Eric Y. Pierre1, David N. Vaughan1,2, Warda T. Syeda3, Bahman Tahayori1, Heath R. Pardoe1, David F. Abbott1, and Graeme D. Jackson1
1Imaging and Epilepsy, The Florey Institute of Neuroscience and Mental Health, Melbourne, Australia, 2Department of Neurology, Austin Health, Melbourne, Australia, 3Department of Psychiatry, The University of Melbourne, Melbourne, Australia
Synopsis
Keywords: Quantitative Imaging, Quantitative Susceptibility mapping, Reproducible Research
We propose the
customization of a commercially-available phantom design to be used for QSM reproducibility
research with long-term stability, allowing reproduction of the phantom itself
commercially. The proposed phantom is based on master-dilution of Gadolinium in
pure water vials, offering long-term stability, with theoretical susceptibility
range chosen to match normal tissue and common pathologies in the brain. We evaluate the suitability of the phantom for QSM reproducible research for different
sequences, orientation within the scanner and regularization algorithms.
INTRODUCTION
QSM validation is challenging but necessary given the vast
parameter space of acquisition and pipelines. To this end many QSM-phantom MR
studies have been done in experimental settings(1–5).
Unlike other quantitative methods like T1-, T2- or
Diffusion mapping(6), to date no QSM standardization phantom is
available commercially, limiting reproducibility analysis mostly to groups with
phantom-design capabilities.
One adverse factor for commercial QSM phantoms is the
challenge posed by long-term stability. Gel-based solutions of susceptibility
agents (e.g. Gadolinium) can mimic tissue properties well(2,3), and do not have the kind of boundary
issues introduced by thick rigid compartments, but do not offer long-term
stability of ground-truth parameters(7).
This work aims to demonstrate the possibility of
repurposing a commercially available phantom design to be used for QSM, with a
view to help democratize long-term reproducibility and standardization studies
in QSM.METHODS - PHANTOM DESIGN
The proposed phantom is adapted from the CaliberMRI
Diffusion phantom design (https://qmri.com) and ordered from the same
manufacturer as a custom design.
A diagram of the phantom is shown in figure 1: 13 vials of 50mL each were filled with different Gd
concentrations as shown in the figure by serial dilution, with initial
concentration confirmed by Inductively-Coupled-Plasma Mass-Spectrometer.
The
concentration range was chosen to cover susceptibility measurements in normal
brain tissue and some common pathologies(8). An additional 3 vials of pure water were
added for reference and orientation. The vials are held in place by an acrylic
plate. The interstitial liquid was pure water.
The phantom also has 10 MR-readable temperature
crystal probes ranging from 15°C to 24°C. METHODS - QSM IMAGING
Imaging was performed on a Siemens 3T Prisma-fit system
(Siemens Healthineers, Erlangen, Germany) with two QSM acquisitions tailored
for epilepsy cohorts(9):
- a 5-echo 3D GRE (5e-GRE) at 1.2mm3 voxels, FOV 230x187x144mm, TE1/DTE/TR=5.84/4.79/30ms, FA 15°, GRAPPA 3 in-plane acceleration, (acquisition
time 3:39 min)
- a 6-echo 3D GRE (6e-GRE) at 1.3 mm3, FOV
208x182x136mm, TE1/DTE/TR=5.38/4.15/30ms, GRAPPA 2 in-plane acceleration.
Phantom
was orientated in the scanner first with vial-axis perpendicular to the B0
field then parallel to it. The experiment was repeated after two weeks on the
same scanner.
METHODS – QSM COMPUTATION
Muti-echo GRE data was analyzed with the MEDI+0 method(10), over a mask of the whole phantom-sphere. Each
Gd-solution, pure water and acrylic plate (AcP) region was segmented by k-means
clustering and connectivity analysis from the magnitude-image averaged across
echoes.
The AcP region was used as an L2-regularization uniformity-penalty and
zero-referencing region(10). The pure-water region for zero-reference
and uniformity-penalty was also used for comparison, as well as the MEDI
algorithm with no zero-reference(11,12).
The upper half of each vial was
excluded in the ROI analysis due to their complex geometry close to the
cap. RESULTS
The built-in thermometric probes indicated a phantom
temperature between 20°C and 21°C for all imaging experiments.
Example QSM maps for the different imaging setting are shown
in figure 2. At perpendicular B0 orientation, streaking boundary artifacts are
visible close to the vial walls. As expected from theory(1), these artifacts are noticeably reduced when
orientation is parallel to the field. Discounting orientation and resolution
differences, 5e-GRE and 6e-GRE produced very similar maps regardless of field
orientation.
The mean QSM values measured in vial ROIs are analyzed in
figure 3. Both sequences were in good-agreement within a +/-0.047ppm 95%CI and a slight bias of -0.01ppm. For
test-retest analysis of the same sequence, the difference CI increased to
+/-0.078. These results appeared to be mostly driven by the vials corresponding
to pathological tissue ranges (0.25-0.93ppm).
All experiments yielded excellent linear fits with r2
> 0.996, with r the Pearson coefficient. However, the slope of the fit, corresponding
to the molar susceptibility of Gd in water, was noticeably different between
field orientations, dropping from 301 ppm.M-1 parallel to the field
to 255 ppm.M-1, against a theoretical molar susceptibility of 320
ppm.M-1(1).
