Yupeng Cao1, Jun Zhao1, Weinan Tang2, Wentao Liu1, and Dong Han1
1National Center for Nanoscience and Technology, Beijing, China, 2Wandong Medical Inc, Beijing, China, Beijing, China
Synopsis
Keywords: New Trajectories & Spatial Encoding Methods, Lung
Single
breath-hold high-resolution lung imaging is challenging due to the short T
2*
and scan efficiency. The stack-of-spirals UTE enables fast lung imaging in a
single breath hold. However, fast scanning of stack-of-spirals results in a
long readout time, introducing adverse effects of the short T
2*
of the lung. Herein, we proposed a density-controllable spiral trajectory
(DCST) design method to design a short readout time trajectory concurrently
satisfying the criteria of optimal compressed sensing (CS), reducing the
adverse effect of the short T
2*. The short readout time
trajectory combined with CS-UTE improves the single breath-hold high-resolution
lung imaging.
Introduction
Stack-of-spirals UTE using the less shot numbers
allows lung imaging under single breath-hold acquisition (1-2). However, the
readout time of the stack-of-spirals is long under high-resolution single
breath-hold imaging, which gives rise to adverse effects resulting from the
short T2* relaxation time of the lung, including blurring
and signal-to-noise ratio decreasing (3). In this study, we propose a
density-controllable spiral trajectory design method to reduce the readout time
in stack-of-spirals UTE. This method includes two parts: (i) the
high-sampling-rate numerical multi-density spliced path, which satisfies
optimal CS criteria and (ii) the gradient-constrained discrete k-space
trajectory pixel-wise selection. The reconstruction was implemented by CS (4). The UTE
sequence employs a short SLR pulse and slice-selective gradient for signal excitation
of the slab region, followed by a gradient merged by the slice-rephasing and
slice encoding gradient, and finally uses the spiral readout gradient. In
addition, the phantom experiment and one healthy volunteer lung imaging were
implemented.Methods
This
study was approved by the National Center for Nanoscience and Technology Ethics
Board, and written informed consent was given by all study participants. All MR
images were obtained with a 1.5-T scanner (Wandong Medical Inc, Beijing, China).
The single-channel body coil was used in the phantom imaging. The six-channel
chest coil was used in the volunteer lung imaging. The DCST design method
includes two parts shown in Figure 1 (a). First, towards the targeted density, the
analytic formula of the spiral (Equation [1] and [2]) was used to generate a
high-sampling-rate numerical path by regulating the deviation of r and the deviation
of φ. Each density region was
smooth-like spliced. Second, according to the allowed gradient amplitude and
slew rate, the k-space trajectory was searched pixel by pixel along the high-sampling-rate
numerical path.
x=rcos(ϕ) [1]
y=rsin(ϕ) [2]
The
undersampling trajectory (32 arms), the long readout time trajectory (32 arms),
and the short readout time trajectory (64 arms) were designed as shown in
Figure 1 (b). The full sampling was performed in the core region of the k-space,
and the multi-density undersampling was performed in the peripheral area of the k-space. The maximum gradient amplitude and slew rate used in this research
were 27 mT/m and 60 T/m/s, respectively. The UTE sequence is based on a stack-of-spirals frame.
As shown in Figure 2, a 100 μs SLR pulse was employed, followed by a gradient which
was merged by the slice rephasing gradient and the slice encoding gradient. First,
the 30cm slice thickness region was excited. Then, the spiral gradient was
implemented, and the signal was acquired concurrently. The CS reconstructed method used in our experiments
has been published by Lustig et al. (4). The
large ACR phantom was used in this experiment, as shown in Figure 3. The lung
imaging of the healthy volunteer is shown in Figure 4. The volunteer was
required to breath-hold in each scan. The scan parameters are shown in Table 1.Results
Images of the ACR phantom in Figure 3 show that the
short-readout time full sampling performed best. The long-readout time full
sampling performed worst. The direct Fourier reconstructed undersampling image
shows blurring but preserves the structure. The CS-reconstructed undersampling
image improves the blurring and appears closer to the image of the
short-readout time full sampling. In Figure 4, the scan time of the undersampling
lung imaging is the shortest. Clear distinction in lung structure is shown in
the CS-reconstructed image, while the lung image of the long readout time full
sampling performs ambiguous structure. The short readout time full sampling
appeared middle in lung imaging.Discussion
Herein, we presented a spiral trajectory design
method, which can control the density in radial and angular directions of
k-space. The density-controllable pattern is suitable for the optimal CS
criteria. Regarding the contradiction between the readout time and the scan
time of stack-of-spirals UTE, this method provided a significantly reduced
readout time trajectory and further improved the blurring in single breath hold
lung imaging for high resolution. The lung imaging of the undersampling CS
reconstruction performed slightly better than the short readout time full
sampling, maybe because the scan time of 38.4 seconds is hard for a single
breath hold. Although, the slice encoding efficiency is low in our experiments,
which leads to the long TE. Our method can be implemented in other
high-efficiency slice encoding patterns, including acquisition-weighted UTE
(5). Noteworthy, the design method can be extended to different approximately
continuous trajectory designs.Conclusion
The proposed method provides an easy approach to
designing a trajectory in which the density is controllable in both radial and
angular directions. In addition, shortening readout time and CS reconstruction
improve the blurring and signal-to-noise ratio of high-resolution lung imaging
in single breath-hold acquisition.Acknowledgements
This
work was supported by National Natural Science Foundation of China
(NO.61971151) and Wandong Medical.References
(1) Mugler
JP et al. Proc Intl Soc Mag Reson Med 2015; 23:1476.
(2) Zeimpekis
K et al. Proc Intl Soc Mag Reson Med 2016; 24:1608.
(3) Willmering
MM, Robison RK, Wang H, Pipe JG, Woods JC. Implementation of the FLORET UTE
sequence for lung imaging. Magnetic Resonance in Medicine 2019; 82:1091–1100.
(4) Lustig
M, Donoho D, Pauly JM. Sparse MRI: The Application of Compressed Sensing
for
Rapid MR Imaging. Magnetic Resonance in Medicine 2007; 58:1182–1195.
(5) Qian YX, Boada FE. Acquisition-weighted stack of
spirals for fast high-resolution three-dimensional ultra-short echo time MR
imaging. Magnetic Resonance in Medicine 2008; 60:135–145.