Matthias Anders1, Carsten Warmuth1, Josef Pfeuffer2, Heiko Tzschätzsch1, Helge Herthum1, Katja Degenhardt3, Oliver Wieben4, Sebastian Schmitter3, Jeanette Schulz-Menger1,5,6, Jürgen Braun1, and Ingolf Sack1
1Charité - Universitätsmedizin Berlin, Berlin, Germany, 2MR Application Development, Siemens Healthcare, Erlangen, Germany, 3Physikalisch-Technische Bundesanstalt(PTB), Braunschweig and Berlin, Berlin, Germany, 4University of Wisconsin, Madison, WI, United States, 5Experimental and Clinical Research Center (ECRC), DZHK partner site Berlin, Berlin, Germany, 6HELIOS Klinikum Berlin Buch, Berlin, Germany
Synopsis
Keywords: Elastography, Data Acquisition
MR elastography (MRE) can noninvasively detect liver
fibrosis based on elevated stiffness values. However, a complete MRE scan is
time-consuming and typically extends over multiple breath-holds. Therefore, a
multi-shot gradient echo sequence with spiral readout was developed to provide
full two-dimensional elastograms of the liver in less than one second. The new
sequence can be used as an MRE navigator to provide immediate feedback on wave
penetration and data consistency for parameter optimization prior to running
full multi-dimensional MRE. Moreover, the method can be used to rapidly track potential
changes in liver stiffness such as those induced by respiration.
Introduction
Liver fibrosis is a prevalent disease that can be monitored noninvasively with MR elastography (MRE) based on shear wave excitation and encoding. Compared with ultrasound elastography, MRE provides better 3D coverage of the liver with higher fidelity of anatomic detail, but suffers from longer scan time. Therefore, MRE examinations of the liver are usually spit into several breath-holds.[1] Alternatively, multifrequency MRE was proposed, which was applied under free breathing to avoid measurement times exceeding five minutes[2]. For an efficient MRE setup with optimal control over shear wave amplitudes and driver settings, it would be beneficial to perform short monofrequency MRE navigator scans before more time-consuming examinations with multifrequency or breath-hold MRE protocols. The aim of this work was to develop an MRE navigator that could either provide immediate feedback on wavefield amplitudes in-vivo or be used for time-resolved mapping of liver stiffness under altered physiological conditions. Methods
Ten healthy volunteers (40±13years) were investigated. Four compressed air drivers were attached
onto the chest with a Velcro belt (two anterior, two posterior) and fed with a
pulsed air flow of 40-Hz frequency. Experiments were performed in a 3-Tesla MRI
scanner (MAGNETOM Lumina, Siemens, Germany) using a 12-channel receiver coil. Figure 1 shows
the custom MRE navigator sequence based on a multi-shot, gradient-echo spiral
sequence (FOV 360×360mm², 2.5×2.5×5.0mm³ voxel size, TR=50ms, TE=14ms, 6/3
Spiral Interleaves designed/played-out, PAT-factor 2). One field component
along the readout direction and 4 instances over a vibration cycle were
acquired. Total acquisition time was 600ms. Overall, two experiments were
performed in each subject. First, the capability of the sequence as MRE
navigator was tested in a transversal slice. The obtained MRE maps were compared
with 2D spin-echo EPI MRE of 11 slices, three encoding directions and 8 time
steps over a vibration cycle which took approximately 25 seconds. In a second
experiment, series of MRE maps were acquired in a coronal slice orientation
continuously for one minute to investigate whether liver stiffness changes with
respiration. Therefore, the breathing state was altered from inspiration to
expiration after every ten seconds following the oral instructions of the operator.
Stiffness maps in terms of shear wave speed (SWS, in m/s) were reconstructed
using the k-MDEV method[3].
For the second experiment, image registration was performed based on spatial
normalization[4] to align the liver position within the 2D-image
slice to the initial region of interest over the entire measurement time. Results
Figure 2 shows wave images and total wave
amplitudes in a volunteer obtained by navigator MRE. Wave amplitudes were in the
order of 60µm and showed shear waves within the entire field of view with indicating
sufficient wave penetration of the abdomen. The attenuation of shear waves
toward the center of the liver provides information about the viscosity of the
tissue.