Comparison between the MEDI and MEDI+0 algorithms are shown
in figure 4. While all regressions had r2 >0.98, the regression
coefficients yielded a Gd-molar-susceptibility < 300 ppm.M-1 without
a zero-reference, or by using pure-water as zero-reference. By comparison, the AcP-region
zero-reference had improved r2 and a measured molar susceptibility closer to
theoretical value. Discussion and Conclusion
The standard deviation within ROIs indicated a
slight overlap of measured susceptibility from one vial to the next, suggesting
that the phantom design could be optimized by increasing the concentration-step
between vials. Regardless, the proposed phantom achieved excellent linear
fitting of mean susceptibility values against known concentrations of
Gd-solutions (r2>0.996).
Interestingly, using an Acrylic Plate
mask for zero-referencing and uniformity regularization appeared to noticeably
improve QSM regression with MEDI+0 over MEDI with no regularization, while pure-water
zero-referencing had the opposite effect.
With vials parallel to the field, values appeared
very stable against change in number of echoes and voxel size, with detected
Molar Susceptibility close to the theoretical value, with a slight
underestimation bias which remains to be investigated. The 95%CI in test-retest analyses were mostly driven by large Gd concentrations, but an order of magnitude less than the ranges themselves, indicating
suitability for neuroimaging studies.
Finally, the proposed customized phantom is commercially-available,
making it easily replicable. The water-based Gd solution yielded good agreement
to theoretical measurements while offering long-term stability of ground-truth susceptibility
parameters. Acknowledgements
We
acknowledge the facilities and scientific and technical assistance of the
National Imaging Facility, a National Collaborative Research Infrastructure
Strategy (NCRIS) capability, at The Florey Institute of Neuroscience and Mental
Health. We also acknowledge the strong support from the Victorian Government
and in particular the funding from the Operational Infrastructure Support Grant,
and support from the Victorian Biomedical Imaging Capability (VBIC). The
Australian Epilepsy Project received funding from the Australian Government
under the Medical Research Future Fund.
The authors would also like to thank William Hollander, Todor Karaulanov
and Chamni Jayarathna for valuable input to this project, and Remika Mito and Lea Vinokur for help with editing this abstract.
References
1.
Olsson E, Wirestam R, Lind E. MRI-Based Quantification of Magnetic
Susceptibility in Gel Phantoms: Assessment of Measurement and Calculation
Accuracy. Radiol Res Pract 2018;2018:1–13 doi: 10.1155/2018/6709525.
2.
Kato H, Kuroda M, Yoshimura K, et al. Composition of MRI phantom equivalent to
human tissues. Med Phys 2005;32:3199–3208 doi: 10.1118/1.2047807.
3.
Hattori K, Ikemoto Y, Takao W, et al. Development of MRI phantom equivalent to
human tissues for 3.0-T MRI. Med Phys 2013;40 doi:
10.1118/1.4790023.
4. Deh K, Kawaji K, Bulk M, et al. Multicenter reproducibility
of quantitative susceptibility mapping in a gadolinium phantom using MEDI+0
automatic zero referencing. Magn Reson Med 2019;81:1229–1236 doi:
10.1002/mrm.27410.
5.
Hobson N, Polster SP, Cao Y, et al. Phantom validation of quantitative
susceptibility and dynamic contrast-enhanced permeability MR sequences across
instruments and sites. Journal of Magnetic Resonance Imaging 2020;51:1192–1199
doi: 10.1002/jmri.26927.
6.
Stupic KF, Ainslie M, Boss MA, et al. A standard system phantom for magnetic
resonance imaging. Magn Reson Med 2021;86:1194–1211 doi: 10.1002/mrm.28779.
7.
Er WX, Lim WJ, Dwihapsari Y, Awang MNA, Yusoff AN. Signal-to-noise ratio
uniformity and stability of agar gel phantom with iron (III) oxide as
relaxation modifier. Beni Suef Univ J Basic Appl Sci 2021;10 doi:
10.1186/s43088-021-00173-0.
8.
Girard R, Fam MD, Zeineddine HA, et al. Vascular permeability and iron
deposition biomarkers in longitudinal follow-up of cerebral cavernous
malformations. J Neurosurg 2017;127:102–110 doi: 10.3171/2016.5.JNS16687.
9.
The Australian Epilepsy Project, https://epilepsyproject.org.au.
10.
Liu Z, Spincemaille P, Yao Y, Zhang Y, Wang Y. MEDI+0: Morphology enabled
dipole inversion with automatic uniform cerebrospinal fluid zero reference for
quantitative susceptibility mapping. Magn Reson Med 2018;79:2795–2803 doi:
10.1002/mrm.26946.
11.
Liu T, Xu W, Spincemaille P, Avestimehr AS, Wang Y. Accuracy of the morphology
enabled dipole inversion (MEDI) algorithm for quantitative susceptibility
mapping in MRI. IEEE Trans Med Imaging 2012;31:816–824 doi:
10.1109/TMI.2011.2182523.
12.
Liu T, Wisnieff C, Lou M, Chen W, Spincemaille P, Wang Y. Nonlinear formulation
of the magnetic field to source relationship for robust quantitative
susceptibility mapping. Magn Reson Med 2013;69:467–476 doi: 10.1002/mrm.24272.