Camille Dubuc1,2, Aimé Labbé1, Sina Marhabaie1, Clément Thibault3,4, Jean-Christophe Ginefri1, Javier Briatico5, and Marie Poirier-Quinot1
1Université Paris-Saclay, BioMaps, ORSAY, France, 2Thales Research & Technology, Palaiseau, France, 3Université Paris-Saclay, CEA, CNRS, Inserm, BioMaps, Orsay, France, 4Université Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif-sur-Yvette, France, 5Unité Mixte de Physique, CNRS, Thales, Université Paris-Saclay, Palaiseau, France
Synopsis
Keywords: Non-Array RF Coils, Antennas & Waveguides, Non-Array RF Coils, Antennas & Waveguides, Superconducting material, Decoupling
A promising strategy to tackle the issue of decoupling High-Temperature Superconducting (HTS) surface coils during emission is to exploit nonlinearities of the HTS coil electromagnetic response. In this work, we present the first Magnetic Resonance Images acquired with a decoupled HTS surface coil and the required coil’s features to obtain these images. The change of the HTS coil Q-factor values leads to different decoupling levels (DL) of the HTS coil and the presence or absence of B1 artifacts on the images.Purpose
The high sensitivity of HTS surface coils is of
profound interest as it allows to achieve at 1.5 T a comparable Signal to Noise
Ratio to the one at 9T (1) without SAR issues, B1+-transmit
field inhomogeneities and with the same NMR contrast as in clinical
applications. However, decoupling the HTS coil from the transmit coil remains a
major issue to overcome. Since conventional decoupling approaches for copper
coils are not suitable for HTS coils, new decoupling strategies are necessary. It
has been reported that the Q-factor drops nonlinearly as the transmit RF-field amplitude
increases (2-3). Consequently, the Q-factor during transmission Qt, can decrease by several
orders of magnitude compared to the one involved during reception Qr. Here, we
present a decoupling method exploiting the non-linear electrical properties of HTS
coils as a function of B1+, the transmit
power level of the body coil, allowing the material to
pass from a near zero-resistance state to a dissipative state. The efficiency
of the decoupling of the HTS coil with the body coil during transmission is then
evaluated using a decoupling level DL (4).Methods
The studied HTS coil is a 6 turn transmission line resonator made of 300 nm thick YBaCuO deposited on a 330 µm thick sapphire r-cut dielectric substrate (Ceraco ceramic coating GmbH, Germany). The HTS coil has been operated in a MR-compatible cryostat (5) after a field-cooling(6). First, Q-factors were directly measured when the HTS coil is placed inside the 1.5T scanner using a Vector Network Analyser (VNA) and the single loop probe of diameter of 5 mm (7). Secondly, MR images of a parallelepipedic Gd-doped water sample (
T1 = 0.0952 s) were acquired using the HTS coil during reception. . The sequence parameters used are : repetition time
TR = 25.0 ms ; echo time
TE = 6.0 ms ; image size : 60 × 144 × 144 voxels ; resolution : 0.5 mm × 0.21 mm × 0.21 mm. Two flips angle, α of 5° and 10°, were used obtained with the following RF pulse parameters, B1+, pulse duration, τ:
α | B1+, τ | B1+, τ |
5° | 0.5, 6.5 ×10-4 s | 1, 3.2 ×10-3 s |
10° | 1, 6.5 ×10-4 s | 10, 6.5 ×10-5 s |
MR signal received by the HTS coil was also computed as a function of B
1+-transmit amplitude, sequence parameters and sample properties, using the equations below. The Q-factor values, Q
t, Q
r, involved in the expression of the Concentration Factor F
c, were then choosen so as to adjust the simulated MR-signal maps to the experimental ones. Obtained Q-factor values are compared to those measured with the VNA.
$$ Signal = |\frac{B_1\times M_0 \times (1-e^{-T_1/T_R})}{\sin(\alpha \sqrt{1+F_c^2 \cos^2(\theta)})\times (1-e^{-T_1/T_R}\cos(\alpha \sqrt{1+F_c^2 \cos^2(\theta)}))}| $$
$$ F_c = \frac{S_{HTS}}{2L_{HTS}}10^{\frac{Q_t}{20 Q_r}} Q_r B_1$$
and $$ \theta = \arctan(\frac{B_x^{HTS}}{B_z^{HTS}})$$
Finally, the decoupling level is defined with values Q
r and Q
t as :
$$ DL = -20\times \log (\frac{I_{emission}}{I_{reception}})= -20\times \log (\frac{Q_{t}}{Q_{r}})$$
Results / Discussion
The measured Q-factor variation as a function of the RF-field amplitude produced by the single-loop probe is displayed in figure 1. It is representative for the nonlinear behavior of the HTS resistivity. For B1 < 0.05 µT the Q-factor is constant and sets Qr of 1650. Up to 0.5 µT, the Q-factor decreases down to 1250 at B1 = 0.5 µT.
