Yu Ding1, Yingmin Liu1, Chong Chen1, Ning Jin2, Rizwan Ahmad1, and Orlando Simonetti1
1The Ohio State University, Columbus, OH, United States, 2Siemens Healthineers, Columbus, OH, United States
Synopsis
Keywords: Myocardium, Low-Field MRI, Late Gadolinium Enhancement
Low-field
cardiac MR imaging has intrinsically low SNR that degrades the image quality,
especially for LGE imaging. Using multiple acquisitions and motion-corrected
averaging is a well-established method to improve SNR. However, the
ineffectiveness of MOCO at low field leads to blurring. We proposed a novel
method that integrates motion correction into compressed sensing and
reconstructs a single motion-corrected image.
Quantitative sharpness measurements demonstrate that the proposed method
improves boundary sharpness, independent of differences in SNR.
Introduction
Low-field
scanners with limited gradient performance potentially offer a cost-effective
alternative to high-field MRI for cardiac imaging. However, the intrinsically
low signal-to-noise ratio (SNR) at low field can degrade image quality,
especially in late gadolinium enhancement (LGE) 1, 2. Reconstruction
methods, e.g., compressed sensing
(CS), can improve individual image SNR. The well-established motion correction (MOCO)
+ averaging method2,3 can also improve SNR when multiple
acquisitions are available. Higher acceleration is needed to overcome limited
gradient performance (26mT/m, 45 mT/m/ms); however, at low field this leads to low
SNR single-shot source images that can reduce the accuracy of motion correction.
Averaging of incorrectly registered source images introduces blurring and degrades
image quality.
In
this study, we propose a novel MOCO compressed sensing (CS) method (MOCO + CS) to
improve LGE image sharpness by incorporating MOCO into the CS image reconstruction
algorithm.Methods
Multiple single-shot LGE images were acquired 10
to 20 minutes following injection of 0.15 mmol/kg gadolinium contrast agent (Gadobutrol,
Bayer Healthcare, LLC, Whippany, NJ) in 10 healthy volunteers on a commercial
0.55T MR scanner (Free.Max, Siemens Healthcare, Erlangen, Germany). The k-space
data from these scans were used to develop and test the effectiveness of the
proposed MOCO + CS method.
24 sets of
k-space raw data from LGE image series were retrospectively included,
comprising 18 short-axis, 3 four-chamber long-axis, and 3 two-chamber image
series. A pseudo-random down-sampling pattern4 with an effective acceleration
rate = 6 was used for data acquisition.
All
k-space data sets were reconstructed using both the MOCO + averaging method
that employs a CS reconstruction of the source images prior to MOCO, and the
proposed MOCO + CS method. MOCO + CS incorporated the estimated motion
displacement field into the CS cost function, i.e.:
$$argmin_x || DFSRx - k_0 ||^2_2 + \rho_x|| TV_xx ||_1$$
where x is the motion-corrected
averaged image; k0 is the acquired k-space raw data; R is the forward MOCO
operator; S is the sensitivity
map; F is the Fourier
transform operator; D is the downsampling
operator; TVx is the spatial total variation
operator; ρx is an adjustable
parameter. For the implementation of R, the “demon” MOCO algorithm was employed5.
Boundary (edge) sharpness was evaluated in each
pixel at the boundary between the LV myocardium and blood pool using the sigmoid
function fitting method along the normal direction6. A paired t-test
was used to evaluate the significance of the boundary sharpness differences. To
account for the impact of SNR on the measured sharpness, Gaussian noise was
added to the MOCO + averaging images to generate a new image series that matched
the SNR of the MOCO + CS method.Results
Figure 1 shows a typical image reconstructed by both
methods, and an SNR matched image. The image reconstructed using the proposed MOCO
+ CS method has visibly improved sharpness. The image sharpness of MOCO + averaging
method, SNR-matched MOCO + averaging method, and the MOCO + CS method are 0.54+/-0.14 pixel-1,
0.64+/-0.18 pixel-1 and 0.84+/-0.26 pixel-1 (p-value <
0.001), respectively (Figure 2, the SNR-matched result is not shown). Conclusion
This study demonstrates the proposed MOCO + CS method
improves the sharpness of free-breathing LGE images acquired using a 0.55T
commercial scanner. The sharpness increase was shown to be independent of SNR
differences. We believe the loss of sharpness in the traditional MOCO + averaging
method is due to the challenges of applying MOCO to low-SNR source images resulting
in residual misregistration, which in turn causes blurring of the averaged
image result. The proposed method, in contrast, leverages all available raw
data to reconstruct one image frame; therefore, it is less sensitive to
inaccuracies in the motion correction displacement field. Further studies are
warranted in a patient population with myocardial scar and fibrosis.Acknowledgements
No acknowledgement found.References
1.
Simonetti, Orlando P., et al. "An improved MR imaging technique for the
visualization of myocardial infarction." Radiology 218.1 (2001): 215-223.
2.
Kellman, Peter, et al. "Motion‐corrected free‐breathing
delayed enhancement imaging of myocardial infarction." Magnetic Resonance
in Medicine: An Official Journal of the International Society for Magnetic
Resonance in Medicine 53.1 (2005): 194-200.
3.
Kellman, Peter, et al. "Dark blood late enhancement imaging." Journal
of Cardiovascular Magnetic Resonance 18.1 (2017): 1-11.
4.
Ahmad, Rizwan, et al. "Cartesian sampling for dynamic magnetic resonance
imaging (MRI)." U.S. Patent No. 11,294,009. 5 Apr. 2022.
5.
Dirk-Jan Kroon (2022). Multimodality non-rigid demon algorithm image
registration
(https://www.mathworks.com/matlabcentral/fileexchange/21451-multimodality-non-rigid-demon-algorithm-image-registration),
MATLAB Central File Exchange. Retrieved July 25, 2022.
6. Ahmad, Rizwan, Yu Ding, and
Orlando P. Simonetti. "Edge sharpness assessment by parametric modeling:
application to magnetic resonance imaging." Concepts in Magnetic Resonance
Part A 44.3 (2015): 138-149.