the maps of MOLED and mDixon are 6.61%, 14.41%, respectively,
and in slice #2 (Figure 3b), the corresponding values are 6.01%, 12.30%, respectively.
This indicates that MOLED can obtain more accurate T2* maps of water and
fat than mDixon.Discussion and conclusion
In this work, we proposed a new
method for fast water/fat separation T2* mapping based on MOLED
technology,
which allows us to obtain T2* maps and
M0 maps of water and fat in a single shot. Multi-peak fat
model, synthetic data, and U-Net were used in image reconstruction. The
proposed method can obtain accurate quantitative parameter values and accurate separation of water
and fat images.Acknowledgements
This work was supported by the
National Natural Science Foundation of China under grant numbers 82071913 and 11775184.References
1.
Ster C L, Gambarota G, et al. Breath-hold MR measurements of fat fraction, T1,
and T2* of water and fat in vertebral bone marrow.
Journal of magnetic resonance imaging, 2016, 44(3):549-555.
2.
Dixon WT. Simple proton spectroscopic imaging. Radiology, 1984, 153:189-194.
3.
Fuller S, Reeder S, et al. Iterative decomposition of water and fat with echo
asymmetry and least-squares estimation (IDEAL) fast spin-echo imaging of the
ankle: initial clinical experience. Ajr Am J Roentgenol, 2006,
187(6):1442-1447.
4.
Ma LC, Wu J, et al. Single-shot multi-parametric mapping based on multiple
overlapping-echo detachment (MOLED) imaging. Neuroimage, 2022, 263:119645.
5.
Yu H, Shimakawa A, et al. Multiecho water-fat separation and simultaneous R2*
estimation with multifrequency fat spectrum modeling. Magnetic resonance in medicine,
2008, 60(5):1122-1134.