Anne-Sophie van Schelt1,2, Nienke Petronella Maria Wassenaar1, Jurgen H. Runge1, Jaap Stoker1,2,3, Aart J Nederveen1, and Eric M Schrauben1
1Radiology and Nuclear Medicine, Amsterdam UMC, location AMC, Amsterdam, Netherlands, 2Imaging and Biomarkers, Cancer Center Amsterdam, Amsterdam, Netherlands, 3Endocrinology, Metabolism, Amsterdam Gastroenterology, Amsterdam, Netherlands
Synopsis
Keywords: Elastography, Pancreas
Breathing
motion can be detrimental to quantitative accuracy in pancreatic MR
elastography (MRE). We investigated motion-correction strategies on sagittal
multi-frequency free-breathing SE-EPI MRE acquisitions to mitigate
breathing-motion and increase accuracy. Breathing-states were determined
through a respiratory-belt and multiple number of bins were tested. Three intensity-based registration methods
(with and without non-rigid post-processing) and non-rigid registration were ranked
on MRE quality measures. The best method showed improved data quality,
inversion precision and repeatability compared to no correction, but resulted
in apparently increased shear wave speed (SWS), which warrants further
investigation.
Introduction
Pancreatic MR elastography
(MRE) is a non-invasive MR technique that is able to quantify visco-elastic
properties of tissue, e.g. assessment of pancreatitis, benign or malignant
tumors.1-3 Breathing motion can be detrimental to quantitative accuracy in
MRE, particularly in the pancreas. Breathing mitigation techniques, such as repeated
breath-holding, are used to overcome this. However, this can be particularly
uncomfortable for patients, prolongs protocol time and limits data richness.4 Multi-frequency MRE allows for time-efficient and highly-resolved stiffness
maps in free-breathing.5 However, the pancreas moves substantially with
respiration, potentially hampering accurate analysis due to misalignment over
the additional dimensions of the MRE acquisition (motion-encoding-gradient
direction and wave-offsets).
Pancreatic MRE
is typically performed in an axial orientation to match anatomical scans. Approaches
have been explored to correct respiratory motion in single-shot echo-planar-imaging
(SE-EPI)4, but they have not considered both foot-head and anterior-posterior
motion, both of which are present and substantial in the pancreas. A sagittal acquisition
is more conducive to resolving these motions and therefore, we investigated
motion-correction strategies on sagittal multi-frequency free-breathing SE-EPI
MRE acquisitions. Methods
All scanning
was performed at 3T (Ingenia,Philips,Best,Netherlands) using a multi-slice,
multi-frequency SE-EPI MRE sequence at four mechanical frequencies
(30,40,50,60Hz) introduced using four pneumatic drivers (figure 1A).5 Acquisition
parameters can be found in figure 1B. Six healthy volunteers (3♀, age=29±3years)
underwent one axial and two consecutive sagittal MRE acquisitions for
repeatability analysis. A respiratory belt (PEAR-belt, Philips) was placed on the
lower abdomen to record the respiratory signal throughout each scan. MRE data
was cropped (50% both in feet-head and anterior-posterior) before
motion-correction.
Three
intensity-based registration methods were tested (figure 1C): monomodal,
multimodal and phase-correlation, using a regular step-gradient-descent-optimizer,
one-plus-one revolutionary-optimizer and windowing in the frequency-domain,
respectively. Pre-processing normalization and similarity-registration were
applied for all three methods. Results with and without post-processing using non-rigid local registration were compared. Lastly, a stand-alone non-rigid
registration was used.
MRE data were
binned into respiratory motion-states (nBins) as determined from the respiratory
belt signal (nBins=5/8/10/12). All motion-states were registered to
end-expiration state using the aforementioned motion-correction methods applied
on the magnitude images, after which the geometric translations were imposed on the real and imaginary parts of the complex data.
Shear-wave-speed (SWS) was calculated using the kMDEV inversion algorithm.6
Regions-of-interest
(ROIs) were manually drawn over the pancreas on end-expiration magnitude-data. Quality parameters were: (1) Displacement
signal-to-noise (Displ-SNR)7, (2) Lap-SNR: the variance (σ) of the Laplacian
(Δ) of the images4, (3) octahedral-shear-strain SNR (OSS-SNR)7, (4) stability:
defined as intensity change over time of ROI boundary-voxels and (5) ratio of
SWS and standard deviation (STDEV) within the ROI. Rank-scores were given for each
motion-correction method for all quality parameters with the sum resulting in
an overall-ranking. The best and worst method were compared to
non-motion-corrected SWS for axial and sagittal scans using a repeated measures
ANOVA and pairwise comparison with Bonferroni correction (p<0.05 was deemed
significant). Repeatability was assessed using Bland-Altman analysis.