Figure 2 displays the experimental and simulated MR-signal maps, which correspond to different Qt factors and DLs. There is a good agreement between the experimental results and the simulations. For B1 = 0.5 µT extracted Qt value of 1150± 50 obtained to adjust the simulated map is in good agreement with the one directly measured with VNA, equals to 1200. Qt of 850 and 50 were respectively obtained for B1 of 1 µT and10 µT.
For B1 up to1 µT, figure 2 highlights the concentration effect of the RF field by the non-decoupled HTS coil. The darker stripes correspond to a flip angle α equal to nπ rad (with n being an integer) higher than the nominal flip angle of 5.0° or 10.0° (8).
For the smaller Qt, obtained with a B1+ of 10 µT, a DL of 30 dB is then calculated and the effect of the reinduced magnetic field by the HTS coil during thansmission is no longer visible on the SNR map. Conclusion
For B1 = 0.5 µT, excellent fit between experimental and simulated MRI signal maps is obtained for a Qt value of 1150±50. This value is in good agreement with the measured one of 1250. Figure 2 illustrates the impact of superconducting state of the HTS coil on B1-artifacts. A high dissipative state is reached, leading to the absence of B1-artifacts, corresponding to a Qt of 50 and a DL up to 30 dB.
This study confirms that nonlinearities of HTS surface coils electric response can be exploited for decoupling the HTS coil during the transmission. Moreover, this decoupling strategy can be applied independently of the HTS geometry, e.g multi turn multi loop transmission line resonator. Finally, thanks to this essential outcome, quantification sequences requiring a full control of the flip angles can now be used to perform MRI with such HTS surface coils.
Acknowledgements
No acknowledgement found.References
1. Marie
Poirier-Quinot, Jean-Christophe Ginefri, Olivier Girard, Philippe Robert, and
Luc Darrasse. Performance of a
miniature high-temperature superconducting (HTS) surface coil for in vivo
microimaging of the mouse in a standard 1.5 T clinical whole-body scanner. Magnetic Resonance in Medicine, 60(4),
feb 2008.
2. Olivier
Girard, Jean-Christophe Ginefri, Marie Poirier-Quinot, and Luc Darrasse. Method for nonlinear
characterization of radio frequency coils made of high temperature
superconducting material in view of magnetic resonance imaging applications. Review of Scientific Instruments, American
Institute of Physics, 78(12), dec 2007.
3. I.
Saniour, G. Authelet, B. Baudouy, R-M. Dubuisson, C. J. van der Beek, L.
Darrasse, J. Briatico, J-C. Ginefri,
and M. Poirier-Quinot. Novel passive decoupling approach for high resolution
HTS RF coils based on the nonlinear electrical properties of superconductors. ESMRMB, Rotterdam, Netherlands, 2019.
4. Sina
Marhabaie, Aime Labbe, B. Q., and Marie POIRIER-QUINOT. Quantitative analysis of minimum satisfactory
detuning level of receive-only surface coils used in magnetic resonance
imaging. Submitted, 2022.
5. Isabelle
Saniour, Gilles Authelet, Bertrand Baudouy, Rose-Marie Dubuisson, Georges
Willoquet Laurène Jourdain, Luc Darrasse, Jean-Christophe Ginefri, and Marie
Poirier-Quinot. A temperature controlled
cryogen free cryostat integrated with transceiver-mode superconducting coil for
high resolution magnetic resonance imaging. Review of Scientific Instruments.
6. Aime
Labbe, Rose-Marie DUBUISSON, Jean-Christophe GINEFRI, Cornelis J. van der BEEK,
Luc DARRASSE, and Marie POIRIER-QUINOT. Static field homogeneity artifacts due to magnetic flux expulsion by HTS
coils for high-resolution magnetic resonance imaging. Applied Physics Letter, dec 2020.
7. Jean-Christophe Ginefri, Emmanuel Durand, and
Luc Darrasse. Quick measurement of nuclear magnetic resonance coil sensitivity
with a single-loop probe. Rev. Sci.
Instrum. 70, 4730 (1999); doi: 10.1063/1.1150142
8. W.A. EDELSTEIN, C.J. HARDY, and O.M.
MUELLER. Electronic
decoupling of surface-coil receivers for nmr imaging and spectroscopy. Journal of Magnetic Resonance, 67, 1986.