Figure 3 shows transversal SWS maps in a
representative subject obtained from 600ms-navigator MRE and 25s-EPI MRE in two
respiration states. SWS values within the liver are slightly higher for the MRE
navigator (1.40±0.46m/s) than
for EPI-MRE (1.26±0.31m/s, p=0.042).
There was no significant difference between liver stiffness measured by EPI-MRE
in inspiration (1.15±0.27m/s) and
expiration (1.26±0.31m/s, p=0.517).
Figure 4 shows the variation of liver
stiffness during alternating breathing states in one volunteer. The respiratory
curve was generated from the displacement along the superior-inferior axis as
obtained from the motion-correction.
The group statistics in figure 5 showed no
significant alteration of SWS with breathing, although some variation in SWS in
synchrony with respiratory motion was observed in individual subjects.Discussion
This work presented a preliminary study using a novel MRE
navigator based on a multi-shot, gradient-echo sequence with spiral readout. With
an acquisition time of 600ms for full SWS maps of the abdomen, this sequence
offers, to our knowledge, the fastest MRE method for liver studies to date. As
shown in two single experiments, the method can be used to optimize the driver
based on in-vivo wave amplitudes prior to lengthy higher-dimensional MRE
acquisitions or to rapidly track potential changes in liver stiffness due to
physiological changes such as geometry, blood flow, or abdominal pressures. Previous
work has shown that liver stiffness decreases with increased abdominal pressure
induced by the Valsalva maneuver, which is associated with a decrease in hepatic
blood flow[5,6]. Similar effects, albeit to a lesser extent, may
occur through respiration, which was indeed observed in some of our subjects. Therefore,
expiration, when hepatic blood flow through the hepatic vein is at its maximum
capacity, is the preferred breathing state for MRE studies. As a limitation, we
observed a significant difference between navigator MRE and EPI-MRE, which
deserves further investigations. We assume that reduction of field encoding to
only one component (read-out) limits the stability of MRE against diffraction effects
resulting in slightly higher values. In ongoing studies, we are testing the encoding
of the thru-plane component as it is used in standard 2D-MRE of the liver[1].Conclusion
Navigator MRE for abdominal stiffness measurements in
less than one second is feasible. This viable method can be used for rapid
testing and parameter optimization of MRE driver hardware and post-processing
pipelines, as well as for measuring potential stiffness variations due to respiratory-induced
deformations of the liver. Acknowledgements
The authors gratefully acknowledge funding
from the German Research Foundation (GRK2260, BIOQIC).References
[1] Venkatesh, S. K., Yin, M., & Ehman, R. L. (2013). Magnetic resonance
elastography of liver: technique, analysis, and clinical applications. Journal
of Magnetic Resonance Imaging, 37(3), 544-555.
[2] Shahryari, M., Meyer, T., Warmuth, C.,
Herthum, H., Bertalan, G., Tzschätzsch, H., ... & Sack, I. (2021).
Reduction of breathing artifacts in multifrequency magnetic resonance
elastography of the abdomen. Magnetic Resonance in Medicine, 85(4),
1962-1973.
[3] Tzschatzsch,
H., Guo, J., Dittmann, F., Hirsch, S., Barnhill, E., Johrens, K., Braun, J.
& Sack, I. (2016). Tomoelastography by multifrequency wave number recovery
from time-harmonic propagating shear waves. Med Image Anal, 30, 1-10.
[4] Ashburner, J. (2009). Computational
anatomy with the SPM software. Magnetic Resonance Imaging, 27(8),
1163-1174.
[5] Ipek‐Ugay, S., Tzschätzsch, H., Braun, J., Fischer, T., & Sack, I. (2017).
Physiologic reduction of hepatic venous blood flow by the Valsalva maneuver
decreases liver stiffness. Journal of Ultrasound in Medicine, 36(7),
1305-1311
[6] Meyer, T., Tzschätzsch, H., Wellge, B., Sack, I., Kröncke, T., &
Martl, A. (2022). Valsalva Maneuver Decreases Liver and Spleen Stiffness
Measured by Time-Harmonic Ultrasound Elastography. Frontiers in Bioengineering and Biotechnology, 10.