Image
registration and analysis, delineation and statistical analysis were performed
in Matlab (R2021b,Mathworks,Natick,MA,USA), ITK-snap (v3.8.0) and SPSS
(version8) respectively. Results
A total of 28
different combinations: nBins, registration methods (MoCo) and post-processing
in the case of intensity based MoCo, were ranked together with no MoCo for all
quality measures, see table 1. The best method was phase-correlation with nBins=10 and non-rigid post-processing. The worst scoring method was phase-correlation registration with nBins=12 and no motion-correction was 28th.
Representative images of the best method can be found in figure 2. The average
SNR-values per nBins, MoCo and post-processing are shown in figure 3. Bland-Altman
analysis showed 95%-limits-of-agreement of [-0.16, 0.10], [-0.06 0.09] and
[-0.16, 0.17] m/s for no MoCo, best and worst respectively, see figure 4. The
average SWS for all volunteers was 1.30±0.07, 1.37±0.10, 1.48±0.10 and 1.41±0.09m/s for axial, no MoCo, best and worst
respectively (F(5,15)=9, p<.001). Pairwise comparison showed significant
differences between best MoCo, axial and sagittal no MoCo (p<.05), see
figure 4.Discussion
Here we show that
motion-correction increases precision in MRE inversion. In particular,
phase-correlation with non-rigid post-processing in 10 respiratory bins
showed the best performance. nBins does not have a large effect on registration
quality overall, though increasing nBins beyond some threshold would eventually
degrade registration. Post-processing showed increased performance, which may
be due to the local registration. Multimodal-intensity can register images
with different contrast, making it robust for different imaging modalities,
however for MRE (single-contrast) it could introduce errors.
Contrary to
recent work, which performed motion-correction of coronal abdominal MRE at 1.5T
using a 2D rigid-body image registration method4, our work showed increased
apparent SWS after motion-correction. This observation needs further analysis,
preferably within a moving phantom with known stiffness inclusions to eliminate
causation by MoCo.
Bland-Altman
analysis showed an increased repeatability after using the best performing MoCo
method, whilst the worst performing showed decreased repeatability, with wider
limits-of-agreement comparable to no MoCo. This may have implications on future
clinical findings – particularly in smaller tissues such as the pancreas for
improved quantification of the heterogeneous tumor microenvironment. Conclusion
Motion-correction in sagittal free-breathing
SE-EPI pancreatic MRE is promising, with improved data quality, inversion precision and
repeatability compared to no correction, but the increased apparent shear wave speed within
the pancreas warrants further investigation.Acknowledgements
No acknowledgement found.References
1. Zhu L, Guo J, Jin Z, et al. Distinguishing pancreatic cancer and autoimmune
pancreatitis with in vivo tomoelastography. Eur Radiol.
2021;31(5):3366-3374. doi:10.1007/s00330-020-07420-5
2. Gültekin E, Wetz C, Braun J, et al. Added value of tomoelastography for
characterization of pancreatic neuroendocrine tumor aggressiveness based on
stiffness. Cancers (Basel). 2021;13(20).
doi:10.3390/cancers13205185
3. Shi Y, Cang L, Zhang X, et al. The use of magnetic resonance elastography in
differentiating autoimmune pancreatitis from pancreatic ductal adenocarcinoma:
A preliminary study. Eur J Radiol. 2018;108:13-20.
doi:10.1016/j.ejrad.2018.09.001
4. Shahryari
M, Meyer T, Warmuth C, et al. Reduction of breathing artifacts in
multifrequency magnetic resonance elastography of the abdomen. Magn
Reson Med. 2021;85(4):1962-1973. doi:10.1002/mrm.28558
5. Dittmann F, Tzschätzsch H, Hirsch S, et al. Tomoelastography of the abdomen:
Tissue mechanical properties of the liver, spleen, kidney, and pancreas from
single MR elastography scans at different hydration states. Magn
Reson Med. 2017;78(3):976-983. doi:10.1002/mrm.26484
6. Tzschätzsch H, Guo J, Dittmann F, et al. Tomoelastography by multifrequency
wave number recovery from time-harmonic propagating shear waves. Med
Image Anal. 2016;30:1-10. doi:10.1016/j.media.2016.01.001
7. Bertalan
G, Guo J, Tzschätzsch H, Klein C, Barnhill E, Sack I, Braun J. Fast
tomoelastography of the mouse brain by multifrequency single-shot MR
elastography. Magn Reson Med. 2019 Apr;81(4):2676-2687. doi: 10.1002/mrm.27586.
Epub 2018 Nov 4. PMID: 30393